An applicable approach for extracting human heart rate and oxygen saturation during physical movements using a multi-wavelength illumination optoelectronic sensor system

被引:9
作者
Alharbi, Samah [1 ]
Hu, Sijung [1 ]
Mulvaney, David [1 ]
Blanos, Panagiotis [1 ]
机构
[1] Loughborough Univ, Wolfson Sch Mech Elect & Mfg Engn, Ashby Rd, Loughborough LE11 3TU, Leics, England
来源
DESIGN AND QUALITY FOR BIOMEDICAL TECHNOLOGIES XI | 2018年 / 10486卷
关键词
Heart rate monitoring; Oxygen saturation monitoring; Optoelectronic patch sensor; deoxyhemoglobin; oxyhemoglobin;
D O I
10.1117/12.2287854
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The ability to gather physiological parameters such as heart rate (HR) and oxygen saturation (SpO2%) during physical movement allows to continuously monitor personal health status without disrupt their normal daily activities. Photoplethysmography (PPG) based pulse oximetry and similar principle devices are unable to extract the HR and SpO2% reliably during physical movement due to interference in the signals that arise from motion artefacts (MAs). In this research, a flexible reflectance multi-wavelength optoelectronic patch sensor (OEPS) has been developed to overcome the susceptibility of conventional pulse oximetry readings to MAs. The OEPS incorporates light embittered diodes as illumination sources with four different wavelengths, e.g. green, orange, red, and infrared unlike the conventional pulse oximetry devices that normally measure the skin absorption of only two wavelengths (red and infrared). The additional green and orange wavelengths were found to be distinguish to the absorption of deoxyhemoglobin (RHb) and oxyhemoglobin (HbO2). The reliability of extracting physiological parameters from the green and orange wavelengths is due to absorbed near to the surface of the skin, thereby shortening the optical path and so effectively reducing the influence of physical movements. To compensate of MAs, a three-axis accelerometer was used as a reference with help of adaptive filter to reduce MAs. The experiments were performed using 15 healthy subjects aged 20 to 30. The primary results show that there are no significant difference of heart rate and oxygen saturation measurements between commercial devices and OEPS Green (r=0.992), Orange(r=0.984), Red(r=0.952) and IR(r=0.97) and SpO2% (r = 0.982, p = 0.894).
引用
收藏
页数:15
相关论文
共 25 条
[1]  
Alzahrani A. A., 2016, THESIS
[2]   A Comparative Study of Physiological Monitoring with a Wearable Opto-Electronic Patch Sensor (OEPS) for Motion Reduction [J].
Alzahrani, Abdullah ;
Hu, Sijung ;
Azorin-Peris, Vicente .
BIOSENSORS-BASEL, 2015, 5 (02) :288-307
[3]   AMON:: A wearable multiparameter medical monitoring and alert system [J].
Anliker, U ;
Ward, JA ;
Lukowicz, P ;
Tröster, G ;
Dolveck, F ;
Baer, M ;
Keita, F ;
Schenker, EB ;
Catarsi, F ;
Coluccini, L ;
Belardinelli, A ;
Shklarski, D ;
Alon, M ;
Hirt, E ;
Schmid, R ;
Vuskovic, M .
IEEE TRANSACTIONS ON INFORMATION TECHNOLOGY IN BIOMEDICINE, 2004, 8 (04) :415-427
[4]  
Fodor L., 2011, AESTHETIC APPL INTEN, P133
[5]   Validity of a multi-sensor armband in estimating rest and exercise energy expenditure [J].
Fruin, ML ;
Rankin, JW .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2004, 36 (06) :1063-1069
[6]  
Grossman Paul, 2004, Stud Health Technol Inform, V108, P133
[7]  
H3 SYS IEM, 2017, HRM 3200 HEART RAT M
[8]  
Huawei, 2017, HON WATCH S1
[9]  
Jayasree V., 2009, Selected cardiovascular studies based on photoplethysmography technique
[10]  
Lee K., 2013, C P IEEE ENG MED BIO, V10, P4577