Revision Total Hip Arthroplasty Using the Modular Proximal Stem Modified for Asians: Medium-Term Clinical Results and Perioperative Complications

被引:2
作者
Abe, Toshiomi [1 ]
Otani, Takuya [2 ]
Hayama, Tetsuo [3 ]
Fujii, Hideki [3 ]
Kawaguchi, Yasuhiko [2 ]
Saito, Mitsuru [3 ]
机构
[1] Jikei Univ, KASHIWA Hosp, Dept Orthopaed Surg, 163-1 Kashiwashita, Kashiwa, Chiba 2770004, Japan
[2] Jikei Univ, DAISAN Hosp, Dept Orthopaed Surg, Tokyo, Japan
[3] Jikei Univ, Sch Med, Dept Orthopaed Surg, Tokyo, Japan
关键词
revision total hip arthroplasty; modular stem; complication; trunnionosis; adverse reaction to metal debris (ARMD); FEMORAL COMPONENT; REPLACEMENT; FIXATION; FAILURE; RETRIEVAL; CORROSION;
D O I
10.1016/j.arth.2021.12.037
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: This study aimed to examine the medium-term clinical and radiological outcomes of revision THA using the S-ROM-A stem, a modification of the S-ROM stem intended for Asians. Methods: Femoral reconstruction using the S-ROM-A stem was performed in 126 hips that underwent revision THA. All patients were followed for perioperative complications. In addition, clinical and radiographic outcomes at a mean of 8 (range 5-14) years postoperatively were evaluated in 96 hips of 86 patients (76%). Results: The most common perioperative complication was a femoral fracture, occurring in 16 hips (13%), including 11 intraoperative and 5 postoperative fractures. Dislocation occurred in five hips (4.0%), infection in three hips (2.4%), and trunnionosis in two hips (1.6%), including late complications. The total second stem revision was performed in two hips while stem only second revision preserving the bone ingrown sleeve was performed in four hips. With a second revision for aseptic loosening as the endpoint, the 13-year stem survival rate was 100%. Hip function as assessed by the Japanese Orthopedic Association score improved from a mean of 48 points preoperatively to 87 points 8 years postoperatively (P < .05). Radiological evaluation at the final follow-up showed that 95 hips (99%) achieved bone ingrowth fixation and one hip (1%) achieved fibrous stable status. Conclusion: Revision THA using the S-ROM-A stem resulted in good medium-term outcomes. Although modifications of the stem length and shape may be effective in preventing fractures in Asians with relatively small body sizes, attention should be paid to the occurrence of trunnionosis, which may be associated with the decreased taper size. (c) 2021 Elsevier Inc. All rights reserved.
引用
收藏
页码:770 / 776
页数:7
相关论文
共 34 条
[21]   NONPARAMETRIC-ESTIMATION FROM INCOMPLETE OBSERVATIONS [J].
KAPLAN, EL ;
MEIER, P .
JOURNAL OF THE AMERICAN STATISTICAL ASSOCIATION, 1958, 53 (282) :457-481
[22]   Short-Term Results of the S-ROM-A Femoral Prosthesis Operative Strategies for Asian Patients With Osteoarthritis [J].
Kido, Kensuke ;
Fujioka, Mikihiro ;
Takahashi, Kenji ;
Ueshima, Keiichiro ;
Goto, Tsuyoshi ;
Kubo, Toshikazu .
JOURNAL OF ARTHROPLASTY, 2009, 24 (08) :1193-1199
[23]   Catastrophic Femoral Head-Stem Trunnion Dissociation Secondary to Corrosion [J].
Ko, Laura Matsen ;
Chen, Antonia F. ;
Deirmengian, Gregory K. ;
Hozack, William J. ;
Sharkey, Peter F. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2016, 98 (16) :1400-1404
[24]   Higher incidence of aseptic loosening caused by a lower canal filling ratio with a modified modular stem in total hip arthroplasty [J].
Kobayashi, Kyosuke ;
Kidera, Kenichi ;
Itose, Masaru ;
Motokawa, Tetsuhiko ;
Chiba, Ko ;
Osaki, Makoto .
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2020, 15 (01)
[25]   Proximal Component Modularity in THA-At What Cost?: An Implant Retrieval Study [J].
Kop, A. M. ;
Keogh, C. ;
Swarts, E. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (07) :1885-1894
[26]   Taper junction failure in large-diameter metal-on-metal bearings [J].
Langton, D. J. ;
Sidaginamale, R. ;
Lord, J. K. ;
Nargol, A. V. F. ;
Joyce, T. J. .
BONE & JOINT RESEARCH, 2012, 1 (04) :56-63
[27]   Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement A CONSEQUENCE OF EXCESS WEAR [J].
Langton, D. J. ;
Jameson, S. S. ;
Joyce, T. J. ;
Hallab, N. J. ;
Natu, S. ;
Nargol, A. V. F. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2010, 92B (01) :38-46
[28]   Trunnion-Head Stresses in THA: Are Big Heads Trouble? [J].
Lavernia, Carlos J. ;
Iacobelli, David A. ;
Villa, Jesus M. ;
Jones, Kinzy ;
Gonzalez, Jose L. ;
Jones, William Kinzy .
JOURNAL OF ARTHROPLASTY, 2015, 30 (06) :1085-1088
[29]  
McCarthy JC, 1993, ORTHOP T, V17, P966
[30]  
Ogawa R, 1995, NIPPON SEIKEIGEKA GA, V69, P860