Modification of the Surgical Strategy for the Dissection of the Recurrent Laryngeal Nerve Using Continuous Intraoperative Nerve Monitoring

被引:24
作者
Arteaga, Andres Marin [1 ,2 ]
Peloni, Giuseppe [2 ,3 ]
Leuchter, Igor [4 ]
Bedat, Benoit [2 ]
Karenovics, Wolfram [2 ]
Triponez, Frederic [2 ]
Sadowski, Samira Mercedes [2 ]
机构
[1] Univ Hosp Geneva, Gen Surg, Geneva, Switzerland
[2] Univ Hosp Geneva, Thorac & Endocrine Surg, Rue Gabrielle Perret Gentil 4, CH-1211 Geneva 14, Switzerland
[3] Osped Reg Mendrisio, Via Turconi 23, CH-6850 Mendrisio, Switzerland
[4] Univ Hosp Geneva, Ear Nose Throat Div, Geneva, Switzerland
关键词
VOCAL FOLD FUNCTION; THYROID-SURGERY; BILATERAL THYROIDECTOMY; NEUROMONITORING SIGNAL; OPERATIVE STRATEGY; PORCINE MODEL; EMG CHANGES; COMPLICATIONS; MULTICENTER; PARALYSIS;
D O I
10.1007/s00268-017-4277-z
中图分类号
R61 [外科手术学];
学科分类号
摘要
The aim of this study was to describe first experiences and changes in management using continuous intraoperative neuromonitoring (C-IONM) in thyroid and parathyroid surgery. Retrospective analysis of patients who underwent surgery with C-IONM since 2012. Surgical maneuvers were modified when electrophysiologic events occurred. Patients with persistent loss of signal (LOS) underwent postoperative laryngoscopy. One hundred and one patients (of 1586 neck surgeries) were included and 19 had events: In 13 these were temporary (resolved before end of surgery) and led to intraoperative modifications in surgical approach; in all cases traction was released, and in 8, recurrent laryngeal nerve (RLN) approach was changed [superior approach (2), inferior approach (2), both (4)]. Six patients had persistent LOS (5.9%, present at end of procedure), with RLN palsy (RLNP) on postoperative day 1: In three, LOS occurred at electrode placement on the vagus nerve, leading to distal placement of the electrode allowing ipsilateral dissection under continuous monitoring; all three had complete recovery at 6 months. In the three other patients, LOS occurred on the RLN: one probable thermal, one traction lesion and one accidental section of the anterior RLN branch. The RLN recovered within 6 months in two patients, and in the third, RLNP persisted after 6 months (1/101 = 1%). C-IONM provides real-time evaluation of the RLN function, allowing for adaptation of surgical maneuvers to prevent RLNP. It seems particularly useful in difficult cases like redo neck surgery, invasive thyroid cancer and intrathoracic or large goiter. Care should be given at electrode placement on the vagus nerve.
引用
收藏
页码:444 / 450
页数:7
相关论文
共 50 条
  • [31] Intraoperative neuromonitoring of the recurrent laryngeal nerve results and learning curve
    Jonas, J.
    Baehr, R.
    ZENTRALBLATT FUR CHIRURGIE, 2006, 131 (06): : 443 - 448
  • [32] Interpretation of intraoperative recurrent laryngeal nerve monitoring signals: The importance of a correct standardization
    Calo, Pietro Giorgio
    Medas, Fabio
    Gordini, Luca
    Podda, Francesco
    Erdas, Enrico
    Pisano, Giuseppe
    Nicolosi, Angelo
    INTERNATIONAL JOURNAL OF SURGERY, 2016, 28 : S54 - S58
  • [33] Prognostic value of intraoperative neural monitoring of the recurrent laryngeal nerve in thyroid surgery
    Stopa, Malgorzata
    Barczynski, Marcin
    LANGENBECKS ARCHIVES OF SURGERY, 2017, 402 (06) : 957 - 964
  • [34] Prognostic value of intraoperative neural monitoring of the recurrent laryngeal nerve in thyroid surgery
    Małgorzata Stopa
    Marcin Barczyński
    Langenbeck's Archives of Surgery, 2017, 402 : 957 - 964
  • [35] Middle fixation of electromyographic endotracheal tube for intraoperative recurrent laryngeal nerve monitoring
    Cherng, Chen-Hwan
    Huang, Yi-Hsuan
    Shih, Ming-Lang
    JOURNAL OF CLINICAL ANESTHESIA, 2014, 26 (03) : 252 - 253
  • [36] Intraoperative Laryngeal Nerve Monitoring During Thyroidectomy
    Donnellan, Kimberly A.
    Pitman, Karen T.
    Cannon, C. Ron
    Replogle, William H.
    Simmons, Jon D.
    ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2009, 135 (12) : 1196 - 1198
  • [37] Intraoperative nerve monitoring is associated with a lower risk of recurrent laryngeal nerve injury: A national analysis of 17,610 patients
    Kim, Jina
    Graves, Claire E.
    Jin, Chengshi
    Duh, Quan-Yang
    Gosnell, Jessica E.
    Shen, Wen T.
    Suh, Insoo
    Sosa, Julie A.
    Roman, Sanziana A.
    AMERICAN JOURNAL OF SURGERY, 2021, 221 (02) : 472 - 477
  • [38] Impact of routine recurrent laryngeal nerve monitoring in prone esophagectomy with mediastinal lymph node dissection
    Hikage, Makoto
    Kamei, Takashi
    Nakano, Toru
    Abe, Shigeo
    Katsura, Kazunori
    Taniyama, Yusuke
    Sakurai, Tadashi
    Teshima, Jin
    Ito, Soichi
    Niizuma, Nobuchika
    Okamoto, Hiroshi
    Fukutomi, Toshiaki
    Yamada, Masato
    Maruyama, Shota
    Ohuchi, Noriaki
    SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2017, 31 (07): : 2986 - 2996
  • [39] Cost-effectiveness of intraoperative nerve monitoring in avoidance of bilateral recurrent laryngeal nerve injury in patients undergoing total thyroidectomy
    Al-Qurayshi, Z.
    Kandil, E.
    Randolph, G. W.
    BRITISH JOURNAL OF SURGERY, 2017, 104 (11) : 1523 - 1531
  • [40] Evaluation of recurrent laryngeal nerve monitoring in thyroid surgery
    Sari, Serkan
    Erbil, Yesim
    Sumer, Aziz
    Agcaoglu, Orhan
    Bayraktar, Adem
    Issever, Halim
    Ozarmagan, Selcuk
    INTERNATIONAL JOURNAL OF SURGERY, 2010, 8 (06) : 474 - 478