Application of argon as the internal standard method in glow discharge mass spectrometry

被引:3
作者
Xing, Yu [1 ]
Xiaojia, Li [1 ]
Haizhou, Wang [1 ]
机构
[1] Cent Iron & Steel Res Inst, Natl Anal Ctr Iron & Steel, Beijing 100081, Peoples R China
关键词
argon internal standard; glow discharge mass spectrometry;
D O I
10.1255/ejms.854
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
Using Ar-78(2)+ as the internal standard (argon internal standard) in glow discharge mass spectrometry (GDMS) was investigated in detail. After comparing ion intensities and interferences, i.e. of argon ions, bi-atom argon ions and tri-atom argon ions, the Ar-78(2)+ was selected as the internal standard in the analysis. Mass spectral behavior of the argon internal standard affected by glow discharge current and voltage were studied. The ion intensity relationship between the argon internal standard and the matrix internal standard showed that the argon internal standard and the matrix internal standard have the same corrective effect on sample analysis. The experiment proved that the effects of the difference in analysis samples and the fluctuation of discharge conditions for analytical results were efficiently reduced if the argon internal standard was used. Moreover, the argon internal standard is similar to the matrix internal standard in correcting analytical results and both give satisfactory results. Elemental content in samples can be accurately determined by using the argon internal standard when the matrix content is unknown with good results.