Polyethylene wear of dual mobility cups: a comparative analysis based on patient-specific finite element modeling

被引:10
|
作者
Wegrzyn, Julien [1 ]
Antoniadis, Alexander [1 ]
Sarshari, Ehsan [2 ]
Boubat, Matthieu [2 ]
Terrier, Alexandre [1 ,2 ]
机构
[1] Ctr Hosp Univ Vaudois CHUV, Lausanne Univ Hosp, Dept Orthoped Surg, Hop Orthoped, Ave Pierre Decker 4, CH-1011 Lausanne, Switzerland
[2] Ecole Polytech Fed Lausanne, Lab Biomech Orthoped, Lausanne, Switzerland
关键词
Dual mobility cup; Polyethylene mobile component; Patient-specific modeling; Polyethylene wear; Highly cross-linked polyethylene; TOTAL HIP-ARTHROPLASTY; DISLOCATION; LINERS; SIMULATOR; BEARINGS; FRICTION;
D O I
10.1007/s00264-022-05305-4
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose Concerns remain about potential increased wear with dual mobility cups related to the multiple articulations involved in this specific design of implant. This finite element analysis study aimed to compare polyethylene (PE) wear between dual mobility cup and conventional acetabular component, and between the use of conventional ultra-high molecular weight PE (UHMWPE) and highly cross-linked PE (XPLE). Methods Patient-specific finite element modeling was developed for 15 patients undergoing primary total hip arthroplasty (THA). Five acetabular components were 3D modeled and compared in THA constructs replicating existing implants: a dual mobility cup with a 22.2-mm-diameter femoral head against UHMWPE or XLPE (DM22PE or DM22XL), a conventional cup with a 22.2-mm-diameter femoral head against UHMWPE (SD22PE) and a conventional cup with a 32-mm-diameter femoral head against UHMWPE or XLPE (SD32PE or SD32XL). Results DM22PE produced 4.6 times and 5.1 times more volumetric wear than SD32XL and DM22XL (p < 0.0001, Cohen's d = 6.97 and 7.11; respectively). However, even if significant, the differences in volumetric wear between DM22XL and SD32XL as well as between DM22PE and SD22PE or SD32PE were small according to their effect size (p < 0.0001, Cohen's vertical bar d vertical bar= 0.48 to 0.65) and could be therefore considered as clinically negligible. Conclusion When using XLPE instead of UHMWPE, dual mobility cup with a 22.2-mm-diameter femoral head produced a similar amount of volumetric wear than conventional acetabular component with a 32-mm-diameter femoral head against XLPE. Therefore, XLPE is advocated in dual mobility cup to improve its wear performance.
引用
收藏
页码:779 / 787
页数:9
相关论文
共 50 条
  • [41] Patient-specific three-dimensional evaluation of interface micromotion in two different short stem designs in cementless total hip arthroplasty: a finite element analysis
    Arata Kanaizumi
    Daisuke Suzuki
    Satoshi Nagoya
    Atsushi Teramoto
    Toshihiko Yamashita
    Journal of Orthopaedic Surgery and Research, 17
  • [42] Patient-specific three-dimensional evaluation of interface micromotion in two different short stem designs in cementless total hip arthroplasty: a finite element analysis
    Kanaizumi, Arata
    Suzuki, Daisuke
    Nagoya, Satoshi
    Teramoto, Atsushi
    Yamashita, Toshihiko
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2022, 17 (01)
  • [43] Effects of Surface Modification and Bulk Geometry on the Biotribological Behavior of Cross-Linked Polyethylene: Wear Testing and Finite Element Analysis
    Watanabe, Kenichi
    Kyomoto, Masayuki
    Saiga, Kenichi
    Taketomi, Shuji
    Inui, Hiroshi
    Kadono, Yuho
    Takatori, Yoshio
    Tanaka, Sakae
    Ishihara, Kazuhiko
    Moro, Toru
    BIOMED RESEARCH INTERNATIONAL, 2015, 2015
  • [44] Potential of computational models in personalized treatment of obstructive sleep apnea: a patient-specific partial 3D finite element study
    Ayyalasomayajula, Venkat
    Moxness, Mads
    Skallerud, Bjorn
    BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2024, 23 (02) : 507 - 524
  • [45] Patient-specific finite element analyses detect significant mechanical therapeutic effects on osteoporotic vertebrae during a three-year treatment
    Tawara D.
    Sakamoto J.
    Murakami H.
    Kawahara N.
    Tomita K.
    Journal of Biomechanical Science and Engineering, 2011, 6 (04): : 248 - 261
  • [46] A Modeling Study of a Patient-specific Safe Zone for THA: Calculation, Validation, and Key Factors Based on Standing and Sitting Sagittal Pelvic Tilt
    Tang, Hao
    Li, Ya
    Zhou, Yixin
    Wang, Siyuang
    Zhao, Yongqiang
    Ma, Zhuyi
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2022, 480 (01) : 191 - 205
  • [47] Potential of computational models in personalized treatment of obstructive sleep apnea: a patient-specific partial 3D finite element study
    Venkat Ayyalasomayajula
    Mads Moxness
    Bjørn Skallerud
    Biomechanics and Modeling in Mechanobiology, 2024, 23 : 507 - 524
  • [48] Modeling Atrial Fiber Orientation in Patient-Specific Geometries: A Semi-automatic Rule-Based Approach
    Krueger, Martin W.
    Schmidt, Viktor
    Tobon, Catalina
    Weber, Frank M.
    Lorenz, Cristian
    Keller, David U. J.
    Barschdorf, Hans
    Burdumy, Michael
    Neher, Peter
    Plank, Gernot
    Rhode, Kawal
    Seemann, Gunnar
    Sanchez-Quintana, Damien
    Saiz, Javier
    Razavi, Reza
    Doessel, Olaf
    FUNCTIONAL IMAGING AND MODELING OF THE HEART, 2011, 6666 : 223 - 232
  • [49] CORR Insights®: A Modeling Study of a Patient-specific Safe Zone for THA: Calculation, Validation, and Key Factors Based on Standing and Sitting Sagittal Pelvic Tilt
    Ebramzadeh, Edward
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2022, 480 (01) : 206 - 208
  • [50] A New Analysis Method of the Dry Sliding Wear Process Based on the Low Cycle Fatigue Theory and the Finite Element Method
    Abdi, Mohammad
    Taheri, Ali Karimi
    Bakhtiarydavijani, Amirhamed
    JOURNAL OF MATERIALS ENGINEERING AND PERFORMANCE, 2014, 23 (03) : 1096 - 1106