Contact conditions for total hip head-neck modular taper junctions with microgrooved stem tapers

被引:20
|
作者
Bechstedt, Maren [1 ]
Gustafson, Jonathan A. [2 ]
Mell, Steven P. [2 ]
Guehrs, Julian [1 ]
Morlock, Michael M. [1 ]
Levine, Brett R. [2 ]
Lundberg, Hannah J. [2 ]
机构
[1] TUHH Hamburg Univ Technol, Inst Biomech, D-21073 Hamburg, Germany
[2] Rush Univ, Dept Orthoped Surg, Med Ctr, 1611 W Harrison St Suite 201, Chicago, IL 60612 USA
关键词
Total hip arthroplasty; Modularity; Surface topography; Finite element analysis; Contact mechanics; FINITE-ELEMENT; FRETTING WEAR; MORSE TAPER; MECHANICAL RESPONSE; REALISTIC FORCES; INTERFACE; STRENGTH; REPLACEMENTS; PREDICTION; IMPLANTS;
D O I
10.1016/j.jbiomech.2020.109689
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Implant failure due to fretting-corrosion of head-neck modular junctions is a rising problem in total hip arthroplasty. Fretting-corrosion initiates when micromotion leads to metal release; however, factors leading to micromotion, such as microgrooves on the stem taper, are not fully understood. The purpose of this study is to describe a finite element analysis technique to determine head-neck contact mechanics and investigate the effect of stem taper microgroove height during head-neck assembly. Two-dimensional axisymmetric finite element models were created. Models were created for a ceramic femoral head and a CoCrMo femoral head against Ti6AI4V stem tapers and compared to available data from prior experiments. Stem taper microgroove height was investigated with a generic 12/14 model. Head-neck assembly was performed to four maximum loads (500 N, 2000 N, 4000 N, 8000 N). For the stem taper coupled with the ceramic head, the number of microgrooves in contact and plastically deformed differed by 2.5 microgrooves (4%) and 6.5 microgrooves (11%), respectively, between the finite element models and experiment. For the stem taper coupled with the CoCrMo head, all microgrooves were in contact after all assembly loads in the finite element model due to an almost identical conical angle between the taper surfaces. In the experiments, all grooves were only in contact for the 8000 N assembly load. Contact area, plastic (permanent) deformation, and contact pressure increased with increasing assembly loads and deeper microgrooves. The described modeling technique can be used to investigate the relationship between implant design factors, allowing for optimal microgroove design within material couples. (C) 2020 Elsevier Ltd. All rights reserved.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] An adaptive finite element simulation of fretting wear damage at the head neck taper junction of total hip replacement: The role of taper angle mismatch
    Fallahnezhad, Khosro
    Oskouei, Reza H.
    Badnava, Hojjat
    Taylor, Mark
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2017, 75 : 58 - 67
  • [42] A combined experimental and finite element approach to analyse the fretting mechanism of the head-stem taper junction in total hip replacement
    Bitter, Thom
    Khan, Imran
    Marriott, Tim
    Lovelady, Elaine
    Verdonschot, Nico
    Janssen, Dennis
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2017, 231 (09) : 862 - 870
  • [43] Spontaneous reduction in both dislocated head and dissociation of head–neck in a modular total hip arthroplasty
    A. Raviraj
    Ashish Anand
    Murali Chakravarthy
    Sanjay Pai
    European Journal of Orthopaedic Surgery & Traumatology, 2012, 22 : 127 - 130
  • [44] A modular femoral neck and head system works well in cementless total hip replacement for patients with developmental dysplasia of the hip
    Sakai, T.
    Ohzono, K.
    Nishii, T.
    Miki, H.
    Takao, M.
    Sugano, N.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2010, 92B (06): : 770 - 776
  • [45] The Role of the Assembly Force in the Tribocorrosion Behaviour of Hip Implant Head-Neck Junctions: An Adaptive Finite Element Approach
    Fallahnezhad, Khosro
    Feyzi, Mohsen
    Hashemi, Reza
    Taylor, Mark
    BIOENGINEERING-BASEL, 2022, 9 (11):
  • [46] Long-term fretting corrosion performance of modular head-neck junctions with self-reinforced composite gaskets from PEEK and UHMWPE
    Ouellette, Eric S.
    Zhu, Dongkai
    Liu, Yangping
    Gilbert, Jeremy L.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2022, 129
  • [47] Assembly force and taper angle difference influence the relative motion at the stem-neck interface of bi-modular hip prostheses
    Haschke, Henning
    Jauch-Matt, Sabrina Y.
    Sellenschloh, Kay
    Huber, Gerd
    Morlock, Michael M.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2016, 230 (07) : 690 - 699
  • [48] Dissociation of modular total hip arthroplasty at the neck-stem interface without dislocation
    Kouzelis A.
    Georgiou C.S.
    Megas P.
    Journal of Orthopaedics and Traumatology, 2012, 13 (4) : 221 - 224
  • [49] Method for Characterization of Material Loss from Modular Head-Stem Taper Surfaces of Hip Replacement Devices
    Racasan, Radu
    Bills, Paul
    Blunt, Liam
    Hart, Alister
    Skinner, John
    MODULARITY AND TAPERS IN TOTAL JOINT REPLACEMENT DEVICES, 2015, 1591 : 132 - 146
  • [50] Clinical Results of Revision Hip Arthroplasty for Neck-Taper Corrosion and Adverse Local Tissue Reactions Around a Modular Neck Stem
    Inoue, Daisuke
    Restrepo, Camilo
    Nourie, Blake
    Hozack, William J.
    JOURNAL OF ARTHROPLASTY, 2020, 35 (06): : S289 - S293