Development and interinstitutional validation of an automatic vertebral-body misalignment error detector for cone-beam CT-guided radiotherapy

被引:6
|
作者
Luximon, Dishane C. [1 ]
Ritter, Timothy [2 ]
Fields, Emma [2 ]
Neylon, John [1 ]
Petragallo, Rachel [1 ]
Abdulkadir, Yasin [1 ]
Charters, John [1 ]
Low, Daniel A. [1 ]
Lamb, James M. [1 ]
机构
[1] Univ Calif Los Angeles, David Geffen Sch Med, Dept Radiat Oncol, 200 UCLA Med Plaza Suite B265,Box 956951, Los Angeles, CA 90095 USA
[2] Virginia Commonwealth Univ, Dept Med Phys, Richmond, VA USA
基金
美国医疗保健研究与质量局;
关键词
deep learning; patient safety; radiation therapy; SURFACE IMAGING-SYSTEM; PATIENT IDENTIFICATION; RADIATION; REGISTRATION; AUDITS;
D O I
10.1002/mp.15927
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background In cone-beam computed tomography (CBCT)-guided radiotherapy, off-by-one vertebral-body misalignments are rare but serious errors that lead to wrong-site treatments. Purpose An automatic error detection algorithm was developed that uses a three-branch convolutional neural network error detection model (EDM) to detect off-by-one vertebral-body misalignments using planning computed tomography (CT) images and setup CBCT images. Methods Algorithm training and test data consisted of planning CTs and CBCTs from 480 patients undergoing radiotherapy treatment in the thoracic and abdominal regions at two radiotherapy clinics. The clinically applied registration was used to derive true-negative (no error) data. The setup and planning images were then misaligned by one vertebral-body in both the superior and inferior directions, simulating the most likely misalignment scenarios. For each of the aligned and misaligned 3D image pairs, 2D slice pairs were automatically extracted in each anatomical plane about a point within the vertebral column. The three slice pairs obtained were then inputted to the EDM that returned a probability of vertebral misalignment. One model (EDM1) was trained solely on data from institution 1. EDM1 was further trained using a lower learning rate on a dataset from institution 2 to produce a fine-tuned model, EDM2. Another model, EDM3, was trained from scratch using a training dataset composed of data from both institutions. These three models were validated on a randomly selected and unseen dataset composed of images from both institutions, for a total of 303 image pairs. The model performances were quantified using a receiver operating characteristic analysis. Due to the rarity of vertebral-body misalignments in the clinic, a minimum threshold value yielding a specificity of at least 99% was selected. Using this threshold, the sensitivity was calculated for each model, on each institution's test set separately. Results When applied to the combined test set, EDM1, EDM2, and EDM3 resulted in an area under curve of 99.5%, 99.4%, and 99.5%, respectively. EDM1 achieved a sensitivity of 96% and 88% on Institution 1 and Institution 2 test set, respectively. EDM2 obtained a sensitivity of 95% on each institution's test set. EDM3 achieved a sensitivity of 95% and 88% on Institution 1 and Institution 2 test set, respectively. Conclusion The proposed algorithm demonstrated accuracy in identifying off-by-one vertebral-body misalignments in CBCT-guided radiotherapy that was sufficiently high to allow for practical implementation. It was found that fine-tuning the model on a multi-facility dataset can further enhance the generalizability of the algorithm.
引用
收藏
页码:6410 / 6423
页数:14
相关论文
共 33 条
  • [31] Retrospective dose reconstruction of prostate stereotactic body radiotherapy using cone-beam CT and a log file during VMAT delivery with flattening-filter-free mode
    Imae, Toshikazu
    Haga, Akihiro
    Watanabe, Yuichi
    Takenaka, Shigeharu
    Shiraki, Takashi
    Nawa, Kanabu
    Ogita, Mami
    Takahashi, Wataru
    Yamashita, Hideomi
    Nakagawa, Keiichi
    Abe, Osamu
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2020, 13 (03) : 238 - 248
  • [32] Stereotactic body radiotherapy for spine and bony pelvis using flattening filter free volumetric modulated arc therapy, 6D cone-beam CT and simple positioning techniques: Treatment time and patient stability
    Dahele, Max
    Slotman, Ben
    Verbakel, Wilko
    ACTA ONCOLOGICA, 2016, 55 (06) : 795 - 798
  • [33] Rotational positional error-corrected linear set-up margin calculation technique for lung stereotactic body radiotherapy in a dual imaging environment of 4-D cone beam CT and ExacTrac stereoscopic imaging
    Sarkar, Biplab
    Ganesh, Tharmarnadar
    Munshi, Anusheel
    Manikandan, Arjunan
    Roy, Soumya
    Krishnankutty, Saneg
    Chitral, Latika
    Sathiya, Jean
    Pradhan, Anirudh
    Mohanti, Bidhu Kalyan
    RADIOLOGIA MEDICA, 2021, 126 (07): : 979 - 988