Three-phase hollow fiber microextraction based on two immiscible organic solvents for determination of tricyclic antidepressant drugs: Comparison with conventional three-phase hollow fiber microextraction

被引:61
作者
Ghambarian, Mahnaz [1 ]
Yamini, Yadollah [1 ]
Esrafili, Ali [1 ]
机构
[1] Tarbiat Modares Univ, Fac Sci, Dept Chem, Tehran, Iran
关键词
Hollow fiber based microextraction; Tricyclic antidepressant drugs; High performance liquid chromatography-diode array detector; Gas chromatography-mass spectrometry; PERFORMANCE LIQUID-CHROMATOGRAPHY; SOLID-PHASE MICROEXTRACTION; SUPERCRITICAL-FLUID EXTRACTION; MICROWAVE-ASSISTED EXTRACTION; TANDEM MASS-SPECTROMETRY; SAMPLE PREPARATION; BIOLOGICAL-FLUIDS; HUMAN PLASMA; HUMAN SERUM; AMITRIPTYLINE;
D O I
10.1016/j.chroma.2011.11.055
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The aim of this research was to compare the extraction efficiencies of two modes of three-phase hollow fiber microextraction (HF-LLLME) based on aqueous and organic acceptor phases for analysis of tricyclic antidepressant (TCA) drugs. High-performance liquid chromatography with photodiode array detection (HPLC-DAD) was applied for determination of the drugs. In order to examine the ability of the new concept of HF-LLLME based on organic acceptor solvent in comparison with aqueous acceptor phase to extract the analytes, four TCAs were selected. The effect of different extraction conditions (i.e., type of acceptor phase, hollow fiber length, ionic strength, stirring rate, and extraction time) on the extraction efficiency of the TCAs was investigated and optimized using central composite design (CCD) as a powerful tool. Both methods were characterized by good linearity and high repeatability, but HF-LLLME with organic acceptor provided higher extraction efficiency and thus lower limits of detection (LODs). Calibration curves were linear (r(2) > 0.996) in the range of 0.2-20014 mu g L-1. LODs for all the TCAs ranged from 0.08 to 0.2 mu g L-1 using HPLC-DAD. Also an improvement in sensitivity of several orders of magnitude was achieved using single-ion monitoring GC-MS analyses (0.04 mu g L-1) due to compatibility of this technique with GC instrument. The applicability of the proposed HF-LLLME/GC-MS and HPLC-DAD methods was demonstrated by analyzing the drugs in spiked urine and plasma samples. The obtained recoveries of the drugs in the range of 87.9-109.2% indicated the excellent capability of the developed method for extraction of TCAs from complex matrices. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:5 / 12
页数:8
相关论文
共 38 条
  • [1] Bolton S., 1984, PHARM STAT PRACTICAL, P430
  • [2] Application of chemometric techniques to the optimization of the solid-phase extraction of 27 pesticides before GC-MIP-AES analysis
    Carro, AM
    Cobas, JC
    Rodríguez, JB
    Lorenzo, RA
    Cela, R
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (12) : 1867 - 1873
  • [3] Stir bar sorptive extraction and liquid chromatography with UV detection for determination of antidepressants in plasma samples
    Chaves, Andrea Rodrigues
    Silva, Silvana Maciel
    Costa Queiroz, Regina Helena
    Lancas, Fernando Mauro
    Costa Queiroz, Maria Eugnia
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 850 (1-2): : 295 - 302
  • [4] High-throughput on-line solid-phase extraction-liquid chromatography-tandem mass spectrometry method for the simultaneous analysis of 14 antidepressants and their metabolites in plasma
    de Castro, Ana
    Femandez, Maria Del Mar Ramirez
    Laloup, Marleen
    Samyn, Nele
    De Boeck, Gert
    Wood, Michelle
    Maes, Viviane
    Lopez-Rivadulla, Manuel
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2007, 1160 (1-2) : 3 - 12
  • [5] Extraction of Azole Antifungal Drugs from Milk and Biological Fluids Using a New Hollow Fiber Liquid-Phase Microextraction and Analysis by GC-FID
    Ebrahimpour, Behnam
    Yamini, Yadollah
    Esrafili, Ali
    [J]. CHROMATOGRAPHIA, 2011, 74 (3-4) : 281 - 289
  • [6] Hollow fiber-based liquid phase microextraction combined with high-performance liquid chromatography for extraction and determination of some antidepressant drugs in biological fluids
    Esrafili, Ali
    Yamini, Yadollah
    Shariati, Shahab
    [J]. ANALYTICA CHIMICA ACTA, 2007, 604 (02) : 127 - 133
  • [7] Dynamic three-phase hollow fiber microextraction based on two immiscible organic solvents with automated movement of the acceptor phase
    Esrafili, Ali
    Yamini, Yadollah
    Ghambarian, Mahnaz
    Moradi, Morteza
    [J]. JOURNAL OF SEPARATION SCIENCE, 2011, 34 (01) : 98 - 106
  • [8] Analysis of eighteen antidepressants, four atypical antipsychotics and active metabolites in serum by liquid chromatography:: a simple tool for therapeutic drug monitoring
    Frahnert, C
    Rao, ML
    Grasmäder, K
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 794 (01): : 35 - 47
  • [9] Toxicokinetics of nortriptyline and amitriptyline: Two case reports
    Franssen, EJF
    Kunst, PWA
    Bet, PM
    van Schijndel, RJMS
    van Loenen, AC
    Wilhelm, AJ
    [J]. THERAPEUTIC DRUG MONITORING, 2003, 25 (02) : 248 - 251
  • [10] PHARMACOKINETIC OPTIMIZATION OF TRICYCLIC ANTIDEPRESSANT THERAPY
    FURLANUT, M
    BENETELLO, P
    SPINA, E
    [J]. CLINICAL PHARMACOKINETICS, 1993, 24 (04) : 301 - 318