Image quality and dose differences caused by vendor-specific image processing of neonatal radiographs

被引:8
|
作者
Sensakovic, William F. [1 ]
O'Dell, M. Cody [1 ]
Letter, Haley [1 ]
Kohler, Nathan [1 ]
Rop, Baiywo [1 ]
Cook, Jane [1 ]
Logsdon, Gregory [1 ]
Varich, Laura [1 ]
机构
[1] Florida Hosp, Imaging Adm, 601 E Rollins St, Orlando, FL 32803 USA
关键词
Chest; Image processing; Image quality; Neonate; Radiation dose; Radiography; IONIZING-RADIATION; CHEST; EXPOSURE; POPULATION; REDUCTION; INFANTS; DISEASE; BACK; UNIT;
D O I
10.1007/s00247-016-3663-2
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Image processing plays an important role in optimizing image quality and radiation dose in projection radiography. Unfortunately commercial algorithms are black boxes that are often left at or near vendor default settings rather than being optimized. We hypothesize that different commercial image-processing systems, when left at or near default settings, create significant differences in image quality. We further hypothesize that image-quality differences can be exploited to produce images of equivalent quality but lower radiation dose. We used a portable radiography system to acquire images on a neonatal chest phantom and recorded the entrance surface air kerma (ESAK). We applied two image-processing systems (Optima XR220amx, by GE Healthcare, Waukesha, WI; and MUSICA(2) by Agfa HealthCare, Mortsel, Belgium) to the images. Seven observers (attending pediatric radiologists and radiology residents) independently assessed image quality using two methods: rating and matching. Image-quality ratings were independently assessed by each observer on a 10-point scale. Matching consisted of each observer matching GE-processed images and Agfa-processed images with equivalent image quality. A total of 210 rating tasks and 42 matching tasks were performed and effective dose was estimated. Median Agfa-processed image-quality ratings were higher than GE-processed ratings. Non-diagnostic ratings were seen over a wider range of doses for GE-processed images than for Agfa-processed images. During matching tasks, observers matched image quality between GE-processed images and Agfa-processed images acquired at a lower effective dose (11 +/- 9 mu Sv; P < 0.0001). Image-processing methods significantly impact perceived image quality. These image-quality differences can be exploited to alter protocols and produce images of equivalent image quality but lower doses. Those purchasing projection radiography systems or third-party image-processing software should be aware that image processing can significantly impact image quality when settings are left near default values.
引用
收藏
页码:1606 / 1613
页数:8
相关论文
共 50 条
  • [21] Effect of additional filtration on radiation dose and image quality
    Gai, Xinghui
    Yang, Dewu
    Li, Xunru
    Sui, Yan
    Liu, Kang
    Wang, Junying
    Zhao, Mengjuan
    2022 IEEE 6TH ADVANCED INFORMATION TECHNOLOGY, ELECTRONIC AND AUTOMATION CONTROL CONFERENCE (IAEAC), 2022, : 1807 - 1811
  • [22] Evaluating the effect of reduced entrance surface dose on neonatal chest imaging using subjective image quality evaluation
    Conradie, Anita
    Herbst, Charles Petrus
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (10): : 1368 - 1374
  • [23] Image quality and dose in computed tomography
    Jurik, AG
    Jessen, KA
    Hansen, J
    EUROPEAN RADIOLOGY, 1997, 7 (01) : 77 - 81
  • [24] Tradeoffs between image quality and dose
    J. Anthony Seibert
    Pediatric Radiology, 2004, 34 : S183 - S195
  • [25] Image quality and dose in computed tomography
    A. G. Jurik
    K. A. Jessen
    J. Hansen
    European Radiology, 1997, 7 : 77 - 81
  • [26] Tradeoffs between image quality and dose
    Seibert, JA
    PEDIATRIC RADIOLOGY, 2004, 34 (Suppl 3) : S183 - S195
  • [27] Reducing CT radiation dose with iterative reconstruction algorithms: The influence of scan and reconstruction parameters on image quality and CTDIvol
    Klink, Thorsten
    Obmann, Verena
    Heverhagen, Johannes
    Stork, Alexander
    Adam, Gerhard
    Begemann, Philipp
    EUROPEAN JOURNAL OF RADIOLOGY, 2014, 83 (09) : 1645 - 1654
  • [28] Radiation dose and image quality in pediatric chest CT: effects of iterative reconstruction in normal weight and overweight children
    Yoon, Haesung
    Kim, Myung-Joon
    Yoon, Choon-Sik
    Choi, Jiin
    Shin, Hyun Joo
    Kim, Hyun Gi
    Lee, Mi-Jung
    PEDIATRIC RADIOLOGY, 2015, 45 (03) : 337 - 344
  • [29] A New Method for the Diagnosis of Urinary tract stone in Radiographs with Image Processing
    Sadeghi, Mostafa
    Memarzadeh-zavareh, Faezeh
    Shafiee, Masoud
    Shafieirad, Hossein
    PROCEEDINGS OF 2012 2ND INTERNATIONAL CONFERENCE ON COMPUTER SCIENCE AND NETWORK TECHNOLOGY (ICCSNT 2012), 2012, : 2242 - 2244
  • [30] New Developed DR Detector Performs Radiographs of Hand, Pelvic and Premature Chest Anatomies at a Lower Radiation Dose and/or a Higher Image Quality
    Precht, Helle
    Tingberg, Anders
    Waaler, Dag
    Outzen, Claus Bjorn
    JOURNAL OF DIGITAL IMAGING, 2014, 27 (01) : 68 - 76