New Non-contact Tonometer HNT-1P Reliability: Comparing Intraocular Pressure, Central Corneal Thickness, and Corrected Intraocular Pressure

被引:2
|
作者
Eah, Kyu Sang [1 ]
Shin, Joong Won [1 ]
Sung, Kyung Rim [1 ]
机构
[1] Univ Ulsan, Coll Med, Dept Ophthalmol Asan, Asan Med Ctr, Seoul, South Korea
来源
JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY | 2020年 / 61卷 / 05期
关键词
Central corneal thickness; Corrected intraocular pressure; HNT-1P; Intraocular pressure; Tonometry; GOLDMANN APPLANATION TONOMETRY; DYNAMIC CONTOUR TONOMETRY; BIOMECHANICAL PROPERTIES; ULTRASOUND PACHYMETRY; AGREEMENT; PARAMETERS;
D O I
10.3341/jkos.2020.61.5.524
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the reliability of a new non-contact tonometer, HNT-1P, by comparing its intraocular pressure (IOP), central corneal thickness (CCT), and corrected IOP (cIOP) measurements with the IOP measured by Goldmann applanation tonometry (GAT), the CCT measured with ultrasound pachymetry (USP), and the cIOP derived from GAT IOP and USP CCT. Methods: A total of 120 eyes of 65 healthy or glaucoma suspects were included in this study. The IOP was measured with the HNT-1P and GAT. The CCT was determined using the HNT-1P and USP. The IOP measured with GAT was corrected according to the CCT measured by USP. The correlations of the measurements obtained with the various instruments were evaluated using Pearson's correlation coefficient, and the agreement was assessed using Bland-Altman plots. Results: The average IOP measured with the HNT-1P and GAT was 14.8 +/- 5.1 mmHg and 15.6 +/- 5.2 mmHg, respectively, and the correlation coefficient between the two IOP measurements was 0.964 (p < 0.001). The mean CCT was 553.5 +/- 27.4 mu m using the HNT-1P and 550.6 +/- 26.3 mu m using USP, and the correlation coefficient between the two measurements was 0.913 (p < 0.001). The average cIOP was 14.6 +/- 5.0 and 13.4 +/- 6.5 mmHg according to the HNT-1P and GAT, respectively, and the correlation coefficient between the two corrected measurements was 0.956 (p < 0.001). Bland-Altman plots showed a high degree of agreement between the HNT-1P measurements and those of the other devices. Conclusions: The new non-contact HNT-1P tonometer provided reliable IOP, CCT, and cIOP measurements when compared with GAT and USP. HNT-1P could therefore be used clinically for reliable and accurate measurements of the IOP, CCT, and cIOP.
引用
收藏
页码:524 / 531
页数:8
相关论文
共 50 条
  • [41] Falsely elevated intraocular pressure due to increased central corneal thickness
    Bron, AM
    Creuzot-Garcher, C
    Goudeau-Boutillon, S
    d'Athis, P
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1999, 237 (03) : 220 - 224
  • [42] Diurnal variation of central corneal thickness and intraocular pressure in eyes with pseudoexfoliation
    Syed, Zeeshan
    Srikanth, Krishnagopal
    Nagarajan, Swathi
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2019, 67 (10) : 1607 - 1609
  • [43] Intraocular pressure and central corneal thickness in full-term newborns
    Maestri Ferreira, Claudia Cardoso
    Tavares, Ivan Maynart
    ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2017, 80 (05) : 313 - 316
  • [44] Influence of central corneal thickness on measured intraocular pressure differentials: Nidek RKT-7700, Topcon CT-80 NCTs and Goldmann Tonometer
    Ogbuehi, Kelechi C.
    Mucke, Sven
    Osuagwu, Uchechukwu L.
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2012, 32 (06) : 547 - 555
  • [45] Corneal hysteresis, central corneal thickness, and intraocular pressure in rheumatoid arthritis, and their relation to disease activity
    Nossair, Ashraf Ahmed
    Kassem, Mona Kassem
    Eltanamly, Rasha Mounir
    Alahmadawy, Yomna Amr
    MIDDLE EAST AFRICAN JOURNAL OF OPHTHALMOLOGY, 2021, 28 (03) : 174 - 179
  • [46] Relation between central corneal thickness and intraocular pressure in Punjabi population
    Kohli P.G.
    Randhawa B.K.
    Singh K.D.
    Randhawa G.S.
    Kohli A.K.
    Journal of Medical Engineering and Technology, 2010, 34 (01) : 1 - 6
  • [47] Effects of 0.5% tetracaine hydrochloride on central corneal thickness and intraocular pressure
    Manassakorn, Anita
    Chaidaroon, Winai
    ASIAN BIOMEDICINE, 2012, 6 (02) : 291 - 296
  • [48] Comparison of the intraocular pressure measurements with the tono-pen and the goldman applanation tonometer and the effect of central corneal thickness on measurements
    Dervisogullari, Mehmet Serdar
    Akarsu, Cengiz
    Ergin, Ahmet
    ANNALS OF CLINICAL AND ANALYTICAL MEDICINE, 2019, 10 (04): : 479 - 484
  • [49] Evaluation of Intraocular Pressure and Central Corneal Thickness in Patients with Thyroid Ophthalmopathy
    Ersoy, Murside
    Saricaoglu, M. Sinan
    Hamurcu, Mualla S.
    Karakurt, Ahmet
    Hasiripi, Hikmet
    TURK OFTALMOLOJI DERGISI-TURKISH JOURNAL OF OPHTHALMOLOGY, 2008, 38 (01): : 64 - 68
  • [50] The Influence of Central Corneal Thickness and Corneal Curvature and Axial Length on the Measurement of Intraocular Pressure
    Qing Li Minru Li Zhigang Fan Ningli WangMedical College of Sun Yatsen University Guangzhou ChinaZhongshan Ophthalmic Center Sun Yatsen University Guangzhou China
    眼科学报, 2002, (03) : 176 - 180