Design, fabrication, and in vitro testing of novel three-dimensionally printed tympanic membrane grafts.

被引:58
作者
Kozin, Elliott D. [1 ,2 ,3 ]
Black, Nicole L. [4 ,5 ]
Cheng, Jeffrey T. [1 ,2 ,3 ]
Cotler, Max J. [4 ,5 ]
McKenna, Michael J. [1 ,2 ,3 ]
Lee, Daniel J. [1 ,2 ,3 ]
Lewis, Jennifer A. [4 ,5 ]
Rosowski, John J. [1 ,2 ,3 ]
Remenschneider, Aaron K. [1 ,2 ,3 ]
机构
[1] Massachusetts Eye & Ear Infirm, Dept Otolaryngol, 243 Charles St, Boston, MA 02118 USA
[2] Massachusetts Eye & Ear Infirm, Eaton Peabody Labs, 243 Charles St, Boston, MA 02118 USA
[3] Harvard Med Sch, Dept Otol & Laryngol, Boston, MA USA
[4] Harvard Univ, Harvard John A Paulson Sch Engn & Appl Sci, Cambridge, MA 02138 USA
[5] Harvard Univ, Wyss Inst Biol Inspired Engn, Cambridge, MA 02138 USA
关键词
Tympanic membrane; 3D printing; Holography; Tympanoplasty; Biomimetic; Tissue engineering; TIME-AVERAGED HOLOGRAPHY; CHRONIC EAR SURGERY; STROBOSCOPIC HOLOGRAPHY; BIOMEDICAL APPLICATIONS; THICKNESS DISTRIBUTION; HYDROGEL SCAFFOLDS; SOUND-TRANSMISSION; FIBER ARRANGEMENT; DRUG-DELIVERY; CARTILAGE;
D O I
10.1016/j.heares.2016.03.005
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
The tympanic membrane (TM) is an exquisite structure that captures and transmits sound from the environment to the ossicular chain of the middle ear. The creation of TM grafts by multi-material three-dimensional (3D) printing may overcome limitations of current graft materials, e.g. temporalis muscle fascia, used for surgical reconstruction of the TM. TM graft scaffolds with either 8 or 16 circumferential and radial filament arrangements were fabricated by 3D printing of polydimethylsiloxane (PDMS), flexpolyactic acid (PIA) and polycaprolactone (PCL) materials followed by uniform infilling with a fibrin collagen composite hydrogel. Digital opto-electronic holography (DOEH) and laser Doppler vibrometry (LDV) were used to measure acoustic properties including surface motions and velocity of TM grafts in response to sound. Mechanical properties were determined using dynamic mechanical analysis (DMA). Results were compared to fresh cadaveric human TMs and cadaveric temporalis fascia. Similar to the human TM, TM grafts exhibit simple surface motion patterns at lower frequencies (400 Hz), with a limited number of displacement maxima. At higher frequencies (>1000 Hz), their displacement patterns are highly organized with multiple areas of maximal displacement separated by regions of minimal displacement. By contrast, temporalis fascia exhibited asymmetric and less regular holographic patterns. Velocity across frequency sweeps (0.2-10 kHz) measured by LDV demonstrated consistent results for 3D printed grafts, while velocity for human fascia varied greatly between specimens. TM composite grafts of different scaffold print materials and varied filament count (8 or 16) displayed minimal, but measurable differences in DOEH and LDV at tested frequencies. TM graft mechanical load increased with higher filament count and is resilient over time, which differs from temporalis fascia, which loses over 70% of its load bearing properties during mechanical testing. This study demonstrates the design, fabrication and preliminary in vitro acoustic and mechanical evaluation of 3D printed TM grafts. Data illustrate the feasibility of creating TM grafts with acoustic properties that reflect sound induced motion patterns of the human TM; furthermore, 3D printed grafts have mechanical properties that demonstrate increased resistance to deformation compared to temporalis fascia. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:191 / 203
页数:13
相关论文
共 69 条
  • [1] Motion of the tympanic membrane after cartilage tympanoplasty determined by stroboscopic holography
    Aarnisalo, Antti A.
    Cheng, Jeffrey T.
    Ravicz, Michael E.
    Furlong, Cosme
    Merchant, Saumil N.
    Rosowski, John J.
    [J]. HEARING RESEARCH, 2010, 263 (1-2) : 78 - 84
  • [2] Middle Ear Mechanics of Cartilage Tympanoplasty Evaluated by Laser Holography and Vibrometry
    Aarnisalo, Antti A.
    Cheng, Jeffrey T.
    Ravicz, Michael E.
    Hulli, Nesim
    Harrington, Ellery J.
    Hernandez-Montes, Maria S.
    Furlong, Cosme
    Merchant, Saumil N.
    Rosowski, John J.
    [J]. OTOLOGY & NEUROTOLOGY, 2009, 30 (08) : 1209 - 1214
  • [3] Direct-Write Assembly of 3D Hydrogel Scaffolds for Guided Cell Growth
    Barry, Robert A., III
    Shepherd, Robert F.
    Hanson, Jennifer N.
    Nuzzo, Ralph G.
    Wiltzius, Pierre
    Lewis, Jennifer A.
    [J]. ADVANCED MATERIALS, 2009, 21 (23) : 2407 - +
  • [4] BOEDTS D, 1990, AM J OTOL, V11, P274
  • [5] Efficacy of Cartilage Palisade Tympanoplasty: Randomized Controlled Trial
    Cabra, Jesus
    Monux, Alfonso
    [J]. OTOLOGY & NEUROTOLOGY, 2010, 31 (04) : 589 - 595
  • [6] The Design and Engineering of the Memory Shape Breast Implant
    Calobrace, M. Bradley
    [J]. PLASTIC AND RECONSTRUCTIVE SURGERY, 2014, 134 (03) : 10S - 15S
  • [7] Wave motion on the surface of the human tympanic membrane: Holographic measurement and modeling analysis
    Cheng, Jeffrey Tao
    Hamade, Mohamad
    Merchant, Saumil N.
    Rosowski, John J.
    Harrington, Ellery
    Furlong, Cosme
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2013, 133 (02) : 918 - 937
  • [8] Motion of the surface of the human tympanic membrane measured with stroboscopic holography
    Cheng, Jeffrey Tao
    Aarnisalo, Antti A.
    Harrington, Ellery
    Hernandez-Montes, Maria del Socorro
    Furlong, Cosme
    Merchant, Saumil N.
    Rosowski, John J.
    [J]. HEARING RESEARCH, 2010, 263 (1-2) : 66 - 77
  • [9] ARTICULAR-CARTILAGE REPAIR USING ALLOGENEIC PERICHONDROCYTE-SEEDED BIODEGRADABLE POROUS POLYLACTIC ACID (PLA) - A TISSUE-ENGINEERING STUDY
    CHU, CR
    COUTTS, RD
    YOSHIOKA, M
    HARWOOD, FL
    MONOSOV, AZ
    AMIEL, D
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1995, 29 (09): : 1147 - 1154
  • [10] Nonlinear material behaviour of spider silk yields robust webs
    Cranford, Steven W.
    Tarakanova, Anna
    Pugno, Nicola M.
    Buehler, Markus J.
    [J]. NATURE, 2012, 482 (7383) : 72 - U91