Toward development of a point-of-care assay of enoxaparin anticoagulant activity in whole blood

被引:8
作者
Inchiosa, Mario A., Jr. [1 ,2 ]
Pothula, Suryanarayana [2 ]
Kubal, Keshar [1 ,2 ]
Sanchala, Vajubhai T. [2 ]
Navarro, Iris [2 ]
机构
[1] New York Med Coll, Dept Pharmacol, Valhalla, NY 10595 USA
[2] New York Med Coll, Dept Anesthesiol, Valhalla, NY 10595 USA
关键词
Low-molecular-weight-heparin; Low-molecular-weight-heparin assay; LMWH point-of-care assay; Anti-factor Xa activity; MOLECULAR-WEIGHT HEPARIN; IN-VITRO; UNFRACTIONATED HEPARIN; COAGULATION ASSAYS; RENAL IMPAIRMENT; PLASMA;
D O I
10.1007/s11239-010-0546-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
There is need for a rapid assay to determine the efficacy of low-molecular-weight-heparin (LMWH) in whole blood. Heparinase was used to eliminate, and thereby quantify, the anticoagulant activity of the low-molecular-weight-heparin, enoxaparin. The percent change in the clotting time of whole blood in the presence of heparinase yielded the anticoagulant contribution of enoxaparin. A minimally activated assay (MAA) of whole blood clotting time was evaluated for the detection and relative quantification of enoxaparin. The assay cartridge consisted of a plain glass tube and detection magnet, with no additional sources of activation. Comparisons were also made with a point-of-care, activated partial thromboplastin time (aPTT) assay. Plasma or whole blood was spiked with enoxaparin at concentrations ranging from 0.1 to 1.0 anti-factor Xa IU/ml. A commercial preparation of heparinase I was used to digest enoxaparin, and clotting times were determined with and without heparinase incubation. Heparinase digestion caused an average shortening of clotting time of 21.1% (47.3 s) in blood spiked with 0.4 anti-Xa IU/ml enoxaparin, an amount expected in the therapeutic range; also, 0.1 anti-Xa IU/ml of enoxaparin could be reliably detected. The assay performed comparably when transferred to a point-of-care setting with heparinase being added directly to citrated blood collection tubes, followed by either MAA or aPTT assay. Strong correlations were obtained with both assays between the percent change in clotting time (after heparinase) and the added concentration of enoxaparin, or in comparison with the chromogenically measured concentration of enoxaparin. The assays for an individual blood sample can be completed within 10 min.
引用
收藏
页码:47 / 53
页数:7
相关论文
共 50 条
[21]   An integrated microfluidic electrochemical assay for cervical cancer detection at point-of-care testing [J].
Keyvani, Fatemeh ;
Debnath, Nandini ;
Saleh, Mahmoud Ayman ;
Poudineh, Mahla .
NANOSCALE, 2022, 14 (18) :6761-6770
[22]   Point-of-Care Microdevices for Blood Plasma Analysis in Viral Infectious Diseases [J].
Yin-Ting Yeh ;
Merisa Nisic ;
Xu Yu ;
Yiqiu Xia ;
Si-Yang Zheng .
Annals of Biomedical Engineering, 2014, 42 :2333-2343
[23]   Point-of-Care Microdevices for Blood Plasma Analysis in Viral Infectious Diseases [J].
Yeh, Yin-Ting ;
Nisic, Merisa ;
Yu, Xu ;
Xia, Yiqiu ;
Zheng, Si-Yang .
ANNALS OF BIOMEDICAL ENGINEERING, 2014, 42 (11) :2333-2343
[24]   Point-of-care Testing of Portable Blood Coagulation Detectors Using Optical Sensors [J].
Yang, Chin-Lung ;
Chiou, Yan-Chao ;
Chou, Cheng-Wei ;
Young, Kung-Chia ;
Huang, Song-Jeng ;
Liu, Chih-Yi .
JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2013, 33 (03) :319-324
[25]   Point-of-care seeding of nitinol stents with blood-derived endothelial cells [J].
Jantzen, Alexandra E. ;
Noviani, Maria ;
Mills, James S. ;
Baker, Katherine M. ;
Lin, Fu-Hsiung ;
Truskey, George A. ;
Achneck, Hardean E. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2016, 104 (08) :1658-1665
[26]   Reliability of a point-of-care device to determine oxidative stress in whole blood before and after acute exercise: A practical approach for the applied sports sciences [J].
Quinn, Karlee M. ;
Cox, Amanda J. ;
Roberts, Llion A. ;
Briskey, David ;
Minahan, Clare .
JOURNAL OF SPORTS SCIENCES, 2021, 39 (06) :673-682
[27]   Development in competitive immunoassay of a point-of-care testing for cotinine (COT) detection in urine [J].
Konziw, Suthinee ;
Tunakhun, Paweena ;
Ngernpimai, Sawinee ;
Srichaiyapol, Oranee ;
Boonsiri, Patcharee ;
Tippayawat, Patcharaporn ;
Techasen, Anchalee ;
Maraming, Pornsuda ;
Choowongkomon, Kiattawee ;
Daduang, Sakda ;
Promdee, Limthong ;
Daduang, Jureerut .
ANALYTICAL METHODS, 2024, 16 (26) :4387-4394
[28]   Blood Glucose Point-of-Care Testing Quality Assessment and Harmonization With Central Laboratory Assays [J].
Jacobs, Ellis ;
Naskalski, Jerzy W. ;
Fogh-Andersen, Niels ;
Ritter, Christoph ;
Lewenstam, Andrezj ;
Maas, Anton ;
Mager, Gerard ;
Okorodudu, Anthony O. .
POINT OF CARE, 2009, 8 (02) :82-86
[29]   Comparison of three point-of-care blood glucose meters for use in adult and juvenile alpacas [J].
Tennent-Brown, Brett S. ;
Koenig, Amie ;
Williamson, Lisa H. ;
Boston, Raymond C. .
JAVMA-JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2011, 239 (03) :380-386
[30]   Standardization of whole blood assay for determination of pyrogenic activity in organic dust samples [J].
Liebers, Verena ;
Stubel, Heike ;
Dueser, Maria ;
Bruening, Thomas ;
Raulf-Heimsoth, Monika .
INTERNATIONAL JOURNAL OF HYGIENE AND ENVIRONMENTAL HEALTH, 2009, 212 (05) :547-556