Current advances and future trends in X-ray digital detectors for medical applications

被引:11
作者
Rowlands, JA [1 ]
机构
[1] Univ Toronto, Sunnybrook & Womens Coll, Hlth Sci Ctr, Toronto, ON M4N 3M5, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
biomedical X-ray imaging; thin-film devices; photoconducting materials;
D O I
10.1109/19.746705
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Despite the progress made in imaging modalities based on nonionizing radiation such as ultrasound and magnetic resonance imaging, X-ray is still the primary imaging method in most radiology departments. In recent years, digital electronics and manufacturing methods have led to many approaches for the design and construction of X-ray detectors. The ideal medical X-ray system is a digital, Bat-panel able to perform all current radiological modalities at reduced dose. It would, immediately after the patient's X-ray exposure, provide a high quality radiograph on a video monitor. It would also be usable for real-time imaging (e.g., digital cine loops and fluoroscopy), The physical form would be similar to a film/screen cassette (except for the addition of an umbilical cord) which could easily fit into the Bucky tray of existing X-ray rooms, There are two main approaches both based on active matrix technology developed for lap-top displays. The first, called the direct method employs a photoconductive layer of amorphous selenium (a-Se) to detect X-rays, The charge image formed on the surface of the a-Se is digitally read out in situ using a two-dimensional array of thin-film transistors (TFT's), i.e., an active matrix. The second flat panel method is called the indirect method and uses an active matrix to readout a phosphor layer. Phosphors give off light on interaction with X-rays and this is sensed by photo diodes on the active matrix, The relative practicality of the direct and indirect active matrix readout systems is an area of current investigation.
引用
收藏
页码:1415 / 1418
页数:4
相关论文
共 15 条
[1]   DEMONSTRATION OF MEGAVOLTAGE AND DIAGNOSTIC-X-RAY IMAGING WITH HYDROGENATED AMORPHOUS-SILICON ARRAYS [J].
ANTONUK, LE ;
BOUDRY, J ;
HUANG, WD ;
MCSHAN, DL ;
MORTON, EJ ;
YORKSTON, J ;
LONGO, MJ ;
STREET, RA .
MEDICAL PHYSICS, 1992, 19 (06) :1455-1466
[2]   RADIOGRAPHIC INFORMATION-THEORY AND APPLICATION TO MAMMOGRAPHY [J].
BRODIE, I ;
GUTCHECK, RA .
MEDICAL PHYSICS, 1982, 9 (01) :79-95
[3]   IMAGE FEATURE ANALYSIS AND COMPUTER-AIDED DIAGNOSIS IN DIGITAL RADIOGRAPHY .1. AUTOMATED DETECTION OF MICROCALCIFICATIONS IN MAMMOGRAPHY [J].
CHAN, HP ;
DOI, K ;
GALHOTRA, S ;
VYBORNY, CJ ;
MACMAHON, H ;
JOKICH, PM .
MEDICAL PHYSICS, 1987, 14 (04) :538-548
[4]   HIGH-SENSITIVITY READOUT OF 2D A-SI IMAGE SENSORS [J].
FUJIEDA, I ;
STREET, RA ;
WEISFIELD, RL ;
NELSON, S ;
NYLEN, P ;
PEREZMENDEZ, V ;
CHO, GS .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1993, 32 (1A) :198-204
[5]  
KASAP SO, 1997, FUTURE DIRECTIONS TH
[6]   A STUDY OF PHOTOELECTRONIC DIGITAL RADIOLOGY .1. THE PHOTOELECTRONIC DIGITAL RADIOLOGY DEPARTMENT [J].
NUDELMAN, S ;
FISHER, HD ;
FROST, MM ;
CAPP, MP ;
OVITT, TW .
PROCEEDINGS OF THE IEEE, 1982, 70 (07) :700-707
[7]  
Powell M.J., 1992, MATER RES SOC S P, V258, P1127
[8]  
ROUGEOT H, 1993, AAPM MONOGRAPH, V22, P49
[9]   SYSTEM FOR DIGITAL ACQUISITION OF GASTROINTESTINAL IMAGES [J].
ROWLANDS, JA ;
HYNES, DM ;
EDMONDS, EW .
MEDICAL PHYSICS, 1989, 16 (04) :553-560
[10]   Digital X-ray systems based on amorphous selenium [J].
Rowlands, JA .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 167 (02) :409-411