An inverse geometry CT system with stationary source arrays

被引:3
|
作者
Hsieh, Scott S. [1 ,2 ]
Heanue, Joseph A. [3 ]
Funk, Tobias [3 ]
Hinshaw, Waldo S. [2 ,3 ]
Pelc, Norbert J. [1 ,2 ]
机构
[1] Stanford Univ, Dept Elect Engn, Stanford, CA 94305 USA
[2] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[3] Triple Ring Technol Inc, Newark, NY 94560 USA
来源
MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING | 2011年 / 7961卷
关键词
computed tomography (CT); inverse-geometry; IGCT; stationary source;
D O I
10.1117/12.878801
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Traditional CT systems face a tradeoff between temporal resolution, volumetric coverage and cone beam artifacts and also have limited ability to customize the distribution of incident x-rays to the imaging task. Inverse geometry CT (IGCT) can overcome some of these limitations by placing a small detector opposite a large, rotating scanned source array. It is difficult to quickly rotate this source array to achieve rapid imaging, so we propose using stationary source arrays instead and investigate the feasibility of such a system. We anticipate that distinct source arrays will need to be physically separated, creating gaps in the sinogram. Symmetry can be used to fill the missing rays except those connecting gaps. With three source arrays, a large triangular field of view emerges. As the small detector orbits the patient, each source spot must be energized at multiple specifically designed times to ensure adequate sampling. A timing scheme is proposed that avoids timing clashes, efficiently uses the detector, and allows for simple collimation. The two-dimensional MTF, noise characteristics, and artifact levels are all found to be comparable to parallel-beam systems. A complete, 100 millisecond volumetric scan may be feasible.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Monoplane Stereoscopic Imaging Method for Inverse Geometry X-ray Fluoroscopy
    Tomkowiak, Michael T.
    Van Lysel, Michael S.
    Speidel, Michael A.
    MEDICAL IMAGING 2013: IMAGE PROCESSING, 2013, 8669
  • [22] A geometric calibration method for inverse geometry computed tomography using P-matrices
    Slagowski, Jordan M.
    Dunkerley, David A. P.
    Hatt, Charles R.
    Speidel, Michael A.
    MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
  • [23] Localization of cardiac volume and patient features in inverse geometry x-ray fluoroscopy
    Speidel, Michael A.
    Slagowski, Jordan M.
    Dunkerley, David A. P.
    Wagner, Martin
    Funk, Tobias
    Raval, Amish N.
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [24] Stationary table CT dosimetry and anomalous scanner-reported values of CTDIvol
    Dixon, Robert L.
    Boone, John M.
    MEDICAL PHYSICS, 2014, 41 (01)
  • [25] Overview of Multi-source CT Systems and Methods
    Zhao, Jun
    Lu, Yang
    Zhuang, Tiange
    Wang, Ge
    DEVELOPMENTS IN X-RAY TOMOGRAPHY VII, 2010, 7804
  • [26] DUAL SOURCE CT: CARDIO-PULMONARY APPLICATIONS
    Hamoir, X.
    Salovic, D.
    Bouziane, T.
    Kirsch, J.
    JBR-BTR, 2007, 90 (02): : 77 - 79
  • [27] Can CT angiography reconstructed from CT perfusion source data on a 320-section volume CT scanner replace conventional CT angiography for the evaluation of intracranial arteries?
    Kidoh, Masafumi
    Hirai, Toshinori
    Oda, Seitaro
    Utsunomiya, Daisuke
    Kawano, Takayuki
    Yano, Shigetoshi
    Nakamura, Hideo
    Makino, Keishi
    Iryo, Yasuhiko
    Azuma, Minako
    Hayashida, Eri
    Nakaura, Takeshi
    Yamashita, Yasuyuki
    JAPANESE JOURNAL OF RADIOLOGY, 2015, 33 (06) : 353 - 359
  • [28] Can CT angiography reconstructed from CT perfusion source data on a 320-section volume CT scanner replace conventional CT angiography for the evaluation of intracranial arteries?
    Masafumi Kidoh
    Toshinori Hirai
    Seitaro Oda
    Daisuke Utsunomiya
    Takayuki Kawano
    Shigetoshi Yano
    Hideo Nakamura
    Keishi Makino
    Yasuhiko Iryo
    Minako Azuma
    Eri Hayashida
    Takeshi Nakaura
    Yasuyuki Yamashita
    Japanese Journal of Radiology, 2015, 33 : 353 - 359
  • [29] TV-Based image reconstruction of multiple objects in a fixed source-detector geometry
    Lu, Yang
    Yang, Zhenyu
    Zhao, Jun
    Wang, Ge
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2012, 20 (03) : 277 - 289
  • [30] Analysis of hip geometry by clinical CT for the assessment of hip fracture risk in elderly Japanese women
    Ito, Masako
    Wakao, Norimitsu
    Hida, Tetsuro
    Matsui, Yasumoto
    Abe, Yasue
    Aoyagi, Kiyoshi
    Uetani, Masataka
    Harada, Atsushi
    BONE, 2010, 46 (02) : 453 - 457