Measurement error due to self-absorption in calibration-free laser-induced breakdown spectroscopy

被引:19
|
作者
Taleb, Aya [1 ,2 ]
Motto-Ros, Vincent [3 ]
Carru, Mauro J. [1 ]
Axente, Emanuel [4 ]
Craciun, Valentin [4 ]
Pelascini, Frederic [2 ]
Hermann, Jorg [1 ]
机构
[1] Aix Marseille Univ, CNRS, LP3, F-13288 Marseille, France
[2] Cetim Grand Est, F-67400 Illkirch Graffenstaden, France
[3] Univ Lyon 1, CNRS, Inst Lumiere Matiere, F-69622 Villeurbanne, France
[4] Natl Inst Laser Plasma & Radiat Phys, Magurele 77125, Romania
基金
欧盟地平线“2020”;
关键词
Elemental analysis; Laser-induced breakdown spectroscopy; Calibration-free; Self-absorption; Error evaluation; INDUCED PLASMA; GROWTH;
D O I
10.1016/j.aca.2021.339070
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Self-absorption of spectral lines is known to lower the performance of analytical measurements via calibration-free laser-induced breakdown spectroscopy. However, the error growth due to this effect is not clearly assessed. Here we propose a method to quantify the measurement error due to self-absorption based on the calculation of the spectral radiance of a plasma in local thermodynamic equilibrium. Validated through spectroscopic measurements for a binary alloy thin film of compositional gradient, the method evidences that measurement performance lowering due to self-absorption depends on the spectral shape of the analytical transition and on the intensity measurement method. Thus, line-integrated intensity measurements of Stark broadened lines enable accurate analysis, even at large optical thickness, if line width and plasma size are precisely known. The error growth due to self-absorption is significantly larger for line shapes dominated by Doppler broadening and for line-center intensity measurements. The findings present a significant advance in compositional measurements via calibration-free laser-induced breakdown spectroscopy, as they enable straightforward selection of most appropriate analytical lines. (c) 2021 Elsevier B.V. All rights reserved.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] A method for improving the accuracy of calibration-free laser-induced breakdown spectroscopy by exploiting self-absorption
    Hu, Zhenlin
    Chen, Feng
    Zhang, Deng
    Chu, Yanwu
    Wang, Weiliang
    Tang, Yun
    Guo, Lianbo
    ANALYTICA CHIMICA ACTA, 2021, 1183
  • [2] Correction of self-absorption effect in calibration-free laser-induced breakdown spectroscopy by an internal reference method
    Sun, Lanxiang
    Yu, Haibin
    TALANTA, 2009, 79 (02) : 388 - 395
  • [3] Self-absorption effect in calibration-free laser-induced breakdown spectroscopy: Analysis of mineral profile in Maerua oblongifolia plant
    Qasim, M.
    Anwar-ul-Haq, M.
    Shah, Amin
    Afgan, M. Sher
    Haq, S. U.
    Khan, Rameez Abbas
    Baig, M. Aslam
    MICROCHEMICAL JOURNAL, 2022, 175
  • [4] Correction of self-absorption effect in calibration-free laser-induced breakdown spectroscopy (CF-LIBS) with blackbody radiation reference
    Li, Tianqi
    Hou, Zongyu
    Fu, Yangting
    Yu, Jianlong
    Gu, Weilun
    Wang, Zhe
    ANALYTICA CHIMICA ACTA, 2019, 1058 : 39 - 47
  • [5] Progress in calibration-free laser-induced breakdown spectroscopy
    Hermann, Jorg
    Gerhard, Christoph
    Burger, Milos
    Craciun, Valentin
    Pelascini, Frederic
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2023, 200
  • [6] A review of calibration-free laser-induced breakdown spectroscopy
    Hu, Zhenlin
    Zhang, Deng
    Wang, Weiliang
    Chen, Feng
    Xu, Yubin
    Nie, Junfei
    Chu, Yanwu
    Guo, Lianbo
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2022, 152
  • [7] A review of calibration-free laser-induced breakdown spectroscopy
    Hu, Zhenlin
    Zhang, Deng
    Wang, Weiliang
    Chen, Feng
    Xu, Yubin
    Nie, Junfei
    Chu, Yanwu
    Guo, Lianbo
    TrAC - Trends in Analytical Chemistry, 2022, 152
  • [8] Calibration-free laser-induced plasma analysis of a metallic alloy with self-absorption correction
    Hamidreza Shirvani-Mahdavi
    Seyede Zahra Shoursheini
    Hamid Gholami
    Ziba Dini-Torkamani
    Sare Ghahari-Korani
    Applied Physics B, 2014, 117 : 823 - 832
  • [9] Calibration-free laser-induced plasma analysis of a metallic alloy with self-absorption correction
    Shirvani-Mahdavi, Hamidreza
    Shoursheini, Seyede Zahra
    Gholami, Hamid
    Dini-Torkamani, Ziba
    Ghahari-Korani, Sare
    APPLIED PHYSICS B-LASERS AND OPTICS, 2014, 117 (03): : 823 - 832
  • [10] Error analysis in optimization problems relevant for calibration-free laser-induced breakdown spectroscopy
    Maali, J. J.