Enhancement of capnogram waveform in the presence of chest compression artefact during cardiopulmonary resuscitation

被引:9
|
作者
Ruiz de Gauna, Sofia [1 ]
Leturiondo, Mikel [1 ]
Gutierrez, J. Julio [1 ]
Ruiz, Jesus M. [1 ]
Gonzalez-Otero, Digna M. [1 ]
Russell, James K. [2 ]
Daya, Mohamud [2 ]
机构
[1] Univ Basque Country, UPV EHU, Dept Commun Engn, Bilbao 48013, Spain
[2] Oregon Hlth & Sci Univ, Dept Emergency Med, Portland, OR 97239 USA
关键词
Cardiopulmonary resuscitation; Advanced life support; Waveform capnography; Ventilation; Chest compressions; Chest compression artefact; VENTILATION DETECTION; MONITOR VENTILATION; CARBON-DIOXIDE; FEASIBILITY; GUIDELINES; ALGORITHMS; REMOVAL; CPR;
D O I
10.1016/j.resuscitation.2018.09.024
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Current resuscitation guidelines emphasize the use of waveform capnography to help guide rescuers during cardiopulmonary resuscitation (CPR). However, chest compressions often cause oscillations in the capnogram, impeding its reliable interpretation, either visual or automated. The aim of the study was to design an algorithm to enhance waveform capnography by suppressing the chest compression artefact. Methods: Monitor-defibrillator recordings from 202 patients in out-of-hospital cardiac arrest were analysed. Capnograms were classified according to the morphology of the artefact. Ventilations were annotated using the transthoracic impedance signal acquired through defibrillation pads. The suppression algorithm is designed to operate in real-time, locating distorted intervals and restoring the envelope of the capnogram. We evaluated the improvement in automated ventilation detection, estimation of ventilation rate, and detection of excessive ventilation rates (over-ventilation) using the capnograms before and after artefact suppression. Results: A total of 44 267 ventilations were annotated. After artefact suppression, sensitivity (Se) and positive predictive value (PPV) of the ventilation detector increased from 91.9/89.5% to 98.0/97.3% in the distorted episodes (83/202). Improvement was most noticeable for high-amplitude artefact, for which Se/PPV raised from 77.6/73.5% to 97.1/96.1%. Estimation of ventilation rate and detection of over-ventilation also upgraded. The suppression algorithm had minimal impact in non-distorted data. Conclusion: Ventilation detection based on waveform capnography improved after chest compression artefact suppression. Moreover, the algorithm enhances the capnogram tracing, potentially improving its clinical interpretation during CPR. Prospective research in clinical settings is needed to understand the feasibility and utility of the method.
引用
收藏
页码:53 / 58
页数:6
相关论文
共 50 条
  • [1] Influence of chest compression artefact on capnogram-based ventilation detection during out-of-hospital cardiopulmonary resuscitation
    Leturiondo, Mikel
    Ruiz de Gauna, Sofia
    Ruiz, Jesus M.
    Julio Gutierrez, J.
    Leturiondo, Luis A.
    Gonzalez-Otero, Digna M.
    Russell, James K.
    Zive, Dana
    Daya, Mohamud
    RESUSCITATION, 2018, 124 : 63 - 68
  • [2] Enhancing ventilation detection during cardiopulmonary resuscitation by filtering chest compression artifact from the capnography waveform
    Julio Gutierrez, Jose
    Leturiondo, Mikel
    Ruiz de Gauna, Sofia
    Maria Ruiz, Jesus
    Alberto Leturiondo, Luis
    Maria Gonzalez-Otero, Digna
    Zive, Dana
    Russell, James Knox
    Daya, Mohamud
    PLOS ONE, 2018, 13 (08):
  • [3] Effect of a dynamic mattress on chest compression quality during cardiopulmonary resuscitation
    Torsy, Tim
    Deswarte, Wim
    Karlberg Traav, Malin
    Beeckman, Dimitri
    NURSING IN CRITICAL CARE, 2022, 27 (02) : 275 - 281
  • [4] Closed-loop Adaptive Filtering for Supressing Chest Compression Oscillations in the Capnogram During Cardiopulmonary Resuscitation
    Leturiondo, Mikel
    Gutierrez, J. J.
    Ruiz de Gauna, Sofia
    Plaza, Sandra
    Veintemillas, Jose F.
    Daya, Mohamud
    2017 COMPUTING IN CARDIOLOGY (CINC), 2017, 44
  • [5] Prediction of blood pressure using chest compression waveform during cardiopulmonary resuscitation
    Han, Jiho
    Ahn, Kyo Jin
    Cha, Kyoung-Chul
    Kim, Sun Ju
    Jung, Woo Jin
    Roh, Young-Il
    Yoon, Young Ro
    Hwang, Sung Oh
    RESUSCITATION, 2024, 202
  • [6] Open-loop Adaptive Filtering for Suppressing Chest Compression Oscillations in the Capnogram During Cardiopulmonary Resuscitation
    Leturiondo, Mikel
    Ruiz, Jesus
    Gutierrez, J. J.
    Leturiondo, Luis A.
    Russell, James K.
    Daya, Mohamud
    2017 COMPUTING IN CARDIOLOGY (CINC), 2017, 44
  • [7] Realizing Optimal Chest Compression Fraction During Cardiopulmonary Resuscitation
    Talkington, Josh
    Dantu, Ram
    2016 IEEE FIRST INTERNATIONAL CONFERENCE ON CONNECTED HEALTH: APPLICATIONS, SYSTEMS AND ENGINEERING TECHNOLOGIES (CHASE), 2016, : 348 - 349
  • [8] Transthoracic impedance for the monitoring of quality of manual chest compression during cardiopulmonary resuscitation
    Zhang, Hehua
    Yang, Zhengfei
    Huang, Zitong
    Chen, Bihua
    Zhang, Lei
    Li, Heng
    Wu, Baoming
    Yu, Tao
    Li, Yongqin
    RESUSCITATION, 2012, 83 (10) : 1281 - 1286
  • [9] Feasibility of the capnogram to monitor ventilation rate during cardiopulmonary resuscitation
    Aramendi, Elisabete
    Elola, Andoni
    Alonso, Erik
    Irusta, Unai
    Daya, Mohamud
    Russell, James K.
    Hubner, Pia
    Sterz, Fritz
    RESUSCITATION, 2017, 110 : 162 - 168
  • [10] The Role of Chest Compressions on Ventilation during Advanced Cardiopulmonary Resuscitation
    Azcarate, Izaskun
    Uriguen, Jose Antonio
    Leturiondo, Mikel
    Sandoval, Camilo Leonardo
    Redondo, Koldo
    Gutierrez, Jose Julio
    Russell, James Knox
    Wallmueller, Pia
    Sterz, Fritz
    Daya, Mohamud Ramzan
    de Gauna, Sofia Ruiz
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (21)