A GRATING-BASED PLASMON BIOSENSOR WITH HIGH RESOLUTION

被引:10
作者
Luo, Z. [1 ]
Suyama, T. [1 ]
Xu, X. [2 ]
Okuno, Y. [1 ]
机构
[1] Kumamoto Univ, Grad Sch Sci & Technol, Kumamoto 8608555, Japan
[2] Kyushu Sangyo Univ, Dept Elect Engn & Informat Technol, Fukuoka 8138503, Japan
来源
PROGRESS IN ELECTROMAGNETICS RESEARCH-PIER | 2011年 / 118卷
关键词
SURFACE-PLASMONS; RESONANCE BIOSENSOR; ENHANCEMENT;
D O I
10.2528/PIER11060103
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We present an idea of grating-based plasmon biosensor utilizing phase detection to realize high resolution in finding a refractive index of a material put on the surface of a metal grating. Considering a trade-off between high resolution and experimental practicability, we show a table of recommended setup that covers a wide range of the index. Keeping the diffraction efficiency no less than 10(-3) and assuming the resolution in phase detection to be 2.5 x 10(-2) degrees, we estimate the resolution of the biosensor as 7.5 x 10(-7) refractive index units. We also discuss the possible improvement to realize a predicted superior limit of resolution around 10(-8) refractive index units.
引用
收藏
页码:527 / 539
页数:13
相关论文
共 18 条
[1]   ENHANCED REFRACTOMETRIC OPTICAL SENSING BY USING ONE-DIMENSIONAL TERNARY PHOTONIC CRYSTALS [J].
Banerjee, A. .
PROGRESS IN ELECTROMAGNETICS RESEARCH-PIER, 2009, 89 :11-22
[2]   PHOTONIC GAPS IN THE DISPERSION OF SURFACE-PLASMONS ON GRATINGS [J].
BARNES, WL ;
PREIST, TW ;
KITSON, SC ;
SAMBLES, JR ;
COTTER, NPK ;
NASH, DJ .
PHYSICAL REVIEW B, 1995, 51 (16) :11164-11167
[3]   SURFACE PLASMON RESONANCES EFFECTS ON DIFFERENT PATTERNS OF SOLID-SILVER AND SILVER-SHELL NANOCYLINDRICAL PAIRS [J].
Chau, Y. -F. ;
Yeh, H. -H. ;
Tsai, D. P. .
JOURNAL OF ELECTROMAGNETIC WAVES AND APPLICATIONS, 2010, 24 (8-9) :1005-1014
[4]  
Hass G., 1963, AM I PHYSICS HDB, P6
[5]   Sensitivity enhancement based on application of multi-pass interferometry in phase-sensitive surface plasmon resonance biosensor [J].
Ho, H. P. ;
Yuan, W. ;
Wong, C. L. ;
Wu, S. Y. ;
Suen, Y. K. ;
Kong, S. K. ;
Lin, Chinlon .
OPTICS COMMUNICATIONS, 2007, 275 (02) :491-496
[6]   Real-time optical biosensor based on differential phase measurement of surface plasmon resonance [J].
Ho, HP ;
Law, WC ;
Wu, SY ;
Lin, CL ;
Kong, SK .
BIOSENSORS & BIOELECTRONICS, 2005, 20 (10) :2177-2180
[7]  
Horowitz P., 1989, The Art of Electronics, V2nd, P641
[8]   Wide dynamic range phase-sensitive surface plasmon resonance biosensor based on measuring the modulation harmonics [J].
Law, Wing-Cheung ;
Markowicz, Przemyslaw ;
Yong, Ken-Tye ;
Roy, Indrajit ;
Baev, Alexander ;
Patskovsky, Sergiy ;
Kabashin, Andrei V. ;
Ho-Pui, Ho ;
Prasad, Paras N. .
BIOSENSORS & BIOELECTRONICS, 2007, 23 (05) :627-632
[9]   Phase-sensitive time-modulated surface plasmon resonance polarimetry for wide dynamic range biosensing [J].
Markowicz, P. P. ;
Law, W. C. ;
Baev, A. ;
Prasad, P. N. ;
Patskovsky, S. ;
Kabashin, A. V. .
OPTICS EXPRESS, 2007, 15 (04) :1745-1754
[10]   High sensitivity surface plasmon resonance sensor based on phase detection [J].
Nelson, SG ;
Johnston, KS ;
Yee, SS .
SENSORS AND ACTUATORS B-CHEMICAL, 1996, 35 (1-3) :187-191