Challenges in MR-only seed localization for postimplant dosimetry in permanent prostate brachytherapy

被引:10
作者
Zijlstra, Frank [1 ]
Moerland, Marinus A. [2 ]
van Zyp, Jochem R. N. van der Voort [2 ]
Noteboom, Juus L. [2 ]
Viergever, Max A. [1 ]
Seevinck, Peter R. [1 ]
机构
[1] Univ Med Ctr Utrecht, Image Sci Inst, Utrecht, Netherlands
[2] Univ Med Ctr Utrecht, Dept Radiotherapy, Utrecht, Netherlands
关键词
brachytherapy; MRI; seed localization; simulation; susceptibility; template matching; ESTRO/EAU/EORTC RECOMMENDATIONS; POSITIVE CONTRAST; IMPLANTATION; MARKER; FIELD; CT;
D O I
10.1002/mp.12505
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: An MR-only postimplant dosimetry workflow for low dose rate (LDR) brachytherapy could reduce patient burden, improve accuracy, and improve cost efficiency. However, localization of brachytherapy seeds on MRI scans remains a major challenge for this type of workflow. In this study, we propose and validate an MR-only seed localization method and identify remaining challenges. Methods and materials: The localization method was based on template matching of simulations of complex-valued imaging artifacts around metal brachytherapy seeds. The method was applied to MRI scans of 25 prostate cancer patients who underwent LDR brachytherapy and for whom postimplant dosimetry was performed after 4 weeks. The seed locations found with the MR-only method were validated against the seed locations found on CT. The circumstances in which detection errors were made were classified to gain an insight in the nature of the errors. Results: A total of 1490 of 1557 (96%) seeds were correctly detected, while 67 false-positive errors were made. The correctly detected seed locations had a high spatial accuracy with an average error of 0.8 mm compared with CT. A majority of the false positives occurred near other seeds. Most false negatives were found in either stranded configurations without spacers or near other seeds. Conclusions: The low detection error rate and high localization accuracy obtained by the complex-valued template matching approach are promising for future clinical application of MR-only dosimetry. The most important remaining challenge is robustness with regard to configurations of multiple seeds in close vicinity, such as in strands of seeds without spacers. This issue could potentially be resolved by simulating specific configurations of multiple seeds or by constraining the treatment planning to avoid these configurations, which could make the proposed method competitive with CT-based seed localization. (C) 2017 The Authors Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.
引用
收藏
页码:5051 / 5060
页数:10
相关论文
共 29 条
[1]   Phase correlation with sub-pixel accuracy: A comparative study in 1D and 2D [J].
Alba, Alfonso ;
Flavio Vigueras-Gomez, J. ;
Arce-Santana, Edgar R. ;
Aguilar-Ponce, Ruth M. .
COMPUTER VISION AND IMAGE UNDERSTANDING, 2015, 137 :76-87
[2]   ESTRO/EAU/EORTC recommendations on permanent seed implantation for localized prostate cancer [J].
Ash, D ;
Flynn, A ;
Battermann, J ;
de Reijke, T ;
Lavagnini, P ;
Blank, L .
RADIOTHERAPY AND ONCOLOGY, 2000, 57 (03) :315-321
[3]   ESUR prostate MR guidelines 2012 [J].
Barentsz, Jelle O. ;
Richenberg, Jonathan ;
Clements, Richard ;
Choyke, Peter ;
Verma, Sadhna ;
Villeirs, Geert ;
Rouviere, Olivier ;
Logager, Vibeke ;
Futterer, Jurgen J. .
EUROPEAN RADIOLOGY, 2012, 22 (04) :746-757
[4]   Alias subtraction more efficient than conventional zero-padding in the Fourier-based calculation of the susceptibility induced perturbation of the magnetic field in MR [J].
Bouwman, Job G. ;
Bakker, Chris J. G. .
MAGNETIC RESONANCE IN MEDICINE, 2012, 68 (02) :621-630
[5]   Improving prostate brachytherapy quality assurance with MRI-CT fusion based sector analysis in a phase II prospective trial of men with intermediate-risk prostate cancer [J].
Brown, Aaron P. ;
Pugh, Thomas J. ;
Swanson, David A. ;
Kudchadker, Rajat J. ;
Bruno, Teresa L. ;
Christensen, Eva N. ;
van Vulpen, Marco ;
Frank, Steven J. .
BRACHYTHERAPY, 2013, 12 (05) :401-407
[6]   Prostate brachytherapy postimplant dosimetry: Automatic plan reconstruction of stranded implants [J].
Chng, N. ;
Spadinger, I. ;
Morris, W. J. ;
Usmani, N. ;
Salcudean, S. .
MEDICAL PHYSICS, 2011, 38 (01) :327-342
[7]   Accuracy of seed reconstruction in prostate postplanning studied with a CT- and MRI-compatible phantom [J].
De Brabandere, Marisol ;
Kirisits, Christian ;
Peeters, Ronald ;
Haustermans, Karin ;
Van den Heuvel, Frank .
RADIOTHERAPY AND ONCOLOGY, 2006, 79 (02) :190-197
[8]   CT- and MRI-based seed localization in postimplant evaluation after prostate brachytherapy [J].
De Brabandere, Marisol ;
Al-Qaisieh, Bashar ;
De Wever, Liesbeth ;
Haustermans, Karin ;
Kirisits, Christian ;
Moerland, Marinus A. ;
Oyen, Raymond ;
Rijnders, Alex ;
Van den Heuvel, Frank ;
Siebert, Frank-Andre .
BRACHYTHERAPY, 2013, 12 (06) :580-588
[9]   MEASURES OF THE AMOUNT OF ECOLOGIC ASSOCIATION BETWEEN SPECIES [J].
DICE, LR .
ECOLOGY, 1945, 26 (03) :297-302
[10]   Susceptibility-based positive contrast MRI of brachytherapy seeds [J].
Dong, Ying ;
Chang, Zheng ;
Xie, Guoxi ;
Whitehead, Gregory ;
Ji, Jim X. .
MAGNETIC RESONANCE IN MEDICINE, 2015, 74 (03) :716-726