New Non-contact Tonometer HNT-1P Reliability: Comparing Intraocular Pressure, Central Corneal Thickness, and Corrected Intraocular Pressure

被引:2
作者
Eah, Kyu Sang [1 ]
Shin, Joong Won [1 ]
Sung, Kyung Rim [1 ]
机构
[1] Univ Ulsan, Coll Med, Dept Ophthalmol Asan, Asan Med Ctr, Seoul, South Korea
来源
JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY | 2020年 / 61卷 / 05期
关键词
Central corneal thickness; Corrected intraocular pressure; HNT-1P; Intraocular pressure; Tonometry; GOLDMANN APPLANATION TONOMETRY; DYNAMIC CONTOUR TONOMETRY; BIOMECHANICAL PROPERTIES; ULTRASOUND PACHYMETRY; AGREEMENT; PARAMETERS;
D O I
10.3341/jkos.2020.61.5.524
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the reliability of a new non-contact tonometer, HNT-1P, by comparing its intraocular pressure (IOP), central corneal thickness (CCT), and corrected IOP (cIOP) measurements with the IOP measured by Goldmann applanation tonometry (GAT), the CCT measured with ultrasound pachymetry (USP), and the cIOP derived from GAT IOP and USP CCT. Methods: A total of 120 eyes of 65 healthy or glaucoma suspects were included in this study. The IOP was measured with the HNT-1P and GAT. The CCT was determined using the HNT-1P and USP. The IOP measured with GAT was corrected according to the CCT measured by USP. The correlations of the measurements obtained with the various instruments were evaluated using Pearson's correlation coefficient, and the agreement was assessed using Bland-Altman plots. Results: The average IOP measured with the HNT-1P and GAT was 14.8 +/- 5.1 mmHg and 15.6 +/- 5.2 mmHg, respectively, and the correlation coefficient between the two IOP measurements was 0.964 (p < 0.001). The mean CCT was 553.5 +/- 27.4 mu m using the HNT-1P and 550.6 +/- 26.3 mu m using USP, and the correlation coefficient between the two measurements was 0.913 (p < 0.001). The average cIOP was 14.6 +/- 5.0 and 13.4 +/- 6.5 mmHg according to the HNT-1P and GAT, respectively, and the correlation coefficient between the two corrected measurements was 0.956 (p < 0.001). Bland-Altman plots showed a high degree of agreement between the HNT-1P measurements and those of the other devices. Conclusions: The new non-contact HNT-1P tonometer provided reliable IOP, CCT, and cIOP measurements when compared with GAT and USP. HNT-1P could therefore be used clinically for reliable and accurate measurements of the IOP, CCT, and cIOP.
引用
收藏
页码:524 / 531
页数:8
相关论文
共 37 条
[1]   Agreement between Central Corneal Thickness Measured Using Pentacam, Ultrasound Pachymetry, Specular Microscopy and Optic Biometer Lenstar IS 900 and the Influence of lntraocular Pressure [J].
Borrego-Sanz, L. ;
Saenz-Frances, F. ;
Bermudez-Vallecilla, M. ;
Morales-Fernandez, L. ;
Martinez-de-la-Casa, J. M. ;
Santos-Bueso, E. ;
Janez, L. ;
Garcia-Feijoo, J. .
OPHTHALMOLOGICA, 2014, 231 (04) :226-235
[2]   Influence of corneal structure, corneal responsiveness, and other ocular parameters on tonometric measurement of Intraocular pressure [J].
Broman, Aimee Teo ;
Congdon, Nathan G. ;
Bandeen-Roche, Karen ;
Quigley, Harry A. .
JOURNAL OF GLAUCOMA, 2007, 16 (07) :581-588
[3]   Comparison of dynamic contour tonometry and Goldmann applanation tonometry in deep lamellar and penetrating keratoplasties [J].
Ceruti, Piero ;
Morbio, Roberta ;
Marraffa, Michele ;
Marchini, Giorgio .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2008, 145 (02) :215-221
[4]   RELIABILITY OF INTRAOCULAR-PRESSURE MEASUREMENT WITH THE GOLDMANN APPLANATION TONOMETER IN EPIDEMIOLOGIC STUDIES [J].
DIELEMANS, I ;
VINGERLING, JR ;
HOFMAN, A ;
GROBBEE, DE ;
DEJONG, PTVM .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1994, 232 (03) :141-144
[5]   Human corneal thickness and its impact on intraocular pressure measures: A review and meta-analysis approach [J].
Doughty, MJ ;
Zaman, ML .
SURVEY OF OPHTHALMOLOGY, 2000, 44 (05) :367-408
[6]  
EHLERS N, 1975, ACTA OPHTHALMOL, V53, P652
[7]  
GOLDMANN H, 1957, Ophthalmologica, V134, P221
[8]   Effect of central corneal thickness on dynamic contour tonometry and goldmann applanation tonometry in primary open-angle glaucoma [J].
Grieshaber, Matthias C. ;
Schoetzau, Andreas ;
Zawinka, Claudia ;
Flammer, Josef ;
Orgul, Selim .
ARCHIVES OF OPHTHALMOLOGY, 2007, 125 (06) :740-744
[9]  
GROLMAN B, 1972, AMER J OPT ARCH AM A, V49, P646
[10]   Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer [J].
Gunvant, P ;
Baskaran, M ;
Vijaya, L ;
Joseph, IS ;
Watkins, RJ ;
Nallapothula, M ;
Broadway, DC ;
O'Leary, DJ .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2004, 88 (04) :518-522