Validation of nested PCR and a selective biochemical method as alternatives for mycoplasma detection

被引:7
作者
Cheong, Kyung Ah [1 ]
Agrawal, Santosh Rani [2 ]
Lee, Ai-Young [1 ]
机构
[1] Dongguk Univ Ilsan Hosp, Dept Dermatol, Goyang, South Korea
[2] Dongguk Univ, Biotechnol Res Inst, Gyeonggi Do, South Korea
关键词
Direct culture; MycoAlert (R); Nested PCR (TaKaRa (R)); Parameters; Validation; LINKED-IMMUNOSORBENT-ASSAY; CELLS; HYOPNEUMONIAE; CONTAMINATION; HOMINIS; SWINE; LUNG; RNA;
D O I
10.1002/jobm.201000066
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Direct culture is the most common way to reliably detect mycoplasma, but it is not practical for the qualitative control of cell therapeutics because of the elaborate culture medium, the prolonged incubation time, and the large sample volumes. Here, we chose two alternative methods using commercial detection kits, the PCR mycoplasma detection kit with nested PCR and the selective biochemical method, MycoAlert (R), and validated them with the direct culture method as a reference. We tested eight mycoplasma species and five validation parameters: specificity, detection limit, robustness, repeatability, and ruggedness, based on the regulatory guidelines in the US Pharmacopoeia. All experiments were performed using fibroblasts spiked with mycoplasma. Specificity tests for both methods included all mycoplasma species, except Mycoplasma pneumonia and M. genitalium for the nested PCR and Ureaplasma urealyticum for the MycoAlert (R) assay. Regarding the detection limit, the nested PCR proved to be as sensitive as the direct culture method and more sensitive than the MycoAlert (R) assay. The predicted median for probit = 0.9 was 54 (44-76) CFU/ml for M. hyorhinis and 16 (13-23) CFU/ml for M. hominis by the nested PCR, but 431 (346-593) CFU/ml and 105 (87-142) CFU/ml, respectively, with MycoAlert (R). Changes in the concentration of reagents, reagent lot, or individual analysts did not influence the results of the examined methods. The results of this study support nested PCR as a valuable alternative for mycoplasma detection.
引用
收藏
页码:215 / 219
页数:5
相关论文
共 24 条
  • [1] An evaluation of PCR methods to detect strains of Mycoplasma fermentans
    Afshar, Baharak
    Pitcher, David
    Nicholas, Robin A. J.
    Miles, Roger J.
    [J]. BIOLOGICALS, 2008, 36 (02) : 117 - 121
  • [2] The use of enzyme-linked immunosorbent assay for detection of Mycoplasma hominis antibodies in infertile women serum samples
    Baczynska, A
    Svenstrup, HF
    Fedder, J
    Birkelund, S
    Christiansen, G
    [J]. HUMAN REPRODUCTION, 2005, 20 (05) : 1277 - 1285
  • [3] Introduction of a validation concept for a PCR-based Mycoplasma detection assay
    Bruchmüller, I
    Pirkl, E
    Herrmann, R
    Stoermer, M
    Eichler, H
    Klüter, H
    Bugert, P
    [J]. CYTOTHERAPY, 2006, 8 (01) : 62 - 69
  • [4] Application and field validation of a PCR assay for the detection of Mycoplasma hyopneumoniae from swine lung tissue samples
    Cai, Hugh Y.
    van Dreumel, Tony
    McEwen, Beverly
    Hornby, Geoff
    Bell-Rogers, Patricia
    McRaild, Pat
    Josephson, Gaylan
    Maxie, Grant
    [J]. JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION, 2007, 19 (01) : 91 - 95
  • [5] MYCOPLASMA-FERMENTANS (INCOGNITUS STRAIN) CELLS ARE ABLE TO FUSE WITH T-LYMPHOCYTES
    DIMITROV, DS
    FRANZOSO, G
    SALMAN, M
    BLUMENTHAL, R
    TARSHIS, M
    BARILE, MF
    ROTTEM, S
    [J]. CLINICAL INFECTIOUS DISEASES, 1993, 17 : S305 - S308
  • [6] Evaluation of the Mycoplasma Duo kit for the detection of Mycoplasma hominis and Ureaplasma urealyticum from urogenital and placental specimens
    Evans, G. E.
    Anderson, T. P.
    Seaward, L. M.
    Murdoch, D. R.
    [J]. BRITISH JOURNAL OF BIOMEDICAL SCIENCE, 2007, 64 (02) : 66 - 69
  • [7] Rapid multiplex nested PCR for detection of respiratory viruses
    Lam, W. Y.
    Yeung, Apple C. M.
    Tang, Julian W.
    Ip, Margaret
    Chan, Edward W. C.
    Hui, Mamie
    Chan, Paul K. S.
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2007, 45 (11) : 3631 - 3640
  • [8] Langdon SP, 2004, METH MOLEC MED, V88, P309
  • [9] LO SC, 1993, MODERN PATHOL, V6, P276
  • [10] Evaluation of NucliSens easyMAG for automated nucleic acid extraction from various clinical specimens
    Loens, K.
    Bergs, K.
    Ursi, D.
    Goossens, H.
    Ieven, M.
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2007, 45 (02) : 421 - 425