Contact conditions for total hip head-neck modular taper junctions with microgrooved stem tapers

被引:20
作者
Bechstedt, Maren [1 ]
Gustafson, Jonathan A. [2 ]
Mell, Steven P. [2 ]
Guehrs, Julian [1 ]
Morlock, Michael M. [1 ]
Levine, Brett R. [2 ]
Lundberg, Hannah J. [2 ]
机构
[1] TUHH Hamburg Univ Technol, Inst Biomech, D-21073 Hamburg, Germany
[2] Rush Univ, Dept Orthoped Surg, Med Ctr, 1611 W Harrison St Suite 201, Chicago, IL 60612 USA
关键词
Total hip arthroplasty; Modularity; Surface topography; Finite element analysis; Contact mechanics; FINITE-ELEMENT; FRETTING WEAR; MORSE TAPER; MECHANICAL RESPONSE; REALISTIC FORCES; INTERFACE; STRENGTH; REPLACEMENTS; PREDICTION; IMPLANTS;
D O I
10.1016/j.jbiomech.2020.109689
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Implant failure due to fretting-corrosion of head-neck modular junctions is a rising problem in total hip arthroplasty. Fretting-corrosion initiates when micromotion leads to metal release; however, factors leading to micromotion, such as microgrooves on the stem taper, are not fully understood. The purpose of this study is to describe a finite element analysis technique to determine head-neck contact mechanics and investigate the effect of stem taper microgroove height during head-neck assembly. Two-dimensional axisymmetric finite element models were created. Models were created for a ceramic femoral head and a CoCrMo femoral head against Ti6AI4V stem tapers and compared to available data from prior experiments. Stem taper microgroove height was investigated with a generic 12/14 model. Head-neck assembly was performed to four maximum loads (500 N, 2000 N, 4000 N, 8000 N). For the stem taper coupled with the ceramic head, the number of microgrooves in contact and plastically deformed differed by 2.5 microgrooves (4%) and 6.5 microgrooves (11%), respectively, between the finite element models and experiment. For the stem taper coupled with the CoCrMo head, all microgrooves were in contact after all assembly loads in the finite element model due to an almost identical conical angle between the taper surfaces. In the experiments, all grooves were only in contact for the 8000 N assembly load. Contact area, plastic (permanent) deformation, and contact pressure increased with increasing assembly loads and deeper microgrooves. The described modeling technique can be used to investigate the relationship between implant design factors, allowing for optimal microgroove design within material couples. (C) 2020 Elsevier Ltd. All rights reserved.
引用
收藏
页数:7
相关论文
共 37 条
[1]   Microgrooved Surface Topography Does Not Influence Fretting Corrosion of Tapers in Total Hip Arthroplasty: Classification and Retrieval Analysis [J].
Arnholt, Christina ;
Underwood, Richard ;
MacDonald, Daniel W. ;
Higgs, Genymphas B. ;
Chen, Antonia F. ;
Klein, Gregg ;
Hamlin, Brain ;
Lee, Gwo-Chin ;
Mont, Michael ;
Cates, Harlod ;
Malkani, Arthur ;
Kraay, Matthew ;
Rimnac, Clare ;
Kurtz, Steven M. .
MODULARITY AND TAPERS IN TOTAL JOINT REPLACEMENT DEVICES, 2015, 1591 :99-112
[2]   Does a micro-grooved trunnion stem surface finish improve fixation and reduce fretting wear at the taper junction of total hip replacements? A finite element evaluation [J].
Ashkanfar, Ariyan ;
Langton, David J. ;
Joyce, Thomas J. .
JOURNAL OF BIOMECHANICS, 2017, 63 :47-54
[3]   A large taper mismatch is one of the key factors behind high wear rates and failure at the taper junction of total hip replacements: A finite element wear analysis [J].
Ashkanfar, Ariyan ;
Langton, David J. ;
Joyce, Thomas J. .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2017, 69 :257-266
[4]  
Bechstedt M., 2014, KONUSVERBINDUNG ENDO
[5]   Finite element wear prediction using adaptive meshing at the modular taper interface of hip implants [J].
Bitter, Thom ;
Khan, Imran ;
Marriott, Tim ;
Lovelady, Elaine ;
Verdonschot, Nico ;
Janssen, Dennis .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2018, 77 :616-623
[6]   A combined experimental and finite element approach to analyse the fretting mechanism of the head-stem taper junction in total hip replacement [J].
Bitter, Thom ;
Khan, Imran ;
Marriott, Tim ;
Lovelady, Elaine ;
Verdonschot, Nico ;
Janssen, Dennis .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2017, 231 (09) :862-870
[7]   Total hip arthroplasty head-neck contact mechanics: A stochastic investigation of key parameters [J].
Donaldson, Finn E. ;
Coburn, James C. ;
Siegel, Karen Lohmann .
JOURNAL OF BIOMECHANICS, 2014, 47 (07) :1634-1641
[8]   Finite element analysis of the head-neck taper interface of modular hip prostheses [J].
Dyrkacz, R. M. R. ;
Brandt, J. M. ;
Morrison, J. B. ;
O' Brien, S. T. ;
Ojo, O. A. ;
Turgeon, T. R. ;
Wyss, U. P. .
TRIBOLOGY INTERNATIONAL, 2015, 91 :206-213
[9]   The effect of different assembly loads on taper junction fretting wear in total hip replacements [J].
English, Russell ;
Ashkanfar, Ariyan ;
Rothwell, Glynn .
TRIBOLOGY INTERNATIONAL, 2016, 95 :199-210
[10]   A computational approach to fretting wear prediction at the head-stem taper junction of total hip replacements [J].
English, Russell ;
Ashkanfar, Ariyan ;
Rothwell, Glynn .
WEAR, 2015, 338 :210-220