Evaluation of Pulse Cooximetry in Patients Undergoing Abdominal or Pelvic Surgery

被引:56
作者
Applegate, Richard L., II [1 ]
Barr, Steven J. [1 ]
Collier, Carl E. [1 ]
Rook, James L. [1 ]
Mangus, Dustin B. [1 ]
Allard, Martin W. [1 ]
机构
[1] Loma Linda Univ, Med Ctr, Dept Anesthesiol, Sch Med, Loma Linda, CA 92354 USA
关键词
OF-THE-LITERATURE; NONCARDIAC SURGERY; BLOOD-TRANSFUSION; PREOPERATIVE ANEMIA; OLDER PATIENTS; SINGLE-CENTER; VOLUME; ASSOCIATION; MORTALITY; RISK;
D O I
10.1097/ALN.0b013e31823d774f
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: Intraoperative transfusion decisions generally are guided by blood loss estimation and periodic invasive hemoglobin measurement. Continuous hemoglobin measurement by pulse cooximetry (pulse hemoglobin; Rainbow (R) SET Pulse CO-Oximeter, Masimo Corporation, Irvine, CA) has good agreement with laboratory hemoglobin in healthy volunteers and could aid transfusion decision-making. Because intraoperative physiology may alter performance of this device, this study investigated pulse hemoglobin during surgery. Methods: Ninety-one adult patients undergoing abdominal or pelvic surgery in which large blood loss was likely were studied. Time-matched pulse hemoglobin measurements were recorded for each intraoperative arterial hemoglobin measurement obtained. Agreement between measurements was assessed by average difference (mean +/- SD, g/dl), linear regression, and multiple measures Bland-Altman analysis. Results: The average difference between 360 time-matched measurements (bias) was 0.50 +/- 1.44 g/dl, with wider limits of agreement (-2.3 to 3.3 g/dl) than reported in healthy volunteers. The average difference between 269 paired sequential pulse and arterial hemoglobin changes was 0.10 +/- 1.11 g/dl, with half between -0.6 and 0.7 g/dl of each other. The bias was larger in patients with blood loss of more than 1,000 ml; hemoglobin less than 9.0 g/dl; any intraoperative transfusion; or intraoperative decrease in arterial hemoglobin at the time of sampling >= 2 g/dl (all P < 0.001). The range of bias was narrower at deeper anesthesia (P < 0.001). Conclusions: Evaluation of the sensor and software version tested suggests that although pulse cooximetry may perform well in ambulatory subjects, in patients undergoing surgery in which large blood loss is likely, an invasive measurement should be used in transfusion decision-making.
引用
收藏
页码:65 / 72
页数:8
相关论文
共 27 条
  • [1] Amer Soc Anesthesiologists Task Fo, 2006, ANESTHESIOLOGY, V105, P198
  • [2] [Anonymous], 2004, PERF CHAR ABL800 REF, P28
  • [3] Acute Surgical Anemia Influences the Cardioprotective Effects of β-Blockade A Single-center, Propensity-matched Cohort Study
    Beattie, W. Scott
    Wijeysundera, Duminda N.
    Karkouti, Keyvan
    McCluskey, Stuart
    Tait, Gordon
    Mitsakakis, Nicholas
    Hare, Gregory M. T.
    [J]. ANESTHESIOLOGY, 2010, 112 (01) : 25 - 33
  • [4] Risk Associated with Preoperative Anemia in Noncardiac Surgery A Single-center Cohort Study
    Beattie, W. Scott
    Karkouti, Keyvan
    Wijeysundera, Duminda N.
    Tait, Gordon
    [J]. ANESTHESIOLOGY, 2009, 110 (03) : 574 - 581
  • [5] A Comparison of Stroke Volume Variation Measured by Vigileo™/FloTraC™ System and Aortic Doppler Echocardiography
    Biais, Matthieu
    Nouette-Gaulain, Karine
    Roullet, Stephanie
    Quinart, Alice
    Revel, Philippe
    Sztark, Francois
    [J]. ANESTHESIA AND ANALGESIA, 2009, 109 (02) : 466 - 469
  • [6] Preoperative anemia in patients undergoing coronary artery bypass grafting predicts acute kidney injury
    De Santo, Luca
    Romano, Gianpaolo
    Della Corte, Alessandro
    de Simone, Vincenzo
    Grimaldi, Francesco
    Cotrufo, Maurizio
    de Feo, Marisa
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2009, 138 (04) : 965 - 970
  • [7] A comparison of stroke volume variation measured by the LiDCOplus and FloTrac-Vigileo system
    de Wilde, R. B. P.
    Geerts, B. F.
    van den Berg, P. C. M.
    Jansen, J. R. C.
    [J]. ANAESTHESIA, 2009, 64 (09) : 1004 - 1009
  • [8] Automated pulse pressure and stroke volume variations from radial artery: evaluation during major abdominal surgery
    Derichard, A.
    Robin, E.
    Tavernier, B.
    Costecalde, M.
    Fleyfel, M.
    Onimus, J.
    Lebuffe, G.
    Chambon, J. -P.
    Vallet, B.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2009, 103 (05) : 678 - 684
  • [9] Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients
    Frasca, Denis
    Dahyot-Fizelier, Claire
    Catherine, Karen
    Levrat, Quentin
    Debaene, Bertrand
    Mimoz, Olivier
    [J]. CRITICAL CARE MEDICINE, 2011, 39 (10) : 2277 - 2282
  • [10] Association between Intraoperative Blood Transfusion and Mortality and Morbidity in Patients Undergoing Noncardiac Surgery
    Glance, Laurent G.
    Dick, Andrew W.
    Mukamel, Dana B.
    Fleming, Fergal J.
    Zollo, Raymond A.
    Wissler, Richard
    Salloum, Rabih
    Meredith, U. Wayne
    Osler, Turner M.
    [J]. ANESTHESIOLOGY, 2011, 114 (02) : 283 - 292