New Non-contact Tonometer HNT-1P Reliability: Comparing Intraocular Pressure, Central Corneal Thickness, and Corrected Intraocular Pressure

被引:2
|
作者
Eah, Kyu Sang [1 ]
Shin, Joong Won [1 ]
Sung, Kyung Rim [1 ]
机构
[1] Univ Ulsan, Coll Med, Dept Ophthalmol Asan, Asan Med Ctr, Seoul, South Korea
来源
JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY | 2020年 / 61卷 / 05期
关键词
Central corneal thickness; Corrected intraocular pressure; HNT-1P; Intraocular pressure; Tonometry; GOLDMANN APPLANATION TONOMETRY; DYNAMIC CONTOUR TONOMETRY; BIOMECHANICAL PROPERTIES; ULTRASOUND PACHYMETRY; AGREEMENT; PARAMETERS;
D O I
10.3341/jkos.2020.61.5.524
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the reliability of a new non-contact tonometer, HNT-1P, by comparing its intraocular pressure (IOP), central corneal thickness (CCT), and corrected IOP (cIOP) measurements with the IOP measured by Goldmann applanation tonometry (GAT), the CCT measured with ultrasound pachymetry (USP), and the cIOP derived from GAT IOP and USP CCT. Methods: A total of 120 eyes of 65 healthy or glaucoma suspects were included in this study. The IOP was measured with the HNT-1P and GAT. The CCT was determined using the HNT-1P and USP. The IOP measured with GAT was corrected according to the CCT measured by USP. The correlations of the measurements obtained with the various instruments were evaluated using Pearson's correlation coefficient, and the agreement was assessed using Bland-Altman plots. Results: The average IOP measured with the HNT-1P and GAT was 14.8 +/- 5.1 mmHg and 15.6 +/- 5.2 mmHg, respectively, and the correlation coefficient between the two IOP measurements was 0.964 (p < 0.001). The mean CCT was 553.5 +/- 27.4 mu m using the HNT-1P and 550.6 +/- 26.3 mu m using USP, and the correlation coefficient between the two measurements was 0.913 (p < 0.001). The average cIOP was 14.6 +/- 5.0 and 13.4 +/- 6.5 mmHg according to the HNT-1P and GAT, respectively, and the correlation coefficient between the two corrected measurements was 0.956 (p < 0.001). Bland-Altman plots showed a high degree of agreement between the HNT-1P measurements and those of the other devices. Conclusions: The new non-contact HNT-1P tonometer provided reliable IOP, CCT, and cIOP measurements when compared with GAT and USP. HNT-1P could therefore be used clinically for reliable and accurate measurements of the IOP, CCT, and cIOP.
引用
收藏
页码:524 / 531
页数:8
相关论文
共 50 条
  • [31] Combining corneal hysteresis with central corneal thickness and intraocular pressure for glaucoma risk assessment
    Pensyl, D.
    Sullivan-Mee, M.
    Torres-Monte, M.
    Halverson, K.
    Qualls, C.
    EYE, 2012, 26 (10) : 1349 - 1356
  • [32] Study of Accuracy of Intraocular Pressure measured by non-contact (air puff) Tonometer confirmed by Goldmann Applanation Tonometer
    Dibaji, Mohammad
    Shaikh, Rehan Moinuddin
    PAKISTAN JOURNAL OF MEDICAL & HEALTH SCIENCES, 2016, 10 (03): : 972 - 974
  • [33] Effects of central corneal thickness and corneal curvature on the intraocular pressure measurement by Goldmann applanation tonometer and ocular blood flow pneumatonometer
    Saleh, Tarek A.
    Adams, Morag
    McDermott, Bill
    Claridge, Kate G.
    Ewings, Paul
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2006, 34 (06) : 516 - 520
  • [34] Implications of Central Corneal Thickness-corrected Intraocular Pressure in Normal-tension Glaucoma with Reduced Central Corneal Thickness
    Lee, Sang Muk
    Baek, Sung Uk
    JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY, 2023, 64 (09): : 811 - 818
  • [35] Correlation between Central Corneal Thickness and Intraocular Pressure in Various Age Groups
    Marjanovic, Ivan
    Kontic, Djordje
    Hentova-Sencanic, Paraskeva
    Markovic, Vujica
    Bozic, Marija
    SRPSKI ARHIV ZA CELOKUPNO LEKARSTVO, 2010, 138 (5-6) : 279 - 286
  • [36] Difference in intraocular pressure measurements between non-contact tonometry and Goldmann applanation tonometry and the role of central corneal thickness in affecting glaucoma referrals
    Khalid Kamel
    Edward Dervan
    Kevin Falzon
    Colm O’Brien
    Irish Journal of Medical Science (1971 -), 2019, 188 : 321 - 325
  • [37] The Relationship between Diurnal Variations in Intraocular Pressure Measurements and Central Corneal Thickness and Corneal Hysteresis
    Kotecha, Aachal
    Crabb, David P.
    Spratt, Alexander
    Garway-Heath, David F.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2009, 50 (09) : 4229 - 4236
  • [38] Central corneal thickness and intraocular pressure relationship in eyes with and without previous LASIK: Comparison of Goldmann applanation tonometer with pneumatonometer
    Baypaktar, S
    Bayraktar, Z
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2005, 15 (01) : 81 - 88
  • [39] Correlation Between Intraocular Pressure and Central Corneal Thickness in Persian Children
    Nejabat M.
    Heidary F.
    Talebnejad M.R.
    Salouti R.
    Nowroozzadeh M.H.
    Masoumpour M.
    Mahdaviazad H.
    Tajbakhsh Z.
    Keshtkar M.
    Jamali H.
    Khalili M.R.
    Movahedan H.
    Roustaei N.
    Gharebaghi R.
    Ophthalmology and Therapy, 2016, 5 (2) : 235 - 243
  • [40] ICare Pro: Age Dependent Effect of Central Corneal Thickness on Intraocular Pressure in Glaucoma and Ocular Hypertension Patients
    Hohberger, Bettina
    Sommerfeld, Carlos
    Lucio, Marianna
    Bergua, Antonio
    CURRENT EYE RESEARCH, 2020, 45 (06) : 668 - 674