Rapid diagnosis of Plasmodium falciparum malaria using a point-of-care loop-mediated isothermal amplification device

被引:9
|
作者
Puri, Madhu [1 ]
Kaur Brar, Harsimran [1 ]
Madan, Evanka [1 ]
Srinivasan, Rajesh [2 ]
Rawat, Kapil [2 ]
Gorthi, Sai Siva [2 ]
Kumari, Geeta [3 ]
Sah, Raj [3 ]
Ojha, Sashi Bhusan [4 ]
Panigrahi, Subhendu [5 ]
Dhangadamajhi, Gunanidhi [4 ]
Muthuswami, Rohini [1 ]
Singh, Shailja [3 ]
Madhubala, Rentala [1 ]
机构
[1] Jawaharlal Nehru Univ, Sch Life Sci, New Delhi, India
[2] Indian Inst Sci, Dept Instrumentat & Appl Phys, Bengaluru, India
[3] Jawaharlal Nehru Univ, Special Ctr Mol Med, New Delhi, India
[4] Maharaja Sriram Chandra Bhanjadeo Univ, Dept Biotechnol, Baripada, India
[5] VSS Med Coll & Hosp, Dept Gen Med, Sambalpur, India
关键词
LAMP; plasmodium falciparum; diagnosis; LOD; parasite density; ADC value; IDENTIFICATION; ENDEMICITY; PARASITES; SAMPLES; PCR;
D O I
10.3389/fcimb.2022.961832
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
LAMP diagnosis of malaria is simple and cost-effective with acceptable sensitivity and specificity as compared to standard diagnostic modules such as microscopy, RDTs and nested PCR, and thus its deployment for onsite screening of malaria in resource-limited regions is under consideration. However, the requirement of an electricity-operated dry bath and bulky read-out unit is still a major concern. In an effort to simplify this limitation, we have developed a portable LAMP device and fluorescence readout unit which can be used in the rapid point-of-care diagnosis of malaria. We have developed a point-of-care diagnostic LAMP device that is easy to operate by a mobile application, and the results can be quantified with a fluorescent readout unit. The diagnostic performance of the device was evaluated in 90 P. falciparum-infected clinical isolates stored at 4 degrees C for 6-7 years and 10 freshly collected isolates from healthy volunteers. The LOD and quantitative ability of LAMP in estimating parasitemia levels were revealed with laboratory-grown P. falciparum strain (3D7). The LAMP assay performed in our device was exclusive for P. falciparum detection with sensitivity and specificity determined to be 98.89% and 100%, respectively, in clinical isolates. The LOD was documented to be 1 parasite/mu l at the cut-off ADC value of 20. Parasite density estimated from ADC values showed concordance with microscopically determined parasite density of the cultured P. falciparum 3D7 strain. The LAMP assay performed in our device provides a possible portable platform for its deployment in the point-of-care diagnosis of malaria. Further validation of the quantitative ability of the assay with freshly collected or properly stored clinical samples of known parasitemia is necessary for field applicability.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] POINT-OF-CARE DIAGNOSIS OF MALARIA BY FLUORESCENCE LOOP MEDIATED ISOTHERMAL AMPLIFICATION
    Lucchi, Naomi W.
    Demas, Allison
    Narayanan, Jothikumar
    Kabanywanyi, Abdunoor M.
    Sumari, Deborah
    Kachur, Patrick S.
    Barnwell, John W.
    Udhayakumar, Venkatachalam
    AMERICAN JOURNAL OF TROPICAL MEDICINE AND HYGIENE, 2010, 83 (05): : 285 - 286
  • [2] Loop-Mediated Isothermal Amplification as Point-of-Care Diagnosis for Neglected Parasitic Infections
    Avendano, Catalina
    Alfonso Patarroyo, Manuel
    INTERNATIONAL JOURNAL OF MOLECULAR SCIENCES, 2020, 21 (21) : 1 - 32
  • [3] Point-of-Care Diagnostic Platforms for Loop-Mediated Isothermal Amplification
    Atceken, Nazente
    Alseed, Muhammad Munzer
    Dabbagh, Sajjad Rahmani
    Yetisen, Ali K.
    Tasoglu, Savas
    ADVANCED ENGINEERING MATERIALS, 2023, 25 (08)
  • [4] Development of loop-mediated isothermal amplification with Plasmodium falciparum unique genes for molecular diagnosis of human malaria
    Zhang, Yijing
    Yao, Yi
    Du, Weixing
    Wu, Kai
    Xu, Wenyue
    Lin, Min
    Tan, Huabing
    Li, Jian
    PATHOGENS AND GLOBAL HEALTH, 2017, 111 (05) : 247 - 255
  • [5] Applying Loop-mediated Isothermal Amplification (LAMP) in the Diagnosis of Malaria, Leishmaniasis and Trypanosomiasis as Point-of-Care Tests (POCTs)
    Auxiliadora Dea-Ayuela, M.
    Galiana-Rosello, Cristina
    Lalatsa, Aikaterini
    Serrano, Dolores R.
    CURRENT TOPICS IN MEDICINAL CHEMISTRY, 2018, 18 (16) : 1358 - 1374
  • [6] Detection of Plasmodium vivax and Plasmodium falciparum DNA in human saliva and urine: Loop-mediated isothermal amplification for malaria diagnosis
    Najafabadi, Zahra Ghayour
    Oormazdi, Hormozd
    Akhlaghi, Lame
    Meamar, Ahmad Reza
    Nateghpour, Mehdi
    Farivar, Leila
    Razmjou, Elham
    ACTA TROPICA, 2014, 136 : 44 - 49
  • [7] Principles and Applications of Loop-Mediated Isothermal Amplification to Point-of-Care Tests
    Park, Jee-Woong
    BIOSENSORS-BASEL, 2022, 12 (10):
  • [8] Loop-Mediated Isothermal Amplification for Diagnosis of Zoonotic Malaria
    Lai, Meng Yee
    Sohairi, Ainaa Nadrah
    Zen, Lee Phone Youth
    Abdullah, Mohd Lutfi
    Lau, Yee Ling
    AMERICAN JOURNAL OF TROPICAL MEDICINE AND HYGIENE, 2024, 111 (04): : 765 - 769
  • [9] Loop-mediated isothermal amplification for point-of-care diagnosis of viral respiratory tract infection in childhood
    Peters, R.
    Pfeil, J.
    Tabatabai, J.
    Grulich-Henn, J.
    Schnitzler, P.
    JOURNAL OF CLINICAL VIROLOGY, 2016, 82 : S51 - S51
  • [10] Utility of loop-mediated isothermal amplification as a point-of-care test in diagnosis of amoebic liver abscess
    Handa, Dipti
    Gupta, Monica
    Lehl, Sarabmeet Singh
    Gupta, Amit
    Singh, Ram
    TROPICAL DOCTOR, 2021, 51 (04) : 488 - 491