Characterization of a double probe for local pulse wave velocity assessment

被引:13
|
作者
Pereira, H. Catarina [1 ,2 ]
Pereira, Tania [1 ]
Almeida, Vania [1 ]
Borges, Elisabeth [1 ]
Figueiras, Edite [1 ]
Simoes, J. Basilio [1 ,2 ]
Malaquias, J. Luis [2 ]
Cardoso, Joao M. R. [1 ]
Correia, Carlos M. B. [1 ]
机构
[1] Univ Coimbra, Instrumentat Ctr, Elect & Instrumentat Grp, Coimbra, Portugal
[2] ISA, Coimbra, Portugal
关键词
haemodynamics; local pulse-wave velocity; double piezoelectric probe; test bench; impulse response; EXPERT CONSENSUS DOCUMENT; CAROTID-ARTERY; STIFFNESS; PRESSURE; ALGORITHM; SENSOR;
D O I
10.1088/0967-3334/31/11/004
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Local pulse-wave velocity (PWV) is an accurate indicator of the degree of arteriosclerosis (stiffness) in an artery, providing a direct characterization of the properties of its wall. Devices currently available for local PWV measurement are mainly based on ultrasound systems and have not yet been generalized to clinical practice since they require high technical expertise and most of them are limited in precision, due to the lack of reliable signal processing methods. The present work describes a new type of probe, based on a double-headed piezoelectric (PZ) sensor. The principle of PWV measurement involves determination of the pulse transit time between the signals acquired simultaneously by both PZs, placed 23 mm apart. The double probe (DP) characterization is accomplished in three main studies, carried out in a dedicated test bench system, capable of reproducing a range of clinically relevant properties of the cardiovascular system. The first study refers to determination of the impulse response (IR) for each PZ sensor, whereas the second one explores the existence of crosstalk between both transducers. In the last one, DP time resolution is inferred from a set of three different algorithms based on (a) the maximum of cross-correlation function, (b) the maximum amplitude detection and (c) the zero-crossing point identification. These values were compared with those obtained by the reference method, which consists of the simultaneous acquisition of pressure waves by means of two pressure sensors. The new probe demonstrates good performance on the test bench system and results show that the signals do not exhibit crosstalk. A good agreement was also verified between the PWV obtained from the DP signals (19.55 +/- 2.02 ms(-1)) and the PWV determined using the reference method (19.26 +/- 0.04 ms(-1)). Although additional studies are still required, this probe seems to be a valid alternative to local PWV stand-alone devices.
引用
收藏
页码:1449 / 1465
页数:17
相关论文
共 50 条
  • [21] In Vitro Validation of 4D Flow MRI for Local Pulse Wave Velocity Estimation
    Ruesink, Timothy
    Medero, Rafael
    Rutkowski, David
    Roldan-Alzate, Alejandro
    CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2018, 9 (04) : 674 - 687
  • [22] On the Feasibility of Pulse Wave Velocity Imaging for Remote Assessment of Physiological Functions
    Saiko, Gennadi
    Dervenis, M.
    Douplik, A.
    OXYGEN TRANSPORT TO TISSUE XLII, 2021, 1269 : 393 - 397
  • [23] Pulse Wave Velocity in Atherosclerosis
    Kim, Hack-Lyoung
    Kim, Sang-Hyun
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2019, 6
  • [24] ASSESSMENT OF CAROTID PULSE WAVE VELOCITY (CARPWV) IN PATIENTS WITH ANKYLOSING SPONDYLITIS
    Becerra Carlos, Ramos
    Alanis-Sanchez, Guillermo
    Cardona-Munoz, Ernesto
    Cardona-Mueller, David
    Khettab, Hakim
    JOURNAL OF HYPERTENSION, 2018, 36 : E259 - E259
  • [25] Assessment of aortic pulse wave velocity by ultrasound: a feasibility study in mice
    Faita, Francesco
    Di Lascio, Nicole
    Stea, Francesco
    Kusmic, Claudia
    Sicari, Rosa
    MEDICAL IMAGING 2014: ULTRASONIC IMAGING AND TOMOGRAPHY, 2014, 9040
  • [26] Current assessment of pulse wave velocity: comprehensive review of validation studies
    Milan, Alberto
    Zocaro, Gaia
    Leone, Dario
    Tosello, Francesco
    Buraioli, Irene
    Schiavone, Domenica
    Veglio, Franco
    JOURNAL OF HYPERTENSION, 2019, 37 (08) : 1547 - 1557
  • [27] Local carotid artery pulse wave velocity: foot to foot or dicrotic notch?
    Muiesan, Maria Lorenza
    Paini, Anna
    JOURNAL OF HYPERTENSION, 2009, 27 (10) : 1962 - 1963
  • [28] Pulse Wave Velocity Prediction and Compliance Assessment in Elastic Arterial Segments
    Lillie, Jeffrey S.
    Liberson, Alexander S.
    Mix, Doran
    Schwarz, Karl Q.
    Chandra, Ankur
    Phillips, Daniel B.
    Day, Steven W.
    Borkholder, David A.
    CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2015, 6 (01) : 49 - 58
  • [29] On the Design of Passive Resonant Circuits to Measure Local Pulse Wave Velocity in a Stent
    Schaechtele, Jonathan
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2016, 10 (03) : 612 - 622
  • [30] Signal Processing For Local Pulse Wave Velocity Estimation Based On Fourier Decomposition
    Mo, Hong
    Zhang, Yufeng
    De, Li
    Li, Zhiyao
    Qi, Guiling
    Zhao, Zhengpeng
    2019 2ND IEEE INTERNATIONAL CONFERENCE ON INFORMATION COMMUNICATION AND SIGNAL PROCESSING (ICICSP), 2019, : 87 - 91