Hand-held nerve conduction device in carpal tunnel syndrome: A prospective study

被引:2
|
作者
Tan, S. Veronica [1 ]
Sandford, Fiona [2 ]
Stevenson, Mark [1 ]
Probert, Sara [3 ]
Sanders, Sue [1 ]
Mills, Kerry R. [4 ]
Koutroumanidis, Michaelis [1 ]
机构
[1] Guys & St Thomas Fdn NHS Trust, St Thomas Hosp, Dept Neurol & Clin Neurophysiol, London, England
[2] Guys & St Thomas Fdn NHS Trust, St Thomas Hosp, Dept Physiotherapy, London, England
[3] Guys & St Thomas Fdn NHS Trust, St Thomas Hosp, Dept Occupat Therapy, London, England
[4] Kings Coll London, London WC2R 2LS, England
关键词
carpal tunnel syndrome; EMG; hand-held nerve conduction device; neurophysiologic testing; peripheral nerve lesions; SCREENING TOOL; NC-STAT;
D O I
10.1002/mus.23279
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Introduction: We assessed the clinical impact of replacing standard neurophysiologic testing with a hand-held device (Mediracer) for diagnosis of carpal tunnel syndrome (CTS). Methods: One hundred patients (200 hands) with suspected CTS were studied by blinded assessors [Hand-therapist (HT) 1 and Consultant Neurophysiologist] using the Mediracer, followed by standard neurophysiologic testing. To simulate testing by personnel without neurological training, Mediracer recordings were analyzed separately by an assessor who had not seen the patients (HT2). Results: Correlation of the CTS grades was 0.94 for the results obtained by HT1, and 0.87 for HT2. The sensitivity and specificity of the Mediracer was 0.85 and 0.9, respectively, by HT1, and 0.84 and 0.89 for HT2. Nine patients had conditions other than CTS, and 35 patients were judged to require further investigation. Conclusions: The Mediracer should only be used in patients with typical CTS symptoms and signs and no muscle wasting who have had careful neurological assessment. Muscle Nerve 45: 635-641, 2012
引用
收藏
页码:635 / 641
页数:7
相关论文
共 50 条
  • [1] The optimum hand temperature to study nerve conduction in patients with carpal tunnel syndrome
    Gavanozi, Evangelia
    Veltsista, Dimitra
    Polychronopoulos, Panagiotis
    Chroni, Elisabeth
    JOURNAL OF ELECTROMYOGRAPHY AND KINESIOLOGY, 2020, 51
  • [2] A handheld nerve conduction measuring device in carpal tunnel syndrome
    Tolonen, U.
    Kallio, M.
    Ryhanen, J.
    Raatikainen, T.
    Honkala, V.
    Lesonen, V.
    ACTA NEUROLOGICA SCANDINAVICA, 2007, 115 (06): : 390 - 397
  • [3] Reliability of Hand-Held Dynamometer Measurement for Thumb Palmar Abduction Strength in Carpal Tunnel Syndrome
    Doi, Kazuteru
    Yasui, Noritada
    Isomura, Yuya
    Tagawa, Masafumi
    Hattori, Yasunori
    Sakamoto, Sotetsu
    JOURNAL OF HAND SURGERY-ASIAN-PACIFIC VOLUME, 2025, : 155 - 165
  • [4] The true utility and indication for hand-held nerve conduction devices
    Ahmed, Aiesha
    Kothari, Milind J.
    MUSCLE & NERVE, 2012, 45 (05) : 621 - 622
  • [5] Sensitivities of sensory nerve conduction study parameters in carpal tunnel syndrome
    Prakash, Kumar M.
    Fook-Chong, Stephanie
    Leoh, Teng Hee
    Dan, Yang Fang
    Nurjannah, Siti
    Tan, Yam Eng
    Lo, Yew Long
    JOURNAL OF CLINICAL NEUROPHYSIOLOGY, 2006, 23 (06) : 565 - 567
  • [6] Motor Nerve Conduction Tests in Carpal Tunnel Syndrome
    Kasius, Kristel M.
    Claes, Franka
    Verhagen, Wim I. M.
    Meulstee, Jan
    FRONTIERS IN NEUROLOGY, 2019, 10
  • [7] Carpal tunnel syndrome with normal nerve conduction studies
    Witt, JC
    Hentz, JG
    Stevens, JC
    MUSCLE & NERVE, 2004, 29 (04) : 515 - 522
  • [8] Use of nerve conduction studies in carpal tunnel syndrome
    Bland, Jeremy D. P.
    JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2023, 48 (10) : 976 - 985
  • [9] Carpal tunnel syndrome in the elderly Nerve conduction parameters
    Naves, Thiago Guimares
    Kouyoumdjian, Joao Aris
    ARQUIVOS DE NEURO-PSIQUIATRIA, 2010, 68 (01) : 87 - 92
  • [10] Effect of heating on nerve conduction in carpal tunnel syndrome
    Tilki, HE
    Stålberg, E
    Coskun, M
    Güngör, L
    JOURNAL OF CLINICAL NEUROPHYSIOLOGY, 2004, 21 (06) : 451 - 456