Real ear measurement methods for open fit hearing aids: Modified pressure concurrent equalization (MPCE) versus modified pressure stored equalization (MPSE)

被引:4
作者
Aazh, Hashir [1 ]
Moore, Brian C. J. [2 ]
Prasher, Deepak [1 ]
机构
[1] Royal Surrey Cty Hosp NHS Fdn Trust, Dept Audiol, Guildford GU2 7XX, Surrey, England
[2] Univ Cambridge, Dept Expt Psychol, Cambridge CB2 1TN, England
关键词
Open fit; hearing aid; real ear measurement; real ear insertion gain; GAIN; VERIFICATION;
D O I
10.3109/14992027.2011.609182
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objective: The aim of this study was to assess differences between real ear insertion gains (REIG) measured with the modified pressure concurrent equalization (MPCE) and modified pressure stored equalization (MPSE) methods for open fittings in a typical audiology patient population. Design: REIGs were compared for the two methods using a warble tone sweep at 65 dB SPL. The differences between the two methods at 0.25, 0.5, 1, 2, 3, 4 and 6 kHz were recorded. Study sample: Eighty-three ears of a consecutive sample of 48 candidates for open-fit hearing aids were included. Results: The mean difference between MPSE and MPCE REIGs was less than 1 dB at all frequencies. Analysis of variance showed that the main effect of method was not significant, and there was no significant interaction between method and frequency. Conclusions: The results for the MPSE and MPCE methods did not differ significantly for the patients with mild-to-moderate hearing losses tested here, for whom REIGs were generally less than 20 dB. Further research is needed to identify the REIG values at which the differences between MPCE and MPSE methods become clinically significant.
引用
收藏
页码:103 / 107
页数:5
相关论文
共 20 条
[1]   Manufacturer predicted and measured REAR values in adult hearing aid fitting: Accuracy and clinical usefulness [J].
Aarts, NL ;
Caffee, CS .
INTERNATIONAL JOURNAL OF AUDIOLOGY, 2005, 44 (05) :293-301
[2]  
Aazh H., 2011, AM J AUDIOL
[3]  
[Anonymous], 2001, HEARING AIDS
[4]  
*ANSI, 1997, S3461997 ANSI
[5]  
ANSI, 2003, S3222003 ANSI S3222003 ANSI
[6]  
British Society of Audiology & British Academy of Audiology, 2007, GUID US REAL EAR MEA
[7]   HEARING-AID GAIN AND FREQUENCY-RESPONSE REQUIREMENTS FOR THE SEVERELY PROFOUNDLY HEARING-IMPAIRED [J].
BYRNE, D ;
PARKINSON, A ;
NEWALL, P .
EAR AND HEARING, 1990, 11 (01) :40-49
[8]   Speech recognition of hearing-impaired listeners: Predictions from audibility and the limited role of high-frequency amplification [J].
Ching, TYC ;
Dillon, H ;
Byrne, D .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1998, 103 (02) :1128-1140
[9]  
Dalsgaard S.C., 1976, J AUD TECH, V15, P170