Watchman FLX vs. Watchman 2.5 in a Dual-Center Left Atrial Appendage Closure Cohort: the WATCH-DUAL study

被引:29
|
作者
Galea, Roberto [1 ]
Mahmoudi, Khalil [2 ]
Grani, Christoph [1 ]
Elhadad, Simon [2 ,3 ]
Huber, Adrian T. [1 ]
Heg, Dik [4 ]
Siontis, George C. M. [1 ]
Brugger, Nicolas [1 ]
Sebag, Frederic [2 ]
Windecker, Stephan [1 ]
Valgimigli, Marco [1 ,5 ]
Landolff, Quentin [2 ]
Roten, Laurent [1 ]
Amabile, Nicolas [2 ]
Raber, Lorenz [1 ]
机构
[1] Univ Bern, Bern Univ Hosp, Dept Cardiol, Inselspital, CH-3010 Bern, Switzerland
[2] Inst Mutualiste Montsouris, Dept Cardiol, Paris, France
[3] Cardiol Dept, Jossigny, France
[4] Univ Bern, Clin Trials Unit, Bern, Switzerland
[5] Inst & Univ Svizzera Italiana USI, Cardioctr Ticino, Lugano, Switzerland
来源
EUROPACE | 2022年 / 24卷 / 09期
关键词
Left atrial appendage closure; Watchman; 2; 5; Watchman FLX; Cardiac computed tomography angiography; Device-related thrombus; Device permeability; OCCLUSION; ECHOCARDIOGRAPHY; FIBRILLATION; SAFETY;
D O I
10.1093/europace/euac021
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims No studies have compared Watchman 2.5 (W2.5) with Watchman FLX (FLX) devices to date. We aimed at comparing the FLX with W2.5 devices with respect to clinical outcomes, left atrial appendage (LAA) sealing properties and device-related thrombus (DRT). Methods and results All consecutive left atrial appendage closure (LAAC) procedures performed at two European centres between November 2017 and February 2021 were included. Procedure-related complications and net adverse cardiovascular events (NACE) at 6 months after LAAC were recorded. At 45-day computed tomography (CT) follow-up, intra- (IDL) and peri- (PDL) device leak, residual patent neck area (RPNA), and DRT were assessed by a Corelab. Out of 144 LAAC consecutive procedures, 71 and 73 interventions were performed using W2.5 and FLX devices, respectively. There were no differences in terms of procedure-related complications (4.2% vs. 2.7%, P = 0.626). At 45-day CT, the FLX was associated with lower frequency of IDL [21.3% vs. 40.0%; P = 0.032; odds ratio (OR): 0.375; 95% confidence interval (CI): 0.160-0.876; P = 0.024], similar rate of PDL (29.5% vs. 42.0%; P = 0.170), and smaller RPNA [6 (0-36) vs. 40 (6-115) mm(2); P = 0.001; OR: 0.240; 95% CI: 0.100-0.577; P = 0.001] compared with the W2.5 group. At 45 days, rate of DRT as detected by CT and/or transoesophageal echocardiography (TOE), was higher with W2.5 (6.0% vs. 0%, P = 0.045). At 6-month follow-up, NACE did not differ between groups. Conclusions In this cohort of consecutive LAACs, FLX as compared to W2.5, was associated with similar procedure-related complications and 6-month NACE, but with improved LAA neck coverage, and lower IDL and DRT.
引用
收藏
页码:1441 / 1450
页数:10
相关论文
共 49 条
  • [11] Intracardiac vs Transesophageal Echocardiography for Left Atrial Appendage Occlusion With Watchman FLX in the US
    Ferro, Enrico G.
    Alkhouli, Mohamad
    Nair, Devi G.
    Kapadia, Samir R.
    Hsu, Jonathan C.
    Gibson, Douglas N.
    V. Freeman, James
    Price, Matthew J.
    Roy, Kristine
    Allocco, Dominic J.
    Yeh, Robert W.
    Piccini, Jonathan P.
    JACC-CLINICAL ELECTROPHYSIOLOGY, 2023, 9 (12) : 2587 - 2599
  • [12] Novel computed tomography angiography-based sizing methodology for WATCHMAN FLX device in left atrial appendage closure
    Dallan, Luis Augusto Palma
    Arruda, Mauricio
    Yoon, Sung-Han
    Rana, Mohammad Atif
    Mogalapalli, Akhil
    Carneiro, Herman A.
    Reed, Joseph
    Rashid, Imran
    Rajagopalan, Sanjay
    Filby, Steven J.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2022, 33 (08) : 1781 - 1787
  • [13] Comparison in Short-Term Safety and Efficacy between New-Generation WATCHMAN FLX and Conventional WATCHMAN 2.5 for Percutaneous Left Atrial Appendage Closure
    Fukuda, Nobuyuki
    Imamura, Teruhiko
    Tanaka, Shuhei
    Kataoka, Naoya
    Ushijima, Ryuichi
    Ueno, Hiroshi
    Kinugawa, Koichiro
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (06)
  • [14] Intracardiac echocardiography-guided implantation of the Watchman FLX left atrial appendage closure device
    Turagam, Mohit K.
    Neuzil, Petr
    Petru, Jan
    Hala, Pavel
    Mraz, Tomas
    Baroch, Jiri
    Lekesova, Veronika
    Prokopova, Milena
    Dukkipati, Srinivas R.
    Reddy, Vivek Y.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2021, 32 (03) : 717 - 725
  • [15] Transition from WATCHMAN generation-2.5 device to WATCHMAN FLX device for percutaneous left atrial appendage closure: Incidence and predictors of device-related thrombosis and short- to mid-term outcomes
    Chatani, Ryuki
    Kubo, Shunsuke
    Tasaka, Hiroshi
    Nishiura, Naoki
    Mushiake, Kazunori
    Ono, Sachiyo
    Maruo, Takeshi
    Kadota, Kazushige
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2024, 104 (02) : 318 - 329
  • [16] Canadian Multi-center Experience With WATCHMAN for Percutaneous Left Atrial Appendage Closure
    Saw, Jacqueline
    Fahmy, Peter
    Azzalini, Lorenzo
    Marquis, Jean-Francois
    Hibbert, Benjamin
    Morillo, Carlos
    Carrizo, Aldo
    Ibrahim, Reda
    CIRCULATION, 2015, 132
  • [17] Gender differences in outcomes after left atrial appendage closure with Watchman FLX device: insights from the Italian-FLX registry
    Bonanni, Michela
    Frazzetto, Marco
    Nardone, Annalisa
    Meucci, Francesco
    Musto, Carmine
    Quaranta, Gaetano
    Sacca, Salvatore
    Bedogni, Francesco
    Maffeo, Diego
    Ugo, Fabrizio
    Guarracini, Fabrizio
    Bocuzzi, Giacomo
    Durante, Alessandro
    Granatelli, Antonino
    Tumminello, Gabriele
    Eusebio, Geppina
    Grasso, Carmelo
    De Marco, Federico
    Cortese, Bernardo
    Mariani, Massimiliano
    Berti, Sergio
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2024, 11
  • [18] Periprocedural Intracardiac Echocardiography for Left Atrial Appendage Closure A Dual-Center Experience
    Berti, Sergio
    Paradossi, Umberto
    Meucci, Francesco
    Trianni, Giuseppe
    Tzikas, Apostolos
    Rezzaghi, Marco
    Stolkova, Miroslava
    Palmieri, Cataldo
    Mori, Fabio
    Santoro, Gennaro
    JACC-CARDIOVASCULAR INTERVENTIONS, 2014, 7 (09) : 1036 - 1044
  • [19] Left Atrial Appendage Closure With the Watchman Device in Patients With a Contraindication for Oral Anticoagulation The ASAP Study (ASA Plavix Feasibility Study With Watchman Left Atrial Appendage Closure Technology)
    Reddy, Vivek Y.
    Moebius-Winkler, Sven
    Miller, Marc A.
    Neuzil, Petr
    Schuler, Gerhard
    Wiebe, Jens
    Sick, Peter
    Sievert, Horst
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 61 (25) : 2551 - 2556
  • [20] Comparison between Amulet and Watchman left atrial appendage closure devices: A real-world, single center experience
    Saad, Mohammed
    Risha, Osama
    Sano, Makoto
    Fink, Thomas
    Heeger, Christian-Hendrik
    Vogler, Julia
    Sciacca, Vanessa
    Eitel, Charlotte
    Stiermaier, Thomas
    Joost, Alexander
    Keelani, Ahmad
    Fuernau, Georg
    Meyer-Saraei, Roza
    Kuck, Karl-Heinz
    Eitel, Ingo
    Tilz, Roland Richard
    IJC HEART & VASCULATURE, 2021, 37