Applicability of bacterial endotoxins test to various blood products by the use of endotoxin-specific lysates

被引:11
作者
Ochiai, Masaki [1 ]
Yamamoto, Akihiko [2 ]
Naito, Seishiro [1 ]
Maeyama, Jun-ichi [3 ]
Masumi, Atsuko [3 ]
Hamaguchi, Isao [3 ]
Horiuchi, Yoshinobu [2 ]
Yamaguchi, Kazunari [1 ,3 ]
机构
[1] Natl Inst Infect Dis, Div Qual Assurance, Musashimurayama, Tokyo 2080011, Japan
[2] Natl Inst Infect Dis, Dept Bacterial Pathogenesis & Infect Control, Musashimurayama, Tokyo 2080011, Japan
[3] Natl Inst Infect Dis, Dept Safety Res Blood & Biol, Musashimurayama, Tokyo 2080011, Japan
关键词
Endotoxin; Pyrogen; LAL test; Blood products; Quality control; RABBIT PYROGEN TEST; LIMULUS; COAGULATION; PROTEIN; LAL;
D O I
10.1016/j.biologicals.2010.07.003
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Endotoxin contamination is a serious threat to the safety of parenteral drugs and the rabbit pyrogen test has played a crucial role in controlling this contamination Although the highly sensitive endotoxin test has replaced the pyrogen test for various pharmaceuticals the pyrogen test is still implemented as the control test for most blood products in Japan We examined the applicability of the endotoxin test to blood products for reliable detection and quantification of endotoxin Nineteen types of blood products were tested for interfering factors based on spike/recovery of endotoxin by using 2 types of endotoxin-specific lysate reagents for photometric techniques Interfering effects on the endotoxin test by the products could be eliminated by diluting from 1/2 to 1/16 with the exception of antithrombin III However conventional lysate reagents that also react with non-pyrogenic substances such as (1-3)-beta-D-glucan produced results that were not relevant to endotoxin content or pyrogenicity Our results showed that the endotoxin test would be applicable to most blood products if used with appropriate endotoxin-specific lysate reagents (C) 2010 The International Association for Biologicals Published by Elsevier Ltd All rights reserved
引用
收藏
页码:629 / 636
页数:8
相关论文
共 31 条
[1]  
[Anonymous], EUR PHARM
[2]  
[Anonymous], 2008, US PHARM
[3]   APPLICATION OF THE LIMULUS TEST FOR PRACTICAL QUALITY-CONTROL ON ENDOTOXIN CONTENT IN COMMERCIAL HUMAN SERUM-ALBUMIN (HSA) PRODUCTS - IN COMPARISON WITH THE RABBIT PYROGEN TEST [J].
ASAKAWA, S ;
FUJIWARA, H ;
NAITO, S ;
HOMMA, R ;
ISHIDA, S ;
CHINO, F ;
TSUCHIYA, M ;
MATSUURA, S ;
TANAKA, S ;
OHKI, M .
YAKUGAKU ZASSHI-JOURNAL OF THE PHARMACEUTICAL SOCIETY OF JAPAN, 1994, 114 (11) :888-893
[4]  
Cooper JF, 1997, PDA J PHARM SCI TECH, V51, P2
[5]  
CROWE BA, 2006, EUR GMP C ALT PYR TE
[6]  
*EUR DIR QUAL MED, 2010, EUR PHARM S, V6
[7]  
Finney DJ, 1978, STAT METHODS BIOL AS
[8]  
GRISMAN SE, 1969, P SOC EXP BIOL MED, V131, P1154
[9]   Detection of pyrogens in intravenous IgG preparations [J].
Huszár, G ;
Jenei, B ;
Szabó, G ;
Medgyesi, GA .
BIOLOGICALS, 2002, 30 (02) :77-83
[10]  
Iwanaga S, 1998, J BIOCHEM, V123, P1