A Review on the Hemodialysis Reliable Outflow (HeRO) Graft for Haemodialysis Vascular Access

被引:52
作者
Al Shakarchi, J. [1 ,2 ]
Houston, J. G. [2 ,3 ]
Jones, R. G. [4 ]
Inston, N. [1 ,2 ]
机构
[1] Univ Hosp Birmingham, Queen Elizabeth Hosp Birmingham, Dept Renal Surg, Birmingham, W Midlands, England
[2] Univ Dundee, ReDVA Res Consortium, Dundee, Scotland
[3] Univ Dundee, Cardiovasc & Diabet Med, Dundee, Scotland
[4] Univ Hosp Birmingham, Dept Radiol, QEHB, Birmingham, W Midlands, England
关键词
Graft; Haemodialysis; HeRO; Outflow; Catheter; ARTERIOVENOUS GRAFTS; DIALYSIS ACCESS; END-STAGE; EXPERIENCE;
D O I
10.1016/j.ejvs.2015.03.059
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objectives: With improved dialysis survival there are increasing numbers of patients who have exhausted definitive access options due to central venous stenosis and are maintaining dialysis on a central venous catheter. The Hemodialysis Reliable Outflow (HeRO) allows an alternative by providing a definitive access solution. The aim of this study is to systematically review the published outcomes of the HeRO graft and discuss the role in complex haemodialysis patients. Methods: Electronic databases were searched for studies assessing the use of the HeRO graft for dialysis in accordance with PRISMA published up to December 31 2014. The primary outcomes for this study were 1-year primary and secondary patency rates. Secondary outcomes were rates of dialysis access associated steal syndrome, HeRO-related bacteraemia rates and rates of interventions. Results: Following strict inclusion/exclusion criteria, eight studies including 409 patients were included in our review. Primary and secondary pooled patency rates in this complex cohort of dialysis patients were found to be 21.9% (9.6-37.2%) and 59.4% (39.4-78%). The rate of dialysis access associated steal syndrome was low at 6.3% (1-14.7%) as was the range of HeRO-related bacteraemia (0.13-0.7 events per 1000 days). Conclusions: This literature review shows that the HeRO graft is an acceptable option for complex dialysis patients who are catheter dependent. Owing to device availability, published data are predominantly North American and further longer-term studies in other populations may be necessary. In this challenging patient group, randomized controlled trials are required to allow comparisons with alternative access options. (C) 2015 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:108 / 113
页数:6
相关论文
共 17 条
  • [1] The use of tunneled central venous catheters: inevitable or system failure?
    Aitken, Emma L.
    Stevenson, Karen S.
    Gingell-Littlejohn, Marc
    Aitken, Margaret
    Clancy, Marc
    Kingsmore, David B.
    [J]. JOURNAL OF VASCULAR ACCESS, 2014, 15 (05) : 344 - 350
  • [2] Hemodialysis Reliable Outflow (HeRO) device in end-stage dialysis access: a decision analysis model
    Dageforde, Leigh Anne
    Bream, Peter R.
    Moore, Derek E.
    [J]. JOURNAL OF SURGICAL RESEARCH, 2012, 177 (01) : 165 - 171
  • [3] Multi-center Experience of 164 Consecutive Hemodialysis Reliable Outflow [HeRO] Graft Implants for Hemodialysis Treatment
    Gage, S. M.
    Katzman, H. E.
    Ross, J. R.
    Hohmann, S. E.
    Sharpe, C. A.
    Butterly, D. W.
    Lawson, J. H.
    [J]. EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2012, 44 (01) : 93 - 99
  • [4] Percutaneous Interventions on the Hemodialysis Reliable Outflow Vascular Access Device
    Gebhard, Thomas A.
    Bryant, Joshua A.
    Grezaffi, J. Adam
    Pabon-Ramos, Waleska M.
    Gage, Shawn M.
    Miller, Michael J.
    Husum, Kurt W.
    Suhocki, Paul V.
    Sopko, David R.
    Lawson, Jeffrey H.
    Smith, Tony P.
    Kim, Charles Y.
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2013, 24 (04) : 543 - 549
  • [5] Initial experience and outcome of a new hemodialysis access device for catheter-dependent patients
    Katzman, Howard E.
    McLafferty, Robert B.
    Ross, John R.
    Glickman, Marc H.
    Peden, Eric K.
    Lawson, Jeffery H.
    [J]. JOURNAL OF VASCULAR SURGERY, 2009, 50 (03) : 600 - 607
  • [6] Experience of HeRO Dialysis Graft Placement in a Challenging Population
    Kokkosis, Angela A.
    Abramowitz, Steven D.
    Schwitzer, Jonathan
    Schanzer, Harry
    Teodorescu, Victoria J.
    [J]. VASCULAR AND ENDOVASCULAR SURGERY, 2013, 47 (04) : 278 - 280
  • [7] The End Stage of Dialysis Access: Femoral Graft or HeRO Vascular Access Device
    Kudlaty, Elizabeth A.
    Pan, Jeanne
    Allemang, Matthew T.
    Kendrick, Daniel E.
    Kashyap, Vikram S.
    Wong, Virginia L.
    [J]. ANNALS OF VASCULAR SURGERY, 2015, 29 (01) : 90 - 97
  • [8] INVERSE OF FREEMAN-TUKEY DOUBLE ARCSINE TRANSFORMATION
    MILLER, JJ
    [J]. AMERICAN STATISTICIAN, 1978, 32 (04) : 138 - 138
  • [9] Moher D, 2009, ANN INTERN MED, V151, P264, DOI [10.7326/0003-4819-151-4-200908180-00135, 10.1136/bmj.b2700, 10.1371/journal.pmed.1000097, 10.1136/bmj.i4086, 10.1016/j.ijsu.2010.02.007, 10.1016/j.ijsu.2010.07.299, 10.1136/bmj.b2535, 10.1186/2046-4053-4-1]
  • [10] A Comparison Between the HeRO Graft and Conventional Arteriovenous Grafts in Hemodialysis Patients
    Nassar, George M.
    Glickman, Marc H.
    McLafferty, Robert B.
    Croston, J. Kevin
    Zarge, Joseph I.
    Katzman, Howard E.
    Peden, Eric K.
    Lawson, Jeffrey H.
    Martinez, Jeffrey M.
    Thackeray, Lisa