Solvent-based de-emulsification dispersive liquid-liquid microextraction of palladium in environmental samples and determination by electrothermal atomic absorption spectrometry

被引:51
作者
Majidi, Behrooz [1 ]
Shemirani, Farzaneh [1 ]
机构
[1] Univ Tehran, Univ Coll Sci, Sch Chem, Tehran, Iran
关键词
Solvent-based de-emulsification dispersive liquid-liquid microextraction; Palladium; Environmental samples; Electrothermal atomic absorption spectrometry; ORTHOGONAL SIGNAL CORRECTION; PLASMA-MASS SPECTROMETRY; PLATINUM-GROUP ELEMENTS; CLOUD-POINT EXTRACTION; SPECTROPHOTOMETRIC DETERMINATION; WATER SAMPLES; PRECONCENTRATION; SEPARATION; SILICA; INTERFERENCES;
D O I
10.1016/j.talanta.2012.02.026
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new simple, rapid and efficient sample pretreatment technique for simultaneous extraction and preconcentration was developed. In this study, solvent-based de-emulsification dispersive liquid-liquid microextraction (SD-DLLME) was developed for preconcentration and highly sensitive determination of Pd(II) using the electrothermal atomic absorption spectrometry. The extraction of Pd(II) was performed in the presence of Thio-Michler's ketone (TMK) as the complexing agent. After dispersing, when an aliquot of acetonitrile was introduced as a chemical demulsifier into the aqueous bulk, the obtained emulsion cleared into two phases quickly. Therefore, an extra centrifugation step is not needed for phase separation. Several factors that influence the microextraction efficiency and ETAAS signal, such as pH. TMK concentration, counter ion, extraction time, disperser, de-emulsifier solvents and extraction solvents, stirring rate, pyrolysis and atomization temperature were investigated. Under the optimized conditions, the calibration graph was linear over the range of 0.025-0.500 mu g L-1 and relative standard deviation of 3.68% at 0.1 mu g L-1 was obtained (n =7). The limit of detection and the enrichment factor (EF) were obtained 0.007 mu g L-1 and 231, respectively. The obtained results indicated that the developed method is an excellent alternative for the routine analysis in the environmental field. Published by Elsevier B.V.
引用
收藏
页码:245 / 251
页数:7
相关论文
共 37 条
[1]   Stopped-flow injection liquid-liquid extraction spectrophotometric determination of palladium in airborne particulate matter and automobile catalysts [J].
Anthemidis, AN ;
Themelis, DG ;
Stratis, JA .
TALANTA, 2001, 54 (01) :37-43
[2]   Coagulation and electrocoagulation of oil-in-water emulsions [J].
Canizares, Pablo ;
Martinez, Fabiola ;
Jimenez, Carlos ;
Saez, Cristina ;
Rodrigo, Manuel A. .
JOURNAL OF HAZARDOUS MATERIALS, 2008, 151 (01) :44-51
[3]   Simultaneous determination of cobalt, nickel and palladium in micellar media using partial least square regression and direct orthogonal signal correction [J].
Chamjangali, MA ;
Bagherian, G ;
Azizi, G .
SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2005, 62 (1-3) :189-196
[4]   Low-density extraction solvent-based solvent terminated dispersive liquid-liquid microextraction combined with gas chromatography-tandem mass spectrometry for the determination of carbamate pesticides in water samples [J].
Chen, Hao ;
Chen, Ruiwen ;
Li, Shengqing .
JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (08) :1244-1248
[5]  
Cheng K.L., 1982, Handbook of Organic Analytical Reagents
[6]   4,4'-BIS(DIMETHYLAMINO)THIOBENZOPHENONE AS A SENSITIVE REAGENT FOR MERCURY AND PALLADIUM [J].
CHENG, KL ;
GOYDISH, BL .
MICROCHEMICAL JOURNAL, 1966, 10 (1-4) :158-&
[7]   Determination of platinum and palladium in environmental samples by graphite furnace atomic absorption spectrometry after separation on dithizone sorbent [J].
Chwastowska, J ;
Skwara, W ;
Sterlinska, E ;
Pszonicki, L .
TALANTA, 2004, 64 (01) :224-229
[8]   Determination of platinum group elements in impact breccias using neutron activation analysis and ultrasonic nebulization inductively coupled plasma mass spectrometry after anion exchange preconcentration [J].
Dai, XX ;
Koeberl, C ;
Fröschl, H .
ANALYTICA CHIMICA ACTA, 2001, 436 (01) :79-85
[9]   Determination of platinum, palladium and rhodium in biological and environmental samples by low temperature electrothermal vaporization inductively coupled plasma atomic emission spectrometry with diethyldithiocarbamate as chemical modifier [J].
Fan, ZF ;
Jiang, ZC ;
Yang, F ;
Hu, B .
ANALYTICA CHIMICA ACTA, 2004, 510 (01) :45-51
[10]   Minimization of mass interferences in quadrupole inductively coupled plasma mass spectrometric (ICP-MS) determination of palladium using a flow injection on-line displacement solid-phase extraction protocol [J].
Fang, Jing ;
Liu, Li-Wen ;
Yan, Xiu-Ping .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2006, 61 (07) :864-869