New Non-contact Tonometer HNT-1P Reliability: Comparing Intraocular Pressure, Central Corneal Thickness, and Corrected Intraocular Pressure

被引:2
|
作者
Eah, Kyu Sang [1 ]
Shin, Joong Won [1 ]
Sung, Kyung Rim [1 ]
机构
[1] Univ Ulsan, Coll Med, Dept Ophthalmol Asan, Asan Med Ctr, Seoul, South Korea
来源
JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY | 2020年 / 61卷 / 05期
关键词
Central corneal thickness; Corrected intraocular pressure; HNT-1P; Intraocular pressure; Tonometry; GOLDMANN APPLANATION TONOMETRY; DYNAMIC CONTOUR TONOMETRY; BIOMECHANICAL PROPERTIES; ULTRASOUND PACHYMETRY; AGREEMENT; PARAMETERS;
D O I
10.3341/jkos.2020.61.5.524
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the reliability of a new non-contact tonometer, HNT-1P, by comparing its intraocular pressure (IOP), central corneal thickness (CCT), and corrected IOP (cIOP) measurements with the IOP measured by Goldmann applanation tonometry (GAT), the CCT measured with ultrasound pachymetry (USP), and the cIOP derived from GAT IOP and USP CCT. Methods: A total of 120 eyes of 65 healthy or glaucoma suspects were included in this study. The IOP was measured with the HNT-1P and GAT. The CCT was determined using the HNT-1P and USP. The IOP measured with GAT was corrected according to the CCT measured by USP. The correlations of the measurements obtained with the various instruments were evaluated using Pearson's correlation coefficient, and the agreement was assessed using Bland-Altman plots. Results: The average IOP measured with the HNT-1P and GAT was 14.8 +/- 5.1 mmHg and 15.6 +/- 5.2 mmHg, respectively, and the correlation coefficient between the two IOP measurements was 0.964 (p < 0.001). The mean CCT was 553.5 +/- 27.4 mu m using the HNT-1P and 550.6 +/- 26.3 mu m using USP, and the correlation coefficient between the two measurements was 0.913 (p < 0.001). The average cIOP was 14.6 +/- 5.0 and 13.4 +/- 6.5 mmHg according to the HNT-1P and GAT, respectively, and the correlation coefficient between the two corrected measurements was 0.956 (p < 0.001). Bland-Altman plots showed a high degree of agreement between the HNT-1P measurements and those of the other devices. Conclusions: The new non-contact HNT-1P tonometer provided reliable IOP, CCT, and cIOP measurements when compared with GAT and USP. HNT-1P could therefore be used clinically for reliable and accurate measurements of the IOP, CCT, and cIOP.
引用
收藏
页码:524 / 531
页数:8
相关论文
共 50 条
  • [21] Difference in intraocular pressure measurements between non-contact tonometry and Goldmann applanation tonometry and the role of central corneal thickness in affecting glaucoma referrals
    Kamel, Khalid
    Dervan, Edward
    Falzon, Kevin
    O'Brien, Colm
    IRISH JOURNAL OF MEDICAL SCIENCE, 2019, 188 (01) : 321 - 325
  • [22] Comparison of three methods of intraocular pressure measurement and their relation to central corneal thickness
    Carbonaro, F.
    Andrew, T.
    Mackey, D. A.
    Spector, T. D.
    Hammond, C. J.
    EYE, 2010, 24 (07) : 1165 - 1170
  • [23] COMPARISON OF INTRAOCULAR PRESSURE MEASURED WITH GOLDMANN APPLANATION TONOMETER AND NON-CONTACT AIR-PUFF TONOMETER
    Mahsud, Hidayatullah
    Saleem, Muhammad Wali
    Muhammad, Rafiq
    Saleem, Muhammad
    Ayaz, Mirqad
    GOMAL JOURNAL OF MEDICAL SCIENCES, 2016, 14 (02): : 63 - 66
  • [24] Central Corneal Thickness and Intraocular Pressure in Women With Gestational Diabetes Mellitus
    Kan, Kok Wei
    Mohd, Mohd-Alif Wan
    Nik-Ahmad-Zuky, Nik Lah
    Shatriah, Ismail
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (03)
  • [25] Effect of central corneal thickness on intraocular pressure and comparison of Topcon CT-80 non-contact tonometry with Goldmann applanation tonometry
    Mansoori, Tarannum
    Balakrishna, Nagalla
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2018, 101 (02) : 206 - 212
  • [26] The influence of central corneal thickness and corneal curvature on Intraocular pressure measured by Tono-Pen and rebound tonometer in children
    Sahin, Afsun
    Basmak, Hikmet
    Yidirim, Nigun
    JOURNAL OF GLAUCOMA, 2008, 17 (01) : 57 - 61
  • [27] Comparability of Icare Pro Rebound Tonometer with Goldmann Applanation and Noncontact Tonometer in a Wide Range of Intraocular Pressure and Central Corneal Thickness
    Tamcelik, Nevbahar
    Atalay, Eray
    Cicik, Erdogan
    Ozkok, Ahmet
    OPHTHALMIC RESEARCH, 2015, 54 (01) : 18 - 25
  • [28] Central corneal thickness and intraocular pressure in the Cameroonian nonglaucomatous population
    Eballe, Andre Omgbwa
    Koki, Godefroy
    Ellong, Augustin
    Owono, Didier
    Epee, Emilienne
    Bella, Lucienne Assumpta
    Mvogo, Come Ebana
    Kouam, Jeanne Mayouego
    CLINICAL OPHTHALMOLOGY, 2010, 4 : 717 - 724
  • [29] Intraocular pressure and central corneal thickness in a healthy student population
    Stamenkovic, Miroslav
    Marjanovic, Ivan
    Maric, Vesna
    Kalezic, Tanja
    Bozic, Marija
    SRPSKI ARHIV ZA CELOKUPNO LEKARSTVO, 2024, 152 (3-4) : 175 - 178
  • [30] Alterations in intraocular pressure and central corneal thickness in premature newborns
    Zengin, Neslihan
    Zengin, Mehmet Ozgur
    Karahan, Eyyup
    Tuncer, Ibrahim
    IZMIR DR BEHCET UZ COCUK HASTANESI DERGISI, 2014, 4 (01): : 20 - 24