Validation of the AGILIS ambulatory blood pressure monitor according to the European Society of Hypertension International protocol for validation of blood pressure measuring devices in adults

被引:9
|
作者
Mallion, JJM [1 ]
Pierre, H [1 ]
Neuder, Y [1 ]
Ormezzano, O [1 ]
Baguet, JP [1 ]
机构
[1] CHU Grenoble, Dept Cardiol & Hypertens Arterielle, F-38043 Grenoble, France
关键词
human; hypertension; blood pressure; ambulatory blood pressure measurement; clinical validation;
D O I
10.1097/00126097-200504000-00008
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background We evaluated the clinical validity of the AGILIS ambulatory blood pressure monitor according to the European Society of Hypertension International Protocol. Methods The blood pressure (BP) was measured at the same arm by two certified observers using two mercury columns and a double stethoscope, then by the volunteer with the automatic device, these measurements were done seven times sequentially using 33 volunteers. For the randomization of the patient, one ABPM measure was compared to the mean of three manual measures. Then, we had three ABPM measures and four manual measures. Each ABPM measure, being compared to two manual measures. Blood pressure readings and recordings were blinded. Position of the subjects was standardized. Results Phase 1: To pass phase 1, the device had to have at least 25 of the 45 measurements within 5 mmHg, 35 within 10 mmHg or 40 within 15 mmHg of the comparative observer measurements. The AGILIS monitor had 35 measurements lying within 5 mmHg, 42 within 10 mmHg and 45 within 15 mmHg for systolic BP (SBP), 35 within 5 mmHg, 44 within 10 mmHg and 45 within 15 mmHg for diastolic BP (DBP). It passed the criteria for SBP and DBR Phase 2: To pass phase 2.1, a device must have at least 60 of 99 measurements within 5 mmHg, 75 within 10 mmHg and 90 within 15 mmHg of the comparative observer measurements, and in addition must also have either 65 within 5 mmHg and 80 within 10 mmHg, or 65 within 5 mmHg and 95 within 15 mmHg, or 80 within 10 mmHg and 95 within 15 mmHg. The AGILIS monitor had 78 measurements within 5 mmHg, 91 within 10 mmHg and 96 within 15 mmHg for SBP; and 70 measurements within 5 mmHg, 92 within 10 mmHg and 96 within 15 mmHg for DBP. It passed the criteria for SBP and DBP. For phase 2.2 using the AGILIS monitor, 27 subjects had at least two of the differences within 5 mmHg and none had any differences within 5 mmHg for SBP; 27 subjects had at least two of the differences within 5 mmHg and three subjects had no differences within 5 mmHg for DBP. The results are in concordance with the requested criteria for phase 2.2. Conclusion Thus the AGILIS device fulfils the validation criteria of the international protocol. © 2005 Lippincott Williams & Wilkins.
引用
收藏
页码:97 / 101
页数:5
相关论文
共 50 条
  • [31] Erratum to: Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010
    Beate Beime
    Cornelia Deutsch
    Ralf Krüger
    Andreas Wolf
    Peter Müller
    Gertrud Hammel
    Peter Bramlage
    European Journal of Pediatrics, 2017, 176 : 835 - 836
  • [32] Validation of the Oregon Scientific BPW810 blood pressure monitor according to the European Society of Hypertension Protocol
    Li, Li
    Hui, Wang
    Lan, Wu Jin
    BLOOD PRESSURE MONITORING, 2006, 11 (06) : 343 - 347
  • [33] European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults (vol 15, pg 23, 2010)
    O'Brien, E.
    Atkins, N.
    Stergiou, G.
    Karpettas, N.
    Parati, G.
    Asmar, R.
    BLOOD PRESSURE MONITORING, 2010, 15 (03) : 171 - 172
  • [34] Validation of the SCIAN LD-735 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Kang, Yuan-Yuan
    Chen, Qi
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2016, 21 (04) : 255 - 258
  • [35] Validation of the AVITA BPM17 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Kang, Yuan-Yuan
    Chen, Qi
    Liu, Chang-Yuan
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2017, 22 (04) : 230 - 233
  • [36] Validation of the Rossmax S150 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Zeng, Wei-Fang
    Huang, Qi-Fang
    Sheng, Chang-Sheng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2012, 17 (05) : 204 - 206
  • [37] Validation of the Somnotouch-NIBP noninvasive continuous blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010
    Bilo, Grzegorz
    Zorzi, Cristina
    Munera, Juan E. Ochoa
    Torlasco, Camilla
    Giuli, Valentina
    Parati, Gianfranco
    BLOOD PRESSURE MONITORING, 2015, 20 (05) : 291 - 294
  • [38] Validation of the Health & Life BSP-4007 blood pressure monitor in adult Chinese according to the European Society of Hypertension International Protocol
    Chen, H. E.
    Cui, Y.
    Sheng, C. S.
    Li, L. H.
    Li, Y.
    Wang, J. G.
    JOURNAL OF HYPERTENSION, 2008, 26 : S496 - S497
  • [39] Improving the accuracy of blood pressure measurement: the influence of the European Society of Hypertension International Protocol (ESH-IP) for the validation of blood pressure measuring devices and future perspectives
    Stergiou, George S.
    Asmar, Roland
    Myers, Martin
    Palatini, Paolo
    Parati, Gianfranco
    Shennan, Andrew
    Wang, Jiguang
    O'Brien, Eoin
    JOURNAL OF HYPERTENSION, 2018, 36 (03) : 479 - 487
  • [40] Validation of the AVITA BPM15S wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Kang, Yuan-Yuan
    Zeng, Wei-Fang
    Zhang, Lu
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2014, 19 (03) : 183 - 186