T1 quantification with inversion recovery TrueFISP

被引:145
作者
Scheffler, K [1 ]
Hennig, J [1 ]
机构
[1] Univ Freiburg, Dept Diagnost Radiol, Sect Med Phys, D-79106 Freiburg, Germany
关键词
TrueFISP; T-1; quantification; inversion recovery; DEFT; off resonance;
D O I
10.1002/mrm.1097
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A snapshot FLASH sequence can be used to acquire the time course of longitudinal magnetization during its recovery after a single inversion pulse. However, excitation pulses disturb the exponential recovery of longitudinal magnetization and may produce systematic errors in T-1 estimations. In this context the possibility of using the TrueFISP sequence to detect the recovery of longitudinal magnetization for quantitative T-1 measurements was examined. Experiments were performed on different Gd-doped water phantoms and on humans. T-1 values derived from inversion recovery TrueFISP were in excellent agreement with the single-point method even for flip angles up to 50 degrees. In terms of T-1 accuracy and SNR, the proposed method seems to be superior to the conventional inversion recovery snapshot FLASH technique. Magn Reson Med 45:720-723, 2001. (C) 2001 Wiley-Liss, Inc.
引用
收藏
页码:720 / 723
页数:4
相关论文
共 19 条