Image quality and dose differences caused by vendor-specific image processing of neonatal radiographs

被引:8
|
作者
Sensakovic, William F. [1 ]
O'Dell, M. Cody [1 ]
Letter, Haley [1 ]
Kohler, Nathan [1 ]
Rop, Baiywo [1 ]
Cook, Jane [1 ]
Logsdon, Gregory [1 ]
Varich, Laura [1 ]
机构
[1] Florida Hosp, Imaging Adm, 601 E Rollins St, Orlando, FL 32803 USA
关键词
Chest; Image processing; Image quality; Neonate; Radiation dose; Radiography; IONIZING-RADIATION; CHEST; EXPOSURE; POPULATION; REDUCTION; INFANTS; DISEASE; BACK; UNIT;
D O I
10.1007/s00247-016-3663-2
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Image processing plays an important role in optimizing image quality and radiation dose in projection radiography. Unfortunately commercial algorithms are black boxes that are often left at or near vendor default settings rather than being optimized. We hypothesize that different commercial image-processing systems, when left at or near default settings, create significant differences in image quality. We further hypothesize that image-quality differences can be exploited to produce images of equivalent quality but lower radiation dose. We used a portable radiography system to acquire images on a neonatal chest phantom and recorded the entrance surface air kerma (ESAK). We applied two image-processing systems (Optima XR220amx, by GE Healthcare, Waukesha, WI; and MUSICA(2) by Agfa HealthCare, Mortsel, Belgium) to the images. Seven observers (attending pediatric radiologists and radiology residents) independently assessed image quality using two methods: rating and matching. Image-quality ratings were independently assessed by each observer on a 10-point scale. Matching consisted of each observer matching GE-processed images and Agfa-processed images with equivalent image quality. A total of 210 rating tasks and 42 matching tasks were performed and effective dose was estimated. Median Agfa-processed image-quality ratings were higher than GE-processed ratings. Non-diagnostic ratings were seen over a wider range of doses for GE-processed images than for Agfa-processed images. During matching tasks, observers matched image quality between GE-processed images and Agfa-processed images acquired at a lower effective dose (11 +/- 9 mu Sv; P < 0.0001). Image-processing methods significantly impact perceived image quality. These image-quality differences can be exploited to alter protocols and produce images of equivalent image quality but lower doses. Those purchasing projection radiography systems or third-party image-processing software should be aware that image processing can significantly impact image quality when settings are left near default values.
引用
收藏
页码:1606 / 1613
页数:8
相关论文
共 50 条
  • [31] New Developed DR Detector Performs Radiographs of Hand, Pelvic and Premature Chest Anatomies at a Lower Radiation Dose and/or a Higher Image Quality
    Helle Precht
    Anders Tingberg
    Dag Waaler
    Claus Bjørn Outzen
    Journal of Digital Imaging, 2014, 27 : 68 - 76
  • [32] Visual grading analysis of digital neonatal chest phantom X-ray images: Impact of detector type, dose and image processing on image quality
    Smet, M. H.
    Breysem, L.
    Mussen, E.
    Bosmans, H.
    Marshall, N. W.
    Cockmartin, L.
    EUROPEAN RADIOLOGY, 2018, 28 (07) : 2951 - 2959
  • [33] Assessment of Radiation Dose and Image Quality of Multidetector Computed Tomography
    Al Ewaidat, Haytham
    Zheng, Xiaoming
    Khader, Yousef
    Abdelrahman, Mostafa
    Alhasan, Mostafa Khaled Mustafa
    Rawashdeh, Mohammad Ahmmad
    Al Mousa, Dana Samir
    Alawneh, Khaled Zayed Ali
    IRANIAN JOURNAL OF RADIOLOGY, 2018, 15 (03)
  • [34] An evaluation of the effect of tube potential on clinical image quality using direct digital detectors for pelvis and lumbar spine radiographs
    Peacock, Nicole E.
    Steward, Adam L.
    Riley, Peter J.
    JOURNAL OF MEDICAL RADIATION SCIENCES, 2020, 67 (04) : 260 - 268
  • [35] Image processing setting adaptions according to image dose and radiologist preference can improve image quality in computed radiography of the equine distal limb: A cadaveric study
    Seeber, Matthias
    Lederer, Kristina A.
    Rowan, Conor
    Strohmayer, Carina
    Ludewig, Eberhard
    VETERINARY RADIOLOGY & ULTRASOUND, 2024, 65 (01) : 19 - 30
  • [36] Chest computed tomography using iterative reconstruction vs filtered back projection (Part 2): image quality of low-dose CT examinations in 80 patients
    Pontana, Francois
    Duhamel, Alain
    Pagniez, Julien
    Flohr, Thomas
    Faivre, Jean-Baptiste
    Hachulla, Anne-Lise
    Remy, Jacques
    Remy-Jardin, Martine
    EUROPEAN RADIOLOGY, 2011, 21 (03) : 636 - 643
  • [37] Patient size matters: Effect of tube current modulation on size-specific dose estimates (SSDE) and image quality in low-dose lung cancer screening CT
    Barreto, Izabella
    Verma, Nupur
    Quails, Nathan
    Olguin, Catherine
    Correa, Nathalie
    Mohammed, Tan-Lucien
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (04): : 87 - 94
  • [38] MAMMOGRAPHY IN SERBIA: IMAGE QUALITY AND RADIATION DOSE
    Rafajlovic, Stefan
    Arandjic, Danijela
    Bozovic, Predrag
    Ceklic, Sandra
    Lazarevic, Djordje
    Ciraj-Bjelac, Olivera
    FOURTH INTERNATIONAL CONFERENCE ON RADIATION AND APPLICATIONS IN VARIOUS FIELDS OF RESEARCH (RAD 2016), 2016, : 29 - 32
  • [39] Dose and image quality in CT
    Walter Huda
    Pediatric Radiology, 2002, 32 : 709 - 713
  • [40] Dose and perceived image quality in chest radiography
    Veldkamp, Wouter J. H.
    Kroft, Lucia J. M.
    Geleijns, Jacob
    EUROPEAN JOURNAL OF RADIOLOGY, 2009, 72 (02) : 209 - 217