Implementation of cellulomonas cholesterol oxidase for total serum cholesterol determination by the endpoint method

被引:5
|
作者
Srisawasdi, Pornpen [1 ]
Chaichanajarernkul, Upsorn [1 ]
Teerakranjana, Narumon [1 ]
Kroll, Martin H. [2 ]
机构
[1] Mahidol Univ, Ramathibodi Hosp, Fac Med, Div Clin Chem,Dept Pathol, Bangkok 10400, Thailand
[2] Boston Univ, Med Ctr, Dept Lab Med, Boston, MA USA
关键词
serum cholesterol; Cellulomonas; Streptomyces; cholesterol oxidase; enzymatic endpoint method; implementation; coronary heart disease;
D O I
10.1002/jcla.20221
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Cellulomonas has been shown to be a good source of cholesterol oxidase in addition to Streptomyces for serum cholesterol determination by the endpoint method, inexpensive in cost, and showing excellent performance. For clinical use, we have assessed the reliability of Cellulomonas reagent for cholesterol determination. We constructed the user-defined endpoint methods on three automated analyzers. The analytical performances (linearity, precision, recovery, interference, stability, and comparison with the standardized method) of Cellulomonas cholesterol reagents were evaluated and compared to those of Streptomyces reagents. Linearity (18.1-23.3 mmol/L) and stability of reagents (6-11 weeks) depended on the analyzers being used. The average within-run and between-day % coefficients of variation (CVs) ranged from 1.44 to 2.45 and 1.98 to 2.99, respectively, and were within National Cholesterol Education Program analytical criteria (<= 3%). All assays using both reagents compared favorably with the commercial method and appeared accurate near the clinical decision cut-points. Hemoglobin concentration at 7.5g/L only affected the assay using single wavelength measurement. Bilirubin decreased in serum cholesterol recovery while lipemia generated a positive interference with all methods. Cellulomonas enzyme is analytically reliable when used for serum cholesterol determination by the endpoint method. Its analytical performance is equivalent to Streptomyces enzymes and meets the analytical goals. It has an advantage over the other enzymes in that it does not ship in the frozen state.
引用
收藏
页码:50 / 58
页数:9
相关论文
共 50 条
  • [1] Performance of four sources of cholesterol oxidase for serum cholesterol determination by the enzymatic endpoint method
    Lolekha, PH
    Srisawasdi, P
    Jearanaikoon, P
    Wetprasit, N
    Sriwanthana, B
    Kroll, MH
    CLINICA CHIMICA ACTA, 2004, 339 (1-2) : 135 - 145
  • [2] Performance characteristics of cholesterol oxidase for kinetic determination of total cholesterol
    Srisawasdi, P
    Jearanaikoon, P
    Kroll, MH
    Lolekha, PH
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2005, 19 (06) : 247 - 252
  • [3] Direct fluorometric determination of total cholesterol in serum using derivatized cholesterol oxidase
    Galbán, J
    Sierra, JF
    Sebastián, JML
    de Marcos, S
    Castillo, JR
    APPLIED SPECTROSCOPY, 2000, 54 (08) : 1157 - 1162
  • [4] Application of Streptomyces and Brevibacterium cholesterol oxidase for total serum cholesterol assay by the enzymatic kinetic method
    Srisawasdi, Pornpen
    Jearanaikoon, Patcharee
    Wetprasit, Nuanchawee
    Sriwanthana, Busarawan
    Kroll, Martin H.
    Lolekha, Porntip H.
    CLINICA CHIMICA ACTA, 2006, 372 (1-2) : 103 - 111
  • [5] STREPTOMYCES - A SUPERIOR SOURCE FOR CHOLESTEROL OXIDASE USED IN SERUM-CHOLESTEROL ASSAY
    LOLEKHA, PH
    JANTAVEESIRIRAT, Y
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 1992, 6 (06) : 405 - 409
  • [6] Advantages and disadvantages of serum cholesterol determination by the kinetic vs the end point method
    Srisawasdi, Pornpen
    Kroll, Martin H.
    Lolekha, Porntip H.
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2007, 127 (06) : 906 - 918
  • [7] Co-immobilization of cholesterol esterase, cholesterol oxidase and peroxidase onto alkylamine glass beads for measurement of total cholesterol in serum
    Suman
    Pundir, CS
    CURRENT APPLIED PHYSICS, 2003, 3 (2-3) : 129 - 133
  • [8] Porcine Pancreas: A Superior Source of Cholesterol Esterase for Total Serum Cholesterol Assay by the Enzymatic Kinetic Method
    Srisawasdi, Pornpen
    Prasertsincharoen, Noppadol
    Kroll, Martin H.
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2012, 26 (06) : 420 - 430
  • [9] Preparation of cholesterol oxidase nanoparticles and their application in amperometric determination of cholesterol
    Sheetal Chawla
    Rachna Rawal
    C. S. Sonia
    Journal of Nanoparticle Research, 2013, 15
  • [10] Preparation of cholesterol oxidase nanoparticles and their application in amperometric determination of cholesterol
    Chawla, Sheetal
    Rawal, Rachna
    Sonia
    Ramrati
    Pundir, C. S.
    JOURNAL OF NANOPARTICLE RESEARCH, 2013, 15 (09)