Hierarchy of stability factors in reverse shoulder arthroplasty

被引:94
作者
Gutierrez, Sergio [1 ,2 ]
Keller, Tony S. [1 ]
Levy, Jonathan C. [3 ]
Lee, William E., III [2 ]
Luo, Zong-Ping [1 ]
机构
[1] Florida Orthopaed Inst, Tampa, FL 33637 USA
[2] Univ S Florida, Coll Engn Chem & Biomed Engn, Tampa, FL USA
[3] Holy Cross Hosp, Inst Orthopaed, Ft Lauderdale, FL USA
关键词
D O I
10.1007/s11999-007-0096-0
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Reverse shoulder arthroplasty is being used more frequently to treat irreparable rotator cuff tears in the presence of glenohumeral arthritis and instability. To date, however, design features and functions of reverse shoulder arthroplasty, which may be associated with subluxation and dislocation of these implants, have been poorly understood. We asked: (l) what is the hierarchy of importance of joint compressive force, prosthetic socket depth, and glenosphere size in relation to stability, and (2) is this hierarchy defined by underlying and theoretically predictable joint contact characteristics? We examined the intrinsic stability in terms of the force required to dislocate the humerosocket from the glenosphere of eight commercially available reverse shoulder arthroplasty devices. The hierarchy of factors was led by compressive force followed by socket depth; glenosphere size played a much lesser role in stability of the reverse shoulder arthroplasty device. Similar results were predicted by a mathematical model, suggesting the stability was determined primarily by compressive forces generated by muscles.
引用
收藏
页码:670 / 676
页数:7
相关论文
共 36 条
[1]  
Anglin C, 2000, J SHOULDER ELB SURG, V9, P104
[2]  
*ASTM, STAND SPEC ULTR HIGH, pF648
[3]   Grammont reverse prosthesis: Design, rationale, and biomechanics [J].
Boileau, P ;
Watkinson, DJ ;
Hatzidakis, AM ;
Balg, F .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2005, 14 (01) :147S-161S
[4]   Confidence intervals for mean absolute deviations [J].
Bonett, DG ;
Seier, E .
AMERICAN STATISTICIAN, 2003, 57 (04) :233-236
[5]  
Boulahia A, 2002, ORTHOPEDICS, V25, P129
[6]   Friction of total hip replacements with different bearings and loading conditions [J].
Brockett, Claire ;
Williams, Sophie ;
Jin, Zhongmin ;
Isaac, Graham ;
Fisher, John .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2007, 81B (02) :508-515
[7]   Reverse shoulder arthroplasty for the treatment of three- and four-part fractures of the proximal humerus in the elderly - A prospective review of 43 cases with a short-term follow-up [J].
Bufquin, T. ;
Hersan, A. ;
Hubert, L. ;
Massin, P. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (04) :516-520
[8]   Grammont reversed prosthesis for acute complex fracture of the proximal humerus in an elderly population with 5 to 12 years follow-up [J].
Cazeneuve, J. -F. ;
Cristofari, D. -J. .
REVUE DE CHIRURGIE ORTHOPEDIQUE ET REPARATRICE DE L APPAREIL MOTEUR, 2006, 92 (06) :543-548
[9]  
*CTR ADV FRICT STU, 2005, FRICT CTR COEFF DAT
[10]  
De Wilde L, 2001, Acta Orthop Belg, V67, P348