Monitoring of dissolved reactive phosphorus in wastewaters by flow injection analysis .2. On-line monitoring system

被引:6
|
作者
Benson, RL
Truong, YB
McKelvie, ID
Hart, BT
Bryant, GW
Hilkmann, WP
机构
[1] MONASH UNIV,WATER STUDIES CTR,CAULFIELD,VIC 3145,AUSTRALIA
[2] MONASH UNIV,DEPT CHEM,CAULFIELD,VIC 3145,AUSTRALIA
[3] BHP RES,NEWCASTLE LABS,WALLSEND,NSW 2287,AUSTRALIA
关键词
flow injection; phosphorus; wastewaters; on-line monitoring; filtration;
D O I
10.1016/0043-1354(96)00009-7
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
An automated-flow injection (FI) analyser for the determination of dissolved reactive phosphorus (DRP) in wastewaters is described. The analytical method is based on the formation of molybdenum blue. The limits of detection using a solid-state photometric detector (LED-photodiode) were 0.05 mg P 1(-1) and the linear range was 0-20 mg P 1(-1). Details of construction, field-testing (including sample filtration) and validation are given. The analyser proved very reliable during an unattended operational period of seven days. Operational costs were minimised, with low reagent consumption and minimal operator intervention. Data retrieval and analyser status were remotely interrogated by means of a cellular telephone-based communications link. The real-time results from the analyser showed good correlation with those generated by a validated laboratory method. Copyright (C) 1996 Elsevier Science Ltd
引用
收藏
页码:1965 / 1971
页数:7
相关论文
共 50 条
  • [31] On-line optical monitoring system for arc welding
    Sforza, P
    de Blasiis, D
    NDT & E INTERNATIONAL, 2002, 35 (01) : 37 - 43
  • [32] Development of On-line Laser Power Monitoring System
    Ding, Chien-Fang
    Lee, Meng-Shiou
    Li, Kuan-Ming
    REAL-TIME MEASUREMENTS, ROGUE EVENTS, AND EMERGING APPLICATIONS, 2016, 9732
  • [33] Development of on-line monitoring and diagnosing system for dredger
    Li, YJ
    Yang, JG
    ICEMI 2005: CONFERENCE PROCEEDINGS OF THE SEVENTH INTERNATIONAL CONFERENCE ON ELECTRONIC MEASUREMENT & INSTRUMENTS, VOL 8, 2005, : 369 - 375
  • [34] On-line monitoring system for downhole temperature and pressure
    Ma Shou
    Guo Jianchun
    Guo Yanli
    Cao Jun
    Zhang Boming
    OPTICAL ENGINEERING, 2014, 53 (08)
  • [35] DETERMINATION OF TOTAL PHOSPHORUS IN WATERS AND WASTEWATERS BY ONLINE MICROWAVE-INDUCED DIGESTION AND FLOW-INJECTION ANALYSIS
    BENSON, RL
    MCKELVIE, ID
    HART, BT
    HAMILTON, IC
    ANALYTICA CHIMICA ACTA, 1994, 291 (03) : 233 - 242
  • [36] On-line Monitoring System for Monitoring SO2 in Fumes (1) - Monitoring Subsystem for SO2 Mass Concentration
    Wang, Shilong
    Wang, Lina
    Wang, Hongbo
    Cai, Yonghui
    EQUIPMENT MANUFACTURING TECHNOLOGY, 2012, 422 : 296 - 299
  • [37] A flow-injection (FI) mediated biosensor for on-line monitoring of lactate in mammalian cell culture
    Male, KB
    Gartu, PO
    Kamen, AA
    Luong, JHT
    ANALYTICA CHIMICA ACTA, 1997, 351 (1-3) : 159 - 167
  • [38] Development of an On-line Monitoring System for Mental Corrosion
    Sun Lingfang
    Li Jing
    Zhang Shujuan
    2008 10TH INTERNATIONAL CONFERENCE ON CONTROL AUTOMATION ROBOTICS & VISION: ICARV 2008, VOLS 1-4, 2008, : 1934 - 1937
  • [39] Flow-injection spectrofluorimetric determination of dissolved inorganic and organic phosphorus in waters using on-line photo-oxidation
    Pérez-Ruiz, T
    Martínez-Lozano, C
    Tomás, V
    Martín, J
    ANALYTICA CHIMICA ACTA, 2001, 442 (01) : 147 - 153
  • [40] On-line Monitoring System for Icing State of Overhead Transmission Line
    Nie, Shiguang
    Qu, Guangyu
    Ye, Haifeng
    Jiang, Xiuchen
    2012 POWER ENGINEERING AND AUTOMATION CONFERENCE (PEAM), 2012, : 506 - 510