Comparisons between commercial salivary testosterone enzyme-linked immunosorbent assay kits

被引:6
作者
Andersson, Christoffer R. [1 ]
Bergquist, Jonas [2 ]
Theodorsson, Elvar [3 ,4 ]
Strom, Jakob O. [1 ,3 ,4 ]
机构
[1] Orebro Univ, Dept Neurol, Fac Med & Hlth, Orebro, Sweden
[2] Uppsala Univ, Analyt Chem & Neurochem, Dept Chem BMC, Uppsala, Sweden
[3] Linkoping Univ, Dept Clin Chem, Linkoping, Sweden
[4] Linkoping Univ, Dept Clin & Expt Med, Linkoping, Sweden
基金
瑞典研究理事会;
关键词
Androgens; immunoassay; methods; saliva; testosterone; TANDEM MASS-SPECTROMETRY; SERUM TESTOSTERONE; HYPOGONADISM; MEN; IMMUNOASSAYS;
D O I
10.1080/00365513.2017.1339231
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Introduction: Measuring testosterone concentrations is of interest both in clinical situations and for research, the latter expanding rapidly during recent years. An increased demand for convenient methods has prompted a number of companies to develop enzyme-linked immunosorbent assay (ELISA) kits to measure testosterone concentrations in saliva. However, the inter-comparability of kits from different manufacturers have yet to be determined. Aim of study: The aim of this study was to compare commercially available ELISA kits from four different manufacturers (Salimetrics, IBL, DRG and Demeditec). Methods: Saliva was collected from 50 participants (25 men and 25 women). Each sample was analysed by the four ELISA kits. Results: The correlations between the ELISA kits from Demeditec, DRG and Salimetrics were moderate to high with r-values >.77; however, proportional errors between the methods calls for caution. The ELISA kit from IBL malfunctioned and no results from this kit was obtained. Conclusions: Results from studies using the ELISA kits from Demeditec, DRG and Salimetrics are generally comparable; however, translation using the formulae presented in the current study could increase the accuracy of these comparisons.
引用
收藏
页码:582 / 586
页数:5
相关论文
共 17 条
[1]   Management of hypogonadism: is there a role for salivary testosterone [J].
Fiers, Tom ;
Kaufman, Jean-Marc .
ENDOCRINE, 2015, 50 (01) :1-3
[2]   A critical evaluation of salivary testosterone as a method for the assessment of serum testosterone [J].
Fiers, Tom ;
Delanghe, Joris ;
T'Sjoen, Guy ;
Van Caenegem, Eva ;
Wierckx, Katrien ;
Kaufman, Jean-Marc .
STEROIDS, 2014, 86 :5-9
[3]   The ''trouble'' with salivary testosterone [J].
Granger, DA ;
Shirtcliff, EA ;
Booth, A ;
Kivlighan, KT ;
Schwartz, EB .
PSYCHONEUROENDOCRINOLOGY, 2004, 29 (10) :1229-1240
[4]   Accuracy of First and Second Generation Testosterone Assays and Improvement through Sample Extraction [J].
Groenestege, Wouter M. Tiel ;
Bui, Hong N. ;
ten Kate, Joop ;
Menheere, Paul P. C. A. ;
Oosterhuis, Wytze P. ;
Vader, Huib L. ;
Heijboer, Annemieke C. ;
Janssen, Marcel J. W. .
CLINICAL CHEMISTRY, 2012, 58 (07) :1154-1156
[5]   Salivary testosterone measurement does not identify biochemical hypogonadism in aging men: a ROC analysis [J].
Hayes, Lawrence D. ;
Sculthorpe, Nicholas ;
Herbert, Peter ;
Baker, Julien S. ;
Hullin, David A. ;
Kilduff, Liam P. ;
Reed, Dewi ;
Spagna, Roberto ;
Grace, Fergal M. .
ENDOCRINE, 2015, 50 (01) :256-259
[6]   Salivary testosterone measurement by liquid chromatography tandem mass spectrometry in adult males and females [J].
Keevil, B. G. ;
MacDonald, P. ;
Macdowall, W. ;
Lee, D. M. ;
Wu, F. C. W. .
ANNALS OF CLINICAL BIOCHEMISTRY, 2014, 51 (03) :368-378
[7]   Cross-reactivity of steroid hormone immunoassays: clinical significance and two-dimensional molecular similarity prediction [J].
Krasowski, Matthew D. ;
Drees, Denny ;
Morris, Cory S. ;
Maakestad, Jon ;
Blau, John L. ;
Ekins, Sean .
BMC CLINICAL PATHOLOGY, 2014, 14
[8]   Low serum testosterone assayed by liquid chromatography-tandem mass spectrometry.: Comparison with five immunoassay techniques [J].
Moal, Valerie ;
Mathieu, Elisabeth ;
Reynier, Pascal ;
Malthiery, Yves ;
Gallois, Yves .
CLINICA CHIMICA ACTA, 2007, 386 (1-2) :12-19
[9]   Validation of salivary testosterone as a screening test for male hypogonadism [J].
Morley, John E. ;
Perry, H. Mitchell, III ;
Patrick, Ping ;
Dollbaum, Charles M. ;
Kells, John M. .
AGING MALE, 2006, 9 (03) :165-169
[10]  
PARDRIDGE WM, 1991, CLIN CHEM, V37, P139