Contact conditions for total hip head-neck modular taper junctions with microgrooved stem tapers

被引:20
|
作者
Bechstedt, Maren [1 ]
Gustafson, Jonathan A. [2 ]
Mell, Steven P. [2 ]
Guehrs, Julian [1 ]
Morlock, Michael M. [1 ]
Levine, Brett R. [2 ]
Lundberg, Hannah J. [2 ]
机构
[1] TUHH Hamburg Univ Technol, Inst Biomech, D-21073 Hamburg, Germany
[2] Rush Univ, Dept Orthoped Surg, Med Ctr, 1611 W Harrison St Suite 201, Chicago, IL 60612 USA
关键词
Total hip arthroplasty; Modularity; Surface topography; Finite element analysis; Contact mechanics; FINITE-ELEMENT; FRETTING WEAR; MORSE TAPER; MECHANICAL RESPONSE; REALISTIC FORCES; INTERFACE; STRENGTH; REPLACEMENTS; PREDICTION; IMPLANTS;
D O I
10.1016/j.jbiomech.2020.109689
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Implant failure due to fretting-corrosion of head-neck modular junctions is a rising problem in total hip arthroplasty. Fretting-corrosion initiates when micromotion leads to metal release; however, factors leading to micromotion, such as microgrooves on the stem taper, are not fully understood. The purpose of this study is to describe a finite element analysis technique to determine head-neck contact mechanics and investigate the effect of stem taper microgroove height during head-neck assembly. Two-dimensional axisymmetric finite element models were created. Models were created for a ceramic femoral head and a CoCrMo femoral head against Ti6AI4V stem tapers and compared to available data from prior experiments. Stem taper microgroove height was investigated with a generic 12/14 model. Head-neck assembly was performed to four maximum loads (500 N, 2000 N, 4000 N, 8000 N). For the stem taper coupled with the ceramic head, the number of microgrooves in contact and plastically deformed differed by 2.5 microgrooves (4%) and 6.5 microgrooves (11%), respectively, between the finite element models and experiment. For the stem taper coupled with the CoCrMo head, all microgrooves were in contact after all assembly loads in the finite element model due to an almost identical conical angle between the taper surfaces. In the experiments, all grooves were only in contact for the 8000 N assembly load. Contact area, plastic (permanent) deformation, and contact pressure increased with increasing assembly loads and deeper microgrooves. The described modeling technique can be used to investigate the relationship between implant design factors, allowing for optimal microgroove design within material couples. (C) 2020 Elsevier Ltd. All rights reserved.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Corrosion at the Head-Neck Taper as a Cause for Adverse Local Tissue Reactions After Total Hip Arthroplasty
    Cooper, H. John
    Della Valle, Craig J.
    Berger, Richard A.
    Tetreault, Matthew
    Paprosky, Wayne G.
    Sporer, Scott M.
    Jacobs, Joshua J.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2012, 94A (18): : 1655 - 1661
  • [22] The wear of the taper junction of the stem trunnion with the femoral head of a total modular hip prosthesis
    Capitanu, Lucian
    Badita, Liliana-Laura
    Tiganesteanu, Constantin
    Florescu, Virgil
    JURNAL TRIBOLOGI, 2019, 23 : 61 - 75
  • [23] Same Same but Different? 12/14 Stem and Head Tapers in Total Hip Arthroplasty
    Mueller, Ulrike
    Braun, Steffen
    Schroeder, Stefan
    Sonntag, Robert
    Kretzer, J. Philippe
    JOURNAL OF ARTHROPLASTY, 2017, 32 (10): : 3191 - 3199
  • [24] In vitro testing for hip head-neck taper tribocorrosion: A review of experimental methods
    Wight, Christian M.
    Schemitsch, Emil H.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2022, 236 (04) : 469 - 482
  • [25] The seating mechanics of head-neck modular tapers in vitro: Load-displacement measurements, moisture, and rate effects
    Ouellette, Eric S.
    Shenoy, Aarti A.
    Gilbert, Jeremy L.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2018, 36 (04) : 1164 - 1172
  • [26] Corrosion of the Head-neck Junction After Total Hip Arthroplasty
    Jennings, Jason M.
    Dennis, Douglas A.
    Yang, Charlie C.
    JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2016, 24 (06) : 349 - 356
  • [27] Risk Factors Associated With Early Complications of Revision Surgery for Head-Neck Taper Corrosion in Metal-on-Polyethylene Total Hip Arthroplasty
    Kwon, Young-Min
    Rossi, Daniel
    MacAuliffe, John
    Peng, Yun
    Arauz, Paul
    JOURNAL OF ARTHROPLASTY, 2018, 33 (10): : 3231 - 3237
  • [28] Parametric analysis of the effect of impaction load on the stability of head-neck junction in total hip arthroplasty
    Zhang, Guoxian
    Yang, Shu
    Cui, Wen
    Peng, Xing
    Zhang, Xiaogang
    Zhang, Yali
    Li, Junyan
    Jin, Zhongmin
    CLINICAL BIOMECHANICS, 2022, 94
  • [29] Dissociation of modular total hip arthroplasty at the femoral head-neck interface after loosening of the acetabular shell following hip dislocation
    Chu, CM
    Wang, SJ
    Lin, LC
    JOURNAL OF ARTHROPLASTY, 2001, 16 (06): : 806 - 809
  • [30] Design, Material, and Seating Load Effects on In Vitro Fretting Corrosion Performance of Modular Head-Neck Tapers
    Ouellette, Eric S.
    Mali, Sachin A.
    Kim, Jua
    Grostefon, Justin
    Gilbert, Jeremy L.
    JOURNAL OF ARTHROPLASTY, 2019, 34 (05): : 991 - 1002