Application of the row action maximum likelihood algorithm with spherical basis functions to clinical PET imaging

被引:85
作者
Daube-Witherspoon, ME [1 ]
Matej, S [1 ]
Karp, JS [1 ]
Lewitt, RM [1 ]
机构
[1] Univ Penn, Dept Radiol, Philadelphia, PA 19104 USA
关键词
D O I
10.1109/23.910827
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Three-dimensional (3D) reconstructions from fully 3D positron emission tomography (PET) data can yield high-quality images but at a high computational cost. The 3D row action maximum likelihood algorithm (3D RAMLA) with spherically-symmetric basis functions (blobs) has recently been modified to reconstruct multi-slice 2D PET data after Fourier rebinning (FORE) but still using 3D basis functions (2.5D RAMLA). In this study 2.5D RAMLA and 3D RAMLA were applied to several patient and phantom PET datasets to assess their clinical performance. RAMLA performance was compared to that for the reconstruction techniques in routine clinical use on our PET scanners. Torso phantom and whole-body patient scans acquired on the C-PET scanner were reconstructed after FORE with filtered back-projection (FORE+FBP), the ordered subsets expectation maximization algorithm (FORE+OSEM), and FORE+2.5D RAMLA for various reconstruction parameters. The 3D Hoffman brain phantom scanned on the HEAD Penn-PET scanner was reconstructed with the 3D reprojection algorithm (3DRP) and 3D RAMLA, as well as FORE+FBP, FORE+OSEM, and FORE+2.5D RAMLA. Our results demonstrate improvement of 3D and 2.5D RAMLA with blob basis functions, compared to the reconstruction methods currently in clinical use, in terms of contrast recovery and noise, especially in regions of Limited statistics.
引用
收藏
页码:24 / 30
页数:7
相关论文
共 27 条
  • [1] A row-action alternative to the EM algorithm for maximizing likelihoods in emission tomography
    Browne, J
    DePierro, AR
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1996, 15 (05) : 687 - 699
  • [2] Exact and approximate rebinning algorithms for 3-D PET data
    Defrise, M
    Kinahan, PE
    Townsend, DW
    Michel, C
    Sibomana, M
    Newport, DF
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (02) : 145 - 158
  • [3] DEPIERRO AR, 1989, MAT APL COMPUT, V8, P153
  • [4] Farquhar TH, 2000, J NUCL MED, V41, P745
  • [5] DESIGN AND PERFORMANCE OF THE HEAD PENN-PET SCANNER
    FREIFELDER, R
    KARP, JS
    GEAGAN, M
    MUEHLLEHNER, G
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) : 1436 - 1440
  • [6] Guzzardi R, 1991, J Nucl Biol Med, V35, P141
  • [7] ALGEBRAIC RECONSTRUCTION TECHNIQUES CAN BE MADE COMPUTATIONALLY EFFICIENT
    HERMAN, GT
    MEYER, LB
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1993, 12 (03) : 600 - 609
  • [8] ASSESSMENT OF ACCURACY OF PET UTILIZING A 3-D PHANTOM TO SIMULATE THE ACTIVITY DISTRIBUTION OF [18F]FLUORODEOXYGLUCOSE UPTAKE IN THE HUMAN BRAIN
    HOFFMAN, EJ
    CUTLER, PD
    GUERRERO, TM
    DIGBY, WM
    MAZZIOTTA, JC
    [J]. JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1991, 11 (02) : A17 - A25
  • [9] Huang SC, 1999, IEEE T NUCL SCI, V46, P603
  • [10] ACCELERATED IMAGE-RECONSTRUCTION USING ORDERED SUBSETS OF PROJECTION DATA
    HUDSON, HM
    LARKIN, RS
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (04) : 601 - 609