Comparison of Self-Expanding and Mechanically Expanded Transcatheter Aortic Valve Prostheses

被引:38
作者
Gooley, Robert P. [1 ,2 ]
Talman, Andrew H. [1 ,2 ]
Cameron, James D. [1 ,2 ]
Lockwood, Siobhan M. [1 ,2 ]
Meredith, Ian T. [1 ,2 ]
机构
[1] Monash Hlth, MonashHeart, Melbourne, Vic 3168, Australia
[2] Monash Univ, Monash Cardiovasc Res Ctr, Clayton, Vic, Australia
基金
英国医学研究理事会;
关键词
aortic stenosis; CoreValve; Lotus; transcatheter aortic valve replacement; LOTUS VALVE; PARAVALVULAR REGURGITATION; VASCULAR COMPLICATIONS; IMPLANTATION IMPACT; EDWARDS-SAPIEN; RISK PATIENTS; REPLACEMENT; COREVALVE; STENOSIS; OUTCOMES;
D O I
10.1016/j.jcin.2015.03.014
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES The aim of this study was to determine whether transcatheter aortic valve replacement (TAVR) with the mechanically expanded Lotus valve (Boston Scientific, Natick Massachusetts) offers potential benefits over treatment with the self-expanding CoreValve (Medtronic, Minneapolis, Minnesota). BACKGROUND New-generation transcatheter aortic valve systems are emerging in clinical trials and practice with design features aimed at improving safety and efficacy. To date, these devices have not been compared systematically with current-generation devices. METHODS A total of 100 patients (83.4 +/- 4.8 years of age, 44% male, Society of Thoracic Surgeons Predicted Risk of Mortality score of 5.5 +/- 2.4) were assessed. Fifty consecutive patients undergoing a Lotus transcatheter aortic valve replacement were enrolled and compared with 50 matched patients treated with a CoreValve. An independent core laboratory reviewed all echocardiographic data, and an independent clinical events committee adjudicated all events. RESULTS Valve Academic Research Consortium 2-defined device success was 84% and 64% in the Lotus and CoreValve cohorts, respectively (p = 0.02). This difference was driven by lower rates of moderate or greater aortic regurgitation (4% vs. 16.7%, respectively; p = 0.04) and higher rates of successfully implanting a single device in the correct anatomic position (100% vs. 86%, respectively; p = 0.06). Cardiovascular mortality rate (0% vs. 4%, respectively; p = 0.32), major stroke rate (4% vs. 2%, respectively; p = 0.56), and permanent pacemaker insertion rate (28% vs. 18%, respectively; p = 0.23) were not different at 30 days in the Lotus and CoreValve cohorts. CONCLUSIONS In this matched comparison of high surgical risk patients undergoing transcatheter aortic valve replacement, the use of the Lotus device was associated with higher rates of Valve Academic Research Consortium 2-defined device success compared with the CoreValve. This was driven by higher rates of correct anatomic positioning and lower incidences of moderate paraprosthetic regurgitation. The clinical significance of these differences needs to be tested in a large randomized, controlled trial. (C) 2015 by the American College of Cardiology Foundation.
引用
收藏
页码:962 / 971
页数:10
相关论文
共 40 条
[1]   Comparison of Balloon-Expandable vs Self-expandable Valves in Patients Undergoing Transcatheter Aortic Valve Replacement The CHOICE Randomized Clinical Trial [J].
Abdel-Wahab, Mohamed ;
Mehilli, Julinda ;
Frerker, Christian ;
Neumann, Franz-Josef ;
Kurz, Thomas ;
Toelg, Ralph ;
Zachow, Dirk ;
Guerra, Elena ;
Massberg, Steffen ;
Schaefer, Ulrich ;
El-Mawardy, Mohamed ;
Richardt, Gert .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2014, 311 (15) :1503-1514
[2]   Transcatheter Aortic-Valve Replacement with a Self-Expanding Prosthesis [J].
Adams, David H. ;
Popma, Jeffrey J. ;
Reardon, Michael J. ;
Yakubov, Steven J. ;
Coselli, Joseph S. ;
Deeb, G. Michael ;
Gleason, Thomas G. ;
Buchbinder, Maurice ;
Hermiller, James, Jr. ;
Kleiman, Neal S. ;
Chetcuti, Stan ;
Heiser, John ;
Merhi, William ;
Zorn, George ;
Tadros, Peter ;
Robinson, Newell ;
Petrossian, George ;
Hughes, G. Chad ;
Harrison, J. Kevin ;
Conte, John ;
Maini, Brijeshwar ;
Mumtaz, Mubashir ;
Chenoweth, Sharla ;
Oh, Jae K. .
NEW ENGLAND JOURNAL OF MEDICINE, 2014, 370 (19) :1790-1798
[3]   Incidence, Predictors, and Outcomes of Aortic Regurgitation After Transcatheter Aortic Valve Replacement Meta-Analysis and Systematic Review of Literature [J].
Athappan, Ganesh ;
Patvardhan, Eshan ;
Tuzcu, E. Murat ;
Svensson, Lars Georg ;
Lemos, Pedro A. ;
Fraccaro, Chiara ;
Tarantini, Giuseppe ;
Sinning, Jan-Malte ;
Nickenig, Georg ;
Capodanno, Davide ;
Tamburino, Corrado ;
Latib, Azeem ;
Colombo, Antonio ;
Kapadia, Samir R. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 61 (15) :1585-1595
[4]   3-year outcomes of self-expanding Corevalve prosthesis - The Italian Registry [J].
Barbanti, Marco ;
Ussia, Gian Paolo ;
Cannata, Stefano ;
Giarratana, Alessandra ;
Sgroi, Carmelo ;
Tamburino, Corrado .
ANNALS OF CARDIOTHORACIC SURGERY, 2012, 1 (02) :182-184
[5]   Echocardiographic Assessment of Valve Stenosis: EAE/ASE Recommendations for Clinical Practice [J].
Baumgartner, Helmut ;
Hung, Judy ;
Bermejo, Javier ;
Chambers, John B. ;
Evangelista, Arturo ;
Griffin, Brian P. ;
Iung, Bernard ;
Otto, Catherine M. ;
Pellikka, Patricia A. ;
Quinones, Miguel .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2009, 22 (01) :1-23
[6]   Transcatheter Aortic Valve Replacement With the SAPIEN 3 A New Balloon-Expandable Transcatheter Heart Valve [J].
Binder, Ronald K. ;
Rodes-Cabau, Josep ;
Wood, David A. ;
Mok, Michael ;
Leipsic, Jonathon ;
De Larochelliere, Robert ;
Toggweiler, Stefan ;
Dumont, Eric ;
Freeman, Melanie ;
Willson, Alexander B. ;
Webb, John G. .
JACC-CARDIOVASCULAR INTERVENTIONS, 2013, 6 (03) :293-300
[7]   A Practical Guide to Multimodality Imaging of Transcatheter Aortic Valve Replacement [J].
Bloomfield, Gerald S. ;
Gillam, Linda D. ;
Hahn, Rebecca T. ;
Kapadia, Samir ;
Leipsic, Jonathon ;
Lerakis, Stamatios ;
Tuzcu, Murat ;
Douglas, Pamela S. .
JACC-CARDIOVASCULAR IMAGING, 2012, 5 (04) :441-455
[8]   Transesophageal echocardiographic scoring for transcatheter aortic valve implantation: Impact of aortic cusp calcification on postoperative aortic regurgitation [J].
Colli, Andrea ;
D'Amico, Roberto ;
Kempfert, Joerg ;
Borger, Michael A. ;
Mohr, Friedrich W. ;
Walther, Thomas .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2011, 142 (05) :1229-1235
[9]   Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis - First human case description [J].
Cribier, A ;
Eltchaninoff, H ;
Bash, A ;
Borenstein, N ;
Tron, C ;
Bauer, F ;
Derumeaux, G ;
Anselme, F ;
Laborde, F ;
Leon, MB .
CIRCULATION, 2002, 106 (24) :3006-3008
[10]   Determinants of Significant Paravalvular Regurgitation After Transcatheter Aortic Valve Implantation Impact of Device and Annulus Discongruence [J].
Detaint, Delphine ;
Lepage, Laurent ;
Himbert, Dominique ;
Brochet, Eric ;
Messika-Zeitoun, David ;
Iung, Bernard ;
Vahanian, Alec .
JACC-CARDIOVASCULAR INTERVENTIONS, 2009, 2 (09) :821-827