Surface dosimetry for oblique tangential photon beams: A Monte Carlo simulation study

被引:22
作者
Chow, James C. L. [1 ,2 ,3 ,4 ]
机构
[1] Univ Toronto, Princess Margaret Hosp, Radiat Med Program, Toronto, ON M5G 2M9, Canada
[2] Univ Toronto, Dept Radiat Oncol, Toronto, ON M5G 2M9, Canada
[3] Univ Waterloo, Dept Phys, Waterloo, ON N2L 3G1, Canada
[4] Ryerson Univ, Dept Phys, Toronto, ON M5B 2K3, Canada
关键词
surface dose; beam obliquity; Monte Carlo simulation; tangential beam;
D O I
10.1118/1.2818956
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The effect of beam obliquity on the surface relative dose profiles for the tangential photon beams was studied. The 6 and 15 MV photon beams with 4 x 4 and 10 x 10 cm(2) field sizes produced by a Varian 21 EX linear accelerator were used. Phase-space models of the photon beams were created using Monte Carlo simulations based on the EGSnrc code, and were verified using film measurements. The relative dose profiles in the phantom skin, at 2 mm depth from the surface of the half-phantom geometry, or HPG, were calculated for increasing gantry angles from 270 to 280 deg clockwise. Relative dose profiles of a full phantom enclosing the whole tangential beam (full phantom geometry, or FPG) were also calculated using Monte Carlo simulation as a control for comparison. The results showed that, although the relative dose profiles in the phantom skin did not change significantly with an oblique beam using a FPG, the surface relative depth dose was increased for the HPG. In the HPG, with 6 MV photon beams and field size= 10 x 10 cm(2), when the beam angle, starting from 270 deg, was increased from 1 to 3 deg, the relative depth doses in the phantom skin were increased from 68% to 79% at 10 cm depth. This increase in dose was slightly larger than the dose from 15 MV photon beams with the same field size and beam angles, where the relative depth doses in phantom skin were increased from 81% to 87% at 10 cm depth. A parameter called the percent depth dose (PDD) ratio, defined as the relative depth dose from the HPG to the relative depth dose from the FPG at a given depth along the phantom skin, was used to evaluate the effect of the phantom-air interface. It is found that the PDD ratio increased significantly when the beam angle was changed from zero to 1-3 degrees. Moreover, the PDD ratio, for a given field size, experienced a greater increase for 6 MV than for 15 MV. For the same photon beam energy, the PDD ratio increased more with a 4 x 4 cm(2) field compared to 10 x 10 cm(2). The results in this study will be useful for physicists and dosimetrists to predict the surface relative dose variations when using clinical tangential-like photon beams in radiation therapy. (c) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:70 / 76
页数:7
相关论文
共 27 条
  • [1] Validation of Monte Carlo calculated surface doses for megavoltage photon beams
    Abdel-Rahman, W
    Seuntjens, JP
    Verhaegen, F
    Deblois, F
    Podgorsak, EB
    [J]. MEDICAL PHYSICS, 2005, 32 (01) : 286 - 298
  • [2] Treatment planning challenges in breast irradiation: The ideal and the practical
    Beavis, AW
    [J]. CLINICAL ONCOLOGY, 2006, 18 (03) : 200 - 209
  • [3] Bielajew A.F., 1984, NUCL INSTRUM METH B, V88, P535
  • [4] BJARNGARD BE, 1995, MED PHYS, V22, P465, DOI 10.1118/1.597477
  • [5] Measurement of 6-MV x-ray surface dose when topical agents are applied prior to external beam irradiation
    Burch, SE
    Parker, SA
    Vann, AM
    Arazie, JC
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (02): : 447 - 451
  • [6] Surface dose extrapolation measurements with radiographic film
    Butson, MJ
    Cheung, T
    Yu, PKN
    Currie, M
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (13) : N197 - N201
  • [7] Study on surface dose generated in prostate intensity-modulated radiation therapy treatment
    Chow, James C. L.
    Grigorov, Grigor N.
    Barnett, Rob B.
    [J]. MEDICAL DOSIMETRY, 2006, 31 (04) : 249 - 258
  • [8] Dose measurements near a non-radioactive gold seed using radiographic film
    Chow, JCL
    Grigorov, GN
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (18) : N227 - N234
  • [9] Evaluation of surface and build-up region dose for intensity-modulated radiation therapy in head and neck cancer
    Chung, HT
    Jin, HS
    Dempsey, JF
    Liu, CH
    Palta, J
    Suh, TS
    Kim, SY
    [J]. MEDICAL PHYSICS, 2005, 32 (08) : 2682 - 2689
  • [10] Accurate skin dose measurements using radiochromic film in clinical applications
    Devic, S
    Seuntjens, J
    Abdel-Rahman, W
    Evans, M
    Olivares, M
    Podgorsak, EB
    Vuong, T
    Soares, CG
    [J]. MEDICAL PHYSICS, 2006, 33 (04) : 1116 - 1124