Study of the IMRT interplay effect using a 4DCT Monte Carlo dose calculation

被引:10
作者
Jensen, Michael D. [1 ]
Abdellatif, Ady [2 ]
Chen, Jeff [1 ,2 ,3 ]
Wong, Eugene [1 ,2 ,3 ,4 ]
机构
[1] Univ Western Ontario, Dept Med Biophys, London, ON, Canada
[2] London Hlth Sci Ctr, London Reg Canc Program, Dept Phys & Engn, London, ON, Canada
[3] Univ Western Ontario, Dept Oncol, London, ON, Canada
[4] Univ Western Ontario, Dept Phys & Astron, London, ON, Canada
关键词
DEFORMABLE IMAGE REGISTRATION; INTRA-FRACTION MOTION; QUALITY-ASSURANCE; VERIFICATION; MANAGEMENT; DOSIMETRY; DELIVERY; BEAM;
D O I
10.1088/0031-9155/57/8/N89
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Respiratory motion may lead to dose errors when treating thoracic and abdominal tumours with radiotherapy. The interplay between complex multileaf collimator patterns and patient respiratory motion could result in unintuitive dose changes. We have developed a treatment reconstruction simulation computer code that accounts for interplay effects by combining multileaf collimator controller log files, respiratory trace log files, 4DCT images and a Monte Carlo dose calculator. Two three-dimensional (3D) IMRT step-and-shoot plans, a concave target and integrated boost were delivered to a 1D rigid motion phantom. Three sets of experiments were performed with 100%, 50% and 25% duty cycle gating. The log files were collected, and five simulation types were performed on each data set: continuous isocentre shift, discrete isocentre shift, 4DCT, 4DCT delivery average and 4DCT plan average. Analysis was performed using 3D gamma analysis with passing criteria of 2%, 2 mm. The simulation framework was able to demonstrate that a single fraction of the integrated boost plan was more sensitive to interplay effects than the concave target. Gating was shown to reduce the interplay effects. We have developed a 4DCT Monte Carlo simulation method that accounts for IMRT interplay effects with respiratory motion by utilizing delivery log files.
引用
收藏
页码:N89 / N99
页数:11
相关论文
共 50 条
[41]   Evaluation of a new commercial Monte Carlo dose calculation algorithm for electron beams [J].
Vandervoort, Eric J. ;
Tchistiakova, Ekaterina ;
La Russa, Daniel J. ;
Cygler, Joanna E. .
MEDICAL PHYSICS, 2014, 41 (02) :021711
[42]   A method using 4D dose accumulation to quantify the interplay effect in lung stereotactic body radiation therapy [J].
Huesa-Berral, Carlos ;
Burguete, Javier ;
Moreno-Jimenez, Marta ;
Diego Azcona, Juan .
PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (03)
[43]   Novel Monte Carlo dose calculation algorithm for robotic radiosurgery with multi leaf collimator: Dosimetric evaluation [J].
Heidorn, Sarah-Charlotta ;
Kilby, Warren ;
Fuerweger, Christoph .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 55 :25-32
[44]   Feasibility of using Geant4 Monte Carlo simulation for IMRT dose calculations for the Novalis Tx with a HD-120 multi-leaf collimator [J].
Jung, Hyunuk ;
Shin, Jungsuk ;
Chung, Kwangzoo ;
Han, Youngyih ;
Kim, Jinsung ;
Choi, Doo Ho .
JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2015, 66 (10) :1489-1494
[45]   Quantifying ventilation change due to radiation therapy using 4DCT Jacobian calculations [J].
Patton, Taylor J. ;
Gerard, Sarah E. ;
Shao, Wei ;
Christensen, Gary E. ;
Reinhardt, Joseph M. ;
Bayouth, John E. .
MEDICAL PHYSICS, 2018, 45 (10) :4483-4492
[46]   A Numerical Approach to Dose Optimization for Moving Targets Using Monte Carlo Simulations [J].
Chun, Kook Jin ;
Park, Youngho ;
Choi, Yongsoo ;
Hyun, Sangil .
RADIATION RESEARCH, 2009, 171 (02) :245-253
[47]   Investigation of stereotactic radiotherapy dose using dosimetry film and Monte Carlo simulations [J].
Kairn, T. ;
Crowe, S. ;
Kenny, J. ;
Trapp, J. V. .
RADIATION MEASUREMENTS, 2011, 46 (12) :1985-1988
[48]   The Effect of Nucleus Size on the Cell Dose in Targeted Radionuclide Therapy - A Monte Carlo Study [J].
Kouhkan, Ebrahim ;
Chegeni, Nahid ;
Hussain, Amjad .
JOURNAL OF MEDICAL SIGNALS & SENSORS, 2020, 10 (02) :113-118
[49]   A Monte Carlo study on the effect of the orbital bone to the radiation dose delivered to the eye lens [J].
Stratis, Andreas ;
Zhang, Guozhi ;
Jacobs, Reinhilde ;
Bogaerts, Ria ;
Bosmans, Hilde .
MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
[50]   Benchmarking and validation of a Geant4-SHADOW Monte Carlo simulation for dose calculations in microbeam radiation therapy [J].
Cornelius, Iwan ;
Guatelli, Susanna ;
Fournier, Pauline ;
Crosbie, Jeffrey C. ;
del Rio, Manuel Sanchez ;
Braeuer-Krisch, Elke ;
Rosenfeld, Anatoly ;
Lerch, Michael .
JOURNAL OF SYNCHROTRON RADIATION, 2014, 21 :518-528