Dry-reagent microfluidic biosensor for simple detection of NT-proBNP via Ag nanoparticles

被引:21
作者
Beck, Franziska [1 ]
Horn, Carina [2 ]
Baeumner, Antje J. [1 ]
机构
[1] Univ Regensburg, Inst Analyt Chem Chemo & Biosensors, D-93043 Regensburg, Germany
[2] Roche Diagnost, D-68305 Mannheim, Germany
关键词
Electrochemical biosensor; Silver nanoparticle; Microfluidic; Blood analysis; Differential pulse voltammetry; ELECTROCHEMICAL DETECTION; HEART-FAILURE; SILVER; PAPER;
D O I
10.1016/j.aca.2021.339375
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The diagnosis of many diseases requires monitoring of biomarker levels over a period of time instead of assessing their concentration only once. For example, in case of heart failure determination, the levels of N-terminal prohormone brain natriuretic peptide (NT-proBNP) in blood vary so strongly amongst individuals, that the current procedure of one-time measurement in combination with clinical examination does not allow for accurate assessment of disease severity and progression. Our microfluidic biosensor addresses key characteristics of desirable home-tests which include low limits of detection, small sample volume (less than 10 mL), simple detection strategies, and ready-to-go all-dried long-term stable reagents. Here, electrochemically superior silver nanoparticles (AgNP) were dried directly within the microfluidic channel in a matrix of trehalose sugar doped with Na2SO3 as oxygen scavenger. This successfully prevented AgNP oxidation and enabled dry and ready-to-use storage for at least 18 weeks. Based on this, laser-cut flow chips were developed containing all bioassay reagents needed in a ready-togo dry format. An oxidation-reduction stripping voltammetry strategy was used for highly sensitive quantification of the AgNPs as electrochemical label. This microfluidic biosensor demonstrated limits of detection for NT-proBNP of 0.57 ng mL-1 with a mean error of 6% (n > 3) in undiluted human serum, which is below the clinically relevant cut-off of 1 ng mL-1. This practical approach has the potential to substitute commonly used lateral-flow assays for various biomarkers, as it offers low patient sample volumes hence supporting simple finger-prick strategies well-known also for other electrochemical biosensors, and independence from the notorious variability in fleece fabrication. (c) 2021 Elsevier B.V. All rights reserved.
引用
收藏
页数:9
相关论文
共 22 条
[21]   Ultra-Sensitive NT-proBNP Quantification for Early Detection of Risk Factors Leading to Heart Failure [J].
Song, Keum-Soo ;
Nimse, Satish Balasaheb ;
Sonawane, Mukesh Digambar ;
Warkad, Shrikant Dashrath ;
Kim, Taisun .
SENSORS, 2017, 17 (09)
[22]   Hybrid paper and 3D-printed microfluidic device for electrochemical detection of Ag nanoparticle labels [J].
Walgama, Charuksha ;
Nguyen, Michael P. ;
Boatner, Lisa M. ;
Richards, Ian ;
Crooks, Richard M. .
LAB ON A CHIP, 2020, 20 (09) :1648-1657