Simulation of the artefact of an iodine seed placed at the needle tip in MRI-guided prostate brachytherapy

被引:14
作者
Lagerburg, V. [1 ]
Moerland, M. A. [1 ]
Seppenwoolde, J. H. [2 ]
Lagendijk, J. J. W. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiotherapy, NL-3584 CX Utrecht, Netherlands
[2] Univ Med Ctr Utrecht, Dept Radiol, NL-3584 CX Utrecht, Netherlands
关键词
D O I
10.1088/0031-9155/53/5/N01
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this research is to study the influence of different needle materials on the artefact of a prostate brachytherapy iodine seed, placed at the needle tip, in MRI (magnetic resonance imaging)-guided prostate brachytherapy. For this research simulations were performed. The simulations showed that with the currently available MRI compatible titanium needles, determination of the exact seed position is difficult, because of the large artefact at the needle tip. This hampers accurate MRI-guided seed delivery. When a plastic needle is used, the image disturbance is caused by the artefact of the iodine seed alone. When a gradient echo sequence is used, the middle of the seed artefact corresponds well with to middle of the real seed position. With the scan parameters we used this deviation was less than 0.4 mm compared to 1.5 mm when a titanium needle is used.
引用
收藏
页码:N59 / N67
页数:9
相关论文
共 19 条
[1]   SIMULATION OF SUSCEPTIBILITY ARTIFACTS IN 2D AND 3D FOURIER-TRANSFORM SPIN-ECHO AND GRADIENT-ECHO MAGNETIC-RESONANCE-IMAGING [J].
BAKKER, CJG ;
BHAGWANDIEN, R ;
MOERLAND, MA ;
RAMOS, LMP .
MAGNETIC RESONANCE IMAGING, 1994, 12 (05) :767-774
[2]   Dosimetric study of a new polymer encapsulated palladium-103 seed [J].
Bernard, S ;
Vynckier, S .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (07) :1493-1504
[3]   NUMERICAL-ANALYSIS OF THE MAGNETIC-FIELD FOR ARBITRARY MAGNETIC-SUSCEPTIBILITY DISTRIBUTIONS IN 3D [J].
BHAGWANDIEN, R ;
MOERLAND, MA ;
BAKKER, CJG ;
BEERSMA, R ;
LAGENDIJK, JJW .
MAGNETIC RESONANCE IMAGING, 1994, 12 (01) :101-107
[4]   NUMERICAL-ANALYSIS OF THE MAGNETIC-FIELD FOR ARBITRARY MAGNETIC-SUSCEPTIBILITY DISTRIBUTIONS IN 2D [J].
BHAGWANDIEN, R ;
VANEE, R ;
BEERSMA, R ;
BAKKER, CJG ;
MOERLAND, MA ;
LAGENDIJK, JJW .
MAGNETIC RESONANCE IMAGING, 1992, 10 (02) :299-313
[5]   Real-time magnetic resonance image-guided interstitial brachytherapy in the treatment of select patients with clinically localized prostate cancer [J].
D'Amico, AV ;
Cormack, R ;
Tempany, CM ;
Kumar, S ;
Topulos, G ;
Kooy, HM ;
Coleman, CN .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (03) :507-515
[6]   DOSE EFFECTS OF SEEDS PLACEMENT DEVIATIONS FROM PRE-PLANNED POSITIONS IN ULTRASOUND-GUIDED PROSTATE IMPLANTS [J].
DAWSON, JE ;
WU, T ;
ROY, T ;
GU, JY ;
KIM, H .
RADIOTHERAPY AND ONCOLOGY, 1994, 32 (03) :268-270
[7]   Prostate biopsy in the supine position in a standard 1.5-T scanner under real time MR-imaging control using a MR-compatible endorectal biopsy device [J].
Engelhard, K ;
Hollenbach, HP ;
Kiefer, B ;
Winkel, A ;
Goeb, K ;
Engehausen, D .
EUROPEAN RADIOLOGY, 2006, 16 (06) :1237-1243
[8]   Monte Carlo dosimetry of the selectSeed 125I interstitial brachytherapy seed [J].
Karaiskos, P ;
Papagiannis, P ;
Sakelliou, L ;
Anagnostopoulos, G ;
Baltas, D .
MEDICAL PHYSICS, 2001, 28 (08) :1753-1760
[9]  
Konings MK, 2000, J MAGN RESON IMAGING, V12, P79, DOI 10.1002/1522-2586(200007)12:1<79::AID-JMRI9>3.3.CO
[10]  
2-K