Calibration transfer and drift counteraction in chemical sensor arrays using Direct Standardization

被引:160
作者
Fonollosa, J. [1 ,2 ,3 ]
Fernandez, L. [2 ,3 ]
Gutierrez-Galvez, A. [2 ,3 ]
Huerta, R. [1 ]
Marco, S. [2 ,3 ]
机构
[1] Univ Calif San Diego, Biocircuit Inst, La Jolla, CA 92093 USA
[2] Inst Bioengn Catalonia, Barcelona 08028, Spain
[3] Univ Barcelona, Dept Elect Engn, E-08028 Barcelona, Spain
关键词
Chemical sensors; Calibration transfer; MOX sensors; Electronic nose; Direct Standardization; Public dataset; SEMICONDUCTOR GAS SENSORS; ORTHOGONAL SIGNAL CORRECTION; ELECTRONIC NOSE; INSTRUMENTS;
D O I
10.1016/j.snb.2016.05.089
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Inherent variability of chemical sensors makes it necessary to calibrate chemical detection systems individually. This shortcoming has traditionally limited usability of systems based on metal oxide gas sensor arrays and prevented mass-production for some applications. Here, aiming at exploring calibration transfer between chemical sensor arrays, we exposed five twin 8-sensor detection units to different concentration levels of ethanol, ethylene, carbon monoxide, or methane. First, we built calibration models using data acquired with a master unit. Second, to explore the transferability of the calibration models, we used Direct Standardization to map the signals of a slave unit to the space of the master unit in calibration. In particular, we evaluated the transferability of the calibration models to other detection units, and within the same unit measuring days apart. Our results show that signals acquired with one unit can be successfully mapped to the space of a reference unit. Hence, calibration models trained with a master unit can be extended to slave units using a reduced number of transfer samples, diminishing thereby calibration costs. Similarly, signals of a sensing unit can be transformed to match sensor behavior in the past to mitigate drift effects. Therefore, the proposed methodology can reduce calibration costs in mass-production and delay recalibrations due to sensor aging. Acquired dataset is made publicly available.
引用
收藏
页码:1044 / 1053
页数:10
相关论文
共 45 条
[1]  
[Anonymous], PLS TOOLB 7 9
[2]   Transportability of data between electronic noses: mathematical methods [J].
Balaban, MO ;
Korel, F ;
Odabasi, AZ ;
Folkes, G .
SENSORS AND ACTUATORS B-CHEMICAL, 2000, 71 (03) :203-211
[3]   On 'Electronic Nose' methodology [J].
Boeker, Peter .
SENSORS AND ACTUATORS B-CHEMICAL, 2014, 204 :2-17
[4]  
Brown SD, 2009, COMPREHENSIVE CHEMOMETRICS: CHEMICAL AND BIOCHEMICAL DATA ANALYSIS, VOLS 1-4, P1
[5]  
Chaiyboun A., 2006, SENSORS, V6
[6]   CHARACTERISTICS OF SEMICONDUCTOR GAS SENSORS .1. STEADY-STATE GAS RESPONSE [J].
CLIFFORD, PK ;
TUMA, DT .
SENSORS AND ACTUATORS, 1983, 3 (03) :233-254
[7]   CHARACTERISTICS OF SEMICONDUCTOR GAS SENSORS .2. TRANSIENT-RESPONSE TO TEMPERATURE-CHANGE [J].
CLIFFORD, PK ;
TUMA, DT .
SENSORS AND ACTUATORS, 1983, 3 (03) :255-281
[8]   Gas concentration estimation in ternary mixtures with room temperature operating sensor array using tapped delay architectures [J].
De Vito, Saverio ;
Castaldo, Anna ;
Loffredo, Fausta ;
Massera, Ettore ;
Polichetti, Tiziana ;
Nasti, Lvana ;
Vacca, Paolo ;
Quercia, Luigi ;
Di Francia, Girolamo .
SENSORS AND ACTUATORS B-CHEMICAL, 2007, 124 (02) :309-316
[9]   Application of electronic nose for industrial odors and gaseous emissions measurement and monitoring - An overview [J].
Deshmukh, Sharvari ;
Bandyopadhyay, Rajib ;
Bhattacharyya, Nabarun ;
Pandey, R. A. ;
Jana, Arun .
TALANTA, 2015, 144 :329-340
[10]   Calibration transfer between electronic nose systems for rapid In situ measurement of pulp and paper industry emissions [J].
Deshmukh, Sharvari ;
Kamde, Kalyani ;
Jana, Arun ;
Korde, Sanjivani ;
Bandyopadhyay, Rajib ;
Sankar, Ravi ;
Bhattacharyya, Nabarun ;
Pandey, R. A. .
ANALYTICA CHIMICA ACTA, 2014, 841 :58-67