Determination of the spring constants of probes for force microscopy/spectroscopy

被引:178
作者
Gibson, CT
Watson, GS
Myhra, S
机构
[1] Faculty of Science and Technology, Griffith University, Nathan
关键词
D O I
10.1088/0957-4484/7/3/014
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
A direct, accurate and convenient procedure for calibrating the spring constants of probes used for force microscopy/spectroscopy is described. It amounts to deflecting an 'unknown' cantilever with a 'standard' lever, where the standard lever has been precalibrated. The absolute and relative accuracies of the procedure are +/-20-30% and +/-10-20%, respectively; the former is limited by uncertainties in the determination of the spring constant for the 'standard' lever, as well as that for the 'unknown'. The method differs from others of the static deflection variety by its exploitation of the routine features of current instruments. The technique is compared with other static and dynamic methods currently being used.
引用
收藏
页码:259 / 262
页数:4
相关论文
共 19 条
  • [1] MICROFABRICATION OF CANTILEVER STYLI FOR THE ATOMIC FORCE MICROSCOPE
    ALBRECHT, TR
    AKAMINE, S
    CARVER, TE
    QUATE, CF
    [J]. JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1990, 8 (04): : 3386 - 3396
  • [2] ATOMIC FORCE MICROSCOPE
    BINNIG, G
    QUATE, CF
    GERBER, C
    [J]. PHYSICAL REVIEW LETTERS, 1986, 56 (09) : 930 - 933
  • [3] DIRECT MEASUREMENT OF HYDROGEN-BONDING IN DNA NUCLEOTIDE BASES BY ATOMIC-FORCE MICROSCOPY
    BOLAND, T
    RATNER, BD
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1995, 92 (12) : 5297 - 5301
  • [4] A NONDESTRUCTIVE METHOD FOR DETERMINING THE SPRING CONSTANT OF CANTILEVERS FOR SCANNING FORCE MICROSCOPY
    CLEVELAND, JP
    MANNE, S
    BOCEK, D
    HANSMA, PK
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (02) : 403 - 405
  • [5] HEUBERGER M, 1994, NANOTECHNOLOGY, V5, P12
  • [6] HOH JH, 1994, J CELL SCI, V107, P1105
  • [7] CALIBRATION OF ATOMIC-FORCE MICROSCOPE TIPS
    HUTTER, JL
    BECHHOEFER, J
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (07) : 1868 - 1873
  • [8] MEASUREMENT AND MANIPULATION OF VAN-DER-WAALS FORCES IN ATOMIC-FORCE MICROSCOPY
    HUTTER, JL
    BECHHOEFER, J
    [J]. JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1994, 12 (03): : 2251 - 2253
  • [9] DIRECT MEASUREMENT OF INTERACTION FORCES BETWEEN COLLOIDAL PARTICLES USING THE SCANNING FORCE MICROSCOPE
    LI, YQ
    TAO, NJ
    PAN, J
    GARCIA, AA
    LINDSAY, SM
    [J]. LANGMUIR, 1993, 9 (03) : 637 - 641
  • [10] DIRECT OBSERVATION OF THE ATOMIC-FORCE MICROSCOPY TIP USING INVERSE ATOMIC-FORCE MICROSCOPY IMAGING
    MONTELIUS, L
    TEGENFELDT, JO
    VANHEEREN, P
    [J]. JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1994, 12 (03): : 2222 - 2226