Study of the IMRT interplay effect using a 4DCT Monte Carlo dose calculation

被引:10
作者
Jensen, Michael D. [1 ]
Abdellatif, Ady [2 ]
Chen, Jeff [1 ,2 ,3 ]
Wong, Eugene [1 ,2 ,3 ,4 ]
机构
[1] Univ Western Ontario, Dept Med Biophys, London, ON, Canada
[2] London Hlth Sci Ctr, London Reg Canc Program, Dept Phys & Engn, London, ON, Canada
[3] Univ Western Ontario, Dept Oncol, London, ON, Canada
[4] Univ Western Ontario, Dept Phys & Astron, London, ON, Canada
关键词
DEFORMABLE IMAGE REGISTRATION; INTRA-FRACTION MOTION; QUALITY-ASSURANCE; VERIFICATION; MANAGEMENT; DOSIMETRY; DELIVERY; BEAM;
D O I
10.1088/0031-9155/57/8/N89
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Respiratory motion may lead to dose errors when treating thoracic and abdominal tumours with radiotherapy. The interplay between complex multileaf collimator patterns and patient respiratory motion could result in unintuitive dose changes. We have developed a treatment reconstruction simulation computer code that accounts for interplay effects by combining multileaf collimator controller log files, respiratory trace log files, 4DCT images and a Monte Carlo dose calculator. Two three-dimensional (3D) IMRT step-and-shoot plans, a concave target and integrated boost were delivered to a 1D rigid motion phantom. Three sets of experiments were performed with 100%, 50% and 25% duty cycle gating. The log files were collected, and five simulation types were performed on each data set: continuous isocentre shift, discrete isocentre shift, 4DCT, 4DCT delivery average and 4DCT plan average. Analysis was performed using 3D gamma analysis with passing criteria of 2%, 2 mm. The simulation framework was able to demonstrate that a single fraction of the integrated boost plan was more sensitive to interplay effects than the concave target. Gating was shown to reduce the interplay effects. We have developed a 4DCT Monte Carlo simulation method that accounts for IMRT interplay effects with respiratory motion by utilizing delivery log files.
引用
收藏
页码:N89 / N99
页数:11
相关论文
共 50 条
  • [31] Spatial fractionation of the dose using neon and heavier ions: A Monte Carlo study
    Peucelle, C.
    Martinez-Rovira, I.
    Prezadoa, Y.
    [J]. MEDICAL PHYSICS, 2015, 42 (10) : 5928 - 5936
  • [32] Evaluation of 4D dose to a moving target with Monte Carlo dose calculation in stereotactic body radiotherapy for lung cancer
    Matsugi K.
    Nakamura M.
    Miyabe Y.
    Yamauchi C.
    Matsuo Y.
    Mizowaki T.
    Hiraoka M.
    [J]. Radiological Physics and Technology, 2013, 6 (1) : 233 - 240
  • [33] Independent absorbed-dose calculation using the Monte Carlo algorithm in volumetric modulated arc therapy
    Haga, Akihiro
    Magome, Taiki
    Takenaka, Shigeharu
    Imae, Toshikazu
    Sakumi, Akira
    Nomoto, Akihiro
    Igaki, Hiroshi
    Shiraishi, Kenshiro
    Yamashita, Hideomi
    Ohtomo, Kuni
    Nakagawa, Keiichi
    [J]. RADIATION ONCOLOGY, 2014, 9
  • [34] Dose Calculation in a Mouse Lung Tumor and in Secondary Organs During Radiotherapy Treatment: A Monte Carlo Study
    Hamdi, Mahdjoub
    Mimi, Malika
    Bentourkia, M'hamed
    [J]. BIOINFORMATICS AND BIOMEDICAL ENGINEERING (IWBBIO 2015), PT I, 2015, 9043 : 360 - 367
  • [35] The impact of low-Z and high-Z metal implants in IMRT: A Monte Carlo study of dose inaccuracies in commercial dose algorithms
    Spadea, Maria Francesca
    Verburg, Joost Mathias
    Baroni, Guido
    Seco, Joao
    [J]. MEDICAL PHYSICS, 2014, 41 (01)
  • [36] GPU-based fast Monte Carlo simulation for radiotherapy dose calculation
    Jia, Xun
    Gu, Xuejun
    Graves, Yan Jiang
    Folkerts, Michael
    Jiang, Steve B.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (22) : 7017 - 7031
  • [37] Monte Carlo simulation on heterogeneous phantom and physical dose calculation on heterogeneous organ
    Chemingui, F. Z.
    Benrachi, F.
    Bali, M. S.
    [J]. JOURNAL OF NEW TECHNOLOGY AND MATERIALS, 2019, 9 (02) : 51 - 56
  • [38] GPU-based fast Monte Carlo dose calculation for proton therapy
    Jia, Xun
    Schuemann, Jan
    Paganetti, Harald
    Jiang, Steve B.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (23) : 7783 - 7797
  • [39] Hybrid Monte Carlo approach for accurate and efficient shutdown dose rate calculation
    Lu, Peng
    Fischer, Ulrich
    Pereslavtsev, Pavel
    Qiu, Yuefeng
    Serikov, Arkady
    Li, Jiangang
    Song, Yuntao
    [J]. FUSION ENGINEERING AND DESIGN, 2018, 136 : 498 - 502
  • [40] Evaluation of a commercial Monte Carlo dose calculation algorithm for electron treatment planning
    Huang, Jessie Y.
    Dunkerley, David
    Smilowitz, Jennifer B.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (06): : 184 - 193