Effects of Inflow and Radiofrequency Spoiling on the Arterial Input Function in Dynamic Contrast-Enhanced MRI: A Combined Phantom and Simulation Study

被引:35
作者
Garpebring, Anders [1 ]
Wirestam, Ronnie [2 ]
Ostlund, Nils [3 ]
Karlsson, Mikael [1 ]
机构
[1] Umea Univ, Dept Radiat Sci, Div Radiat Phys, SE-90187 Umea, Sweden
[2] Lund Univ, Dept Med Radiat Phys, Lund, Sweden
[3] Umea Univ Hosp, Dept Radiat Sci, Div Biomed Engn & Informat, S-90185 Umea, Sweden
基金
瑞典研究理事会;
关键词
dynamic contrast-enhanced MRI; arterial input function; blood flow effects; RF spoiling; HIGH-TEMPORAL-RESOLUTION; FAST GRADIENT-ECHO; BLOOD-VOLUME; IMPROVED ACCURACY; T-1; MEASUREMENT; FLOWING BLOOD; PERMEABILITY; PERFUSION; T1; QUANTIFICATION;
D O I
10.1002/mrm.22760
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The arterial input function is crucial in pharmacokinetic analysis of dynamic contrast-enhanced MRI data. Among other artifacts in arterial input function quantification, the blood inflow effect and nonideal radiofrequency spoiling can induce large measurement errors with subsequent reduction of accuracy in the pharmacokinetic parameters. These errors were investigated for a 3D spoiled gradient-echo sequence using a pulsatile flow phantom and a total of 144 typical imaging settings. In the presence of large inflow effects, results showed poor average accuracy and large spread between imaging settings, when the standard spoiled gradient-echo signal equation was used in the analysis. For example, one of the investigated inflow conditions resulted in a mean error of about 40% and a spread, given by the coefficient of variation, of 20% for K-trans. Minimizing inflow effects by appropriate slice placement, combined with compensation for nonideal radiofrequency spoiling, significantly improved the results, but they remained poorer than without flow (e. g., 3-4 times larger coefficient of variation for K-trans). It was concluded that the 3D spoiled gradient-echo sequence is not optimal for accurate arterial input function quantification and that correction for nonideal radiofrequency spoiling in combination with inflow minimizing slice placement should be used to reduce the errors. Magn Reson Med 65:1670-1679, 2011. (C)2011 Wiley-Liss, Inc.
引用
收藏
页码:1670 / 1679
页数:10
相关论文
共 46 条
  • [1] Brain hydrodynamics study by phase-contrast magnetic resonance imaging and transcranial color Doppler
    Baledent, Olivier
    Fin, Loic
    Khuoy, Louis
    Ambarki, Khalid
    Gauvin, Anne-Cecile
    Gondry-Jouet, Catherine
    Meyer, Marc-Etienne
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2006, 24 (05) : 995 - 1004
  • [2] AZD2171, a pan-VEGF receptor tyrosine kinase inhibitor, normalizes tumor vasculature and alleviates edema in glioblastoma patients
    Batchelor, Tracy T.
    Sorensen, A. Gregory
    di Tomaso, Emmanuelle
    Zhang, Wei-Ting
    Duda, Dan G.
    Cohen, Kenneth S.
    Kozak, Kevin R.
    Cahill, Daniel P.
    Chen, Poe-Jou
    Zhu, Mingwang
    Ancukiewicz, Marek
    Mrugala, Maciej M.
    Plotkin, Scott
    Drappatz, Jan
    Louis, David N.
    Ivy, Percy
    Scadden, David T.
    Benner, Thomas
    Loeffler, Jay S.
    Wen, Patrick Y.
    Jain, Rakesh K.
    [J]. CANCER CELL, 2007, 11 (01) : 83 - 95
  • [3] Bernstein MA., 2004, Handbook of MRI pulse sequences, P659
  • [4] Brookes JA, 1999, JMRI-J MAGN RESON IM, V9, P163, DOI 10.1002/(SICI)1522-2586(199902)9:2<163::AID-JMRI3>3.0.CO
  • [5] 2-L
  • [6] Investigation and optimization of parameter accuracy in dynamic contrast-enhanced MRI
    Cheng, Hai-Ling Margaret
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 28 (03) : 736 - 743
  • [7] T1 measurement of flowing blood and arterial input function determination for quantitative 3D T1-weighted DCE-MRI
    Cheng, Hai-Ling Margaret
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 25 (05) : 1073 - 1078
  • [8] Rapid high-resolution T1 mapping by variable flip angles:: Accurate and precise measurements in the presence of radiofrequency field inhomogeneity
    Cheng, HLM
    Wright, GA
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2006, 55 (03) : 566 - 574
  • [9] Dynamic contrast-enhanced magnetic resonance imaging of radiation therapy-induced microcirculation changes in rectal cancer
    De Lussanet, QG
    Backes, WH
    Griffioen, AW
    Padhani, AR
    Baeten, CI
    Van Baardwijk, A
    Lambin, P
    Beets, GL
    Van Engelshoven, JMA
    Beets-Tan, RGH
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (05): : 1309 - 1315
  • [10] The accuracy of pharmacokinetic parameter measurement in DCE-MRI of the breast at 3 T
    Di Giovanni, P.
    Azlan, C. A.
    Ahearn, T. S.
    Semple, S. I.
    Gilbert, F. J.
    Redpath, T. W.